摘要
A simple, highly sensitive, precise and accurate high-performance liquid chromatographic (LCMSMS) method with mass detection was developed and validated for the rapid quantification of flupirtine (CAS 75507-68-5) in rat plasma samples. The chromatographic separation was achieved with a reverse phase column (4.6 x 50 mm, 5 mu m) and the mobile phase consisted of cyanomethane and 5 mM ammonium formate buffer pH 4.5 (70:30 v/v) as eluent, at a flow rate of 0.6 mL/min. Labetalol (CAS 36894-69-6) was used as an internal standard. The effluence was ionized by positive electrospray ionization and measured by mass spectrometry. The retention times of fiupirtine and labetalol were found to be 2.16 and 1.66 min respectively. The calibration curve was linear (r(2) > or = 0.99) ranging from 0.98 to 1000 ng/ml and the lower limit of quantification was 0.98 ng/mL. Inter-day and Intra-day precision were lower than 5% (CV) and accuracy ranged from 90 to 110% in terms of percent accuracy. Mean extraction recovery was found to be above 86.5%. The method was successfully applied for evaluation of the pharmacokinetic profile of flupirtine in male Sprague-Dawley rats and validated for excellent selectivity, accuracy, precision, recovery and stability.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 50 条
  • [1] Development and validation of a liquid chromatography-tandem mass spectrometry method for quantification of decitabine in rat plasma
    Xu, Haiyan
    Iv, Shaoqiong
    Qiao, Mingxi
    Fu, Yao
    Jiang, Xue
    Jin, Yi
    Li, Chibing
    Yuan, Bo
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 899 : 81 - 85
  • [2] A Liquid Chromatography-Tandem Mass Spectrometry Bioanalytical Method Development and Validation for the Determination of Pacritinib in Plasma
    Kumar, Mulamraju Aruna
    Bangaraiah, Pagala
    SEPARATION SCIENCE PLUS, 2024, 7 (12):
  • [3] Bioanalytical liquid chromatography-tandem mass spectrometry method development and validation for quantification of lurasidone in rat plasma using ion pairing agent
    Rajadhyaksha, Madhura
    Londhe, Vaishali
    BIOMEDICAL CHROMATOGRAPHY, 2024, 38 (01)
  • [4] Bioanalytical method development and validation of a liquid chromatography-tandem mass spectrometry method for determination of β-lapachone in human plasma
    Putnam, William C.
    Kallem, Raja Reddy
    Subramaniyan, Indhumathy
    Beg, M. Shaalan
    Edpuganti, Vindhya
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2020, 188
  • [5] Liquid Chromatography-tandem Mass Spectrometry for Quantification of Dioscin in Rat Plasma
    Kong, Tae Yeon
    Ji, Hye Young
    Choi, Sang-Zin
    Son, Miwon
    Lee, Hye Suk
    MASS SPECTROMETRY LETTERS, 2013, 4 (03) : 55 - 58
  • [6] Development and validation of an ultra performance liquid chromatography-tandem mass spectrometry method for the quantification of daptomycin in human plasma
    Bazoti, Fotini N.
    Gikas, Evangelos
    Skoutelis, Athanasios
    Tsarbopoulos, Anthony
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 56 (01) : 78 - 85
  • [7] Development of a liquid chromatography-tandem mass spectrometry method for assaying cenobamate in rat plasma
    Oh, Ji-Hoon
    Jeong, Jong-Woo
    Ji, Yu-Geun
    Shin, Yu-Mi
    Lee, Kyeong-Ryoon
    Cho, Kwan Hyung
    Koo, Tae-Sung
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2018, 41 (17-18) : 992 - 997
  • [8] Development and Validation of Liquid Chromatography-Tandem Mass Spectrometric Method for the Quantification of Diclofenac in Human Plasma
    Grover, Parul
    Bhardwaj, Monika
    Mehta, L.
    Kumar, Sandeep
    INDIAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2023, 85 (03) : 740 - 752
  • [9] Development and Validation of Liquid Chromatography-Tandem Mass Spectrometry Method for Quantification of Plasma Metanephrines for Differential Diagnosis of Adrenal Incidentaloma
    Lee, Sae-Mi
    Lee, Mi-Na
    Oh, Hyeon Ju
    Cho, Yoon Young
    Kim, Jae Hyeon
    Woo, Hye In
    Park, Hyung-Doo
    Lee, Soo-Youn
    ANNALS OF LABORATORY MEDICINE, 2015, 35 (05) : 519 - 522
  • [10] Determination of typhaneoside in rat plasma by liquid chromatography-tandem mass spectrometry
    Chen, Peidong
    Liu, Shijia
    Dai, Guoliang
    Xie, Liyan
    Xu, Jie
    Zhou, Ling
    Ju, Wenzheng
    Ding, Anwei
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2012, 70 : 636 - 639