Respiratory mechanics during high-frequency oscillatory ventilation: a physical model and preterm infant study

被引:14
|
作者
Singh, Rachana [2 ]
Courtney, Sherry E. [3 ]
Weisner, Michael D. [4 ]
Habib, Robert H. [1 ]
机构
[1] Amer Univ Beirut, Med Ctr, Clin Res Inst, Outcomes Res Unit,Dept Internal Med, Beirut 11072020, Lebanon
[2] Baystate Med Ctr, Div Newborn Med, Springfield, MA USA
[3] SUNY Stony Brook, Med Ctr, Div Neonatol, Stony Brook, NY USA
[4] Equilibrated Bio Syst Inc, Smithtown, NY USA
关键词
mechanical ventilation; endotracheal tube leak; tracheal pressure monitoring; very low birth weight; UNCUFFED ENDOTRACHEAL-TUBES; PRESSURE; FUTURE;
D O I
10.1152/japplphysiol.01120.2011
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Singh R, Courtney SE, Weisner MD, Habib RH. Respiratory mechanics during high-frequency oscillatory ventilation: a physical model and preterm infant study. J Appl Physiol 112: 1105-1113, 2012. First published December 29, 2011; doi:10.1152/japplphysiol.01120.2011.- Accurate mechanics measurements during high-frequency oscillatory ventilation (HFOV) facilitate optimizing ventilator support settings. Yet, these are influenced substantially by endotracheal tube (ETT) contributions, which may dominate when leaks around uncuffed ETT are present. We hypothesized that 1) the effective removal of ETT leaks may be confirmed via direct comparison of measured vs. model-predicted mean intratracheal pressure [mPtr (meas) vs. mPtr (pred)], and 2) reproducible respiratory system resistance (Rrs) and compliance (Crs) may be derived from no-leak oscillatory Ptr and proximal flow. With the use of ETT test-lung models, proximal airway opening (Pao) and distal (Ptr) pressures and flows were measured during slow-cuff inflations until leaks are removed. These were repeated for combinations of HFOV settings [frequency, mean airway pressure (Paw), oscillation amplitudes (Delta P), and inspiratory time (%t(I))] and varying test-lung Crs. Results showed that leaks around the ETT will 1) systematically reduce the effective distending pressures and lung-delivered oscillatory volumes, and 2) derived mechanical properties are increasingly nonphysiologic as leaks worsen. Mean pressures were systematically reduced along the ventilator circuit and ETT (Paw > Pao > Ptr), even for no-leak conditions. ETT size-specific regression models were then derived for predicting mPtr based on mean Pao (mPao), Delta P, %t(I), and frequency. Next, in 10 of 11 studied preterm infants (0.77 +/- 0.24 kg), no-to-minimal leak was confirmed based on excellent agreement between mPtr (meas) and mPtr (pred), and consequently, their oscillatory respiratory mechanics were evaluated. Infant resistance at the proximal ETT (R-ETT; resistance airway opening = R-ETT + Rrs; P < 0.001) and ETT inertance (P = 0.014) increased significantly with increasing Delta n P (50%, 100%, and 150% baseline), whereas Rrs showed a modest, nonsignificant increase (P = 0.14), and Crs was essentially unchanged (P = 0.39). We conclude that verifying no-leak conditions is feasible by comparison of model-derived vs. distending mPtr (meas). This facilitated the reliable and accurate assessment of physiologic respiratory mechanical properties that can objectively guide ventilatory management of HFOV-treated preterm infants.
引用
收藏
页码:1105 / 1113
页数:9
相关论文
共 50 条
  • [41] Effect of high-frequency oscillatory ventilation combined with volume guarantee on preterm infants with hypoxic respiratory failure
    Chen, Lih-Ju
    Chen, Jia-Yuh
    JOURNAL OF THE CHINESE MEDICAL ASSOCIATION, 2019, 82 (11) : 861 - 864
  • [42] Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study
    Osamu Nagano
    Tetsuya Yumoto
    Atsunori Nishimatsu
    Shunsuke Kanazawa
    Takahisa Fujita
    Sunao Asaba
    Hideo Yamanouchi
    Intensive Care Medicine Experimental, 6 (1)
  • [43] Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study
    Nagano, Osamu
    Yumoto, Tetsuya
    Nishimatsu, Atsunori
    Kanazawa, Shunsuke
    Fujita, Takahisa
    Asaba, Sunao
    Yamanouchi, Hideo
    INTENSIVE CARE MEDICINE EXPERIMENTAL, 2018, 6
  • [44] Non-invasive High-Frequency Oscillatory Ventilation as Initial Respiratory Support for Preterm Infants With Respiratory Distress Syndrome
    Lai, Shu-Hua
    Xie, Ying-Ling
    Chen, Zhi-Qing
    Chen, Rong
    Cai, Wen-Hong
    Wu, Luo-Cheng
    Lin, Yun-Feng
    Zheng, Yi-Rong
    FRONTIERS IN PEDIATRICS, 2022, 9
  • [45] Survival of a 300-g infant ventilated by high-frequency oscillatory ventilation for respiratory distress syndrome
    Kohelet, D
    Arbel, E
    Shochat, R
    Tavori, I
    CRITICAL CARE MEDICINE, 2002, 30 (05) : 1152 - 1155
  • [46] High-Frequency Oscillatory Ventilation in an Infant With Necrotizing Pneumonia and Bronchopleural Fistula
    Hasan, Rashed A.
    Al-Neyadi, Shaikha
    Abuhasna, Said
    Black, Craig P.
    RESPIRATORY CARE, 2011, 56 (03) : 351 - 354
  • [47] Carbon Dioxide Entrainment During High-Frequency Oscillatory Ventilation: A Model Analysis
    Chatburn, Robert L.
    El-Khatib, Mohamad F.
    RESPIRATORY CARE, 2012, 57 (11) : 1984 - 1988
  • [48] Inhaled Prostacyclin and High-Frequency Oscillatory Ventilation in a Premature Infant With Respiratory Syncytial Virus-Associated Respiratory Failure
    Gupta, Manoj
    Guertin, Stephen
    Martin, Steve
    Omar, Said
    PEDIATRICS, 2012, 130 (02) : E442 - E445
  • [49] HIGH-FREQUENCY OSCILLATORY VENTILATION AS A RESCUE STRATEGY FOR PATIENTS WITH RESPIRATORY FAILURE
    Thind, Guramrinder
    Duggal, Abhijit
    Chatburn, Robert
    Hatipoglu, Umur
    Mireles-Cabodevila, Eduardo
    CRITICAL CARE MEDICINE, 2020, 48
  • [50] High-frequency oscillatory ventilation in adults with acute respiratory distress syndrome
    Hager, David N.
    CURRENT OPINION IN ANESTHESIOLOGY, 2012, 25 (01) : 17 - 23