Whole-body diffusion-weighted imaging: technical improvement and preliminary results

被引:41
作者
Li, Shuo
Sun, Fei
Jin, Zheng-yu
Xue, Hua-dan
Li, Ming-li
机构
[1] Beijing Union Med Coll Hosp, Dept Radiol, Chinese Acad Med Sci, Beijing 100730, Peoples R China
[2] GE Healthcare, MR Modal, Beijing, Peoples R China
关键词
diffusion magnetic resonance imaging; whole-body; imaging; STIR; tumor imaging;
D O I
10.1002/jmri.21074
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To optimize the free-breathing whole body diffusion-weighted imaging (WB-DWI) protocol by using the, short TI inversion-recovery diffusion-weighted echo-planar imaging (STIR-DWEPI) sequence and the built-in body coil. Additionally, to evaluate the feasibility of tumor screening using high-resolution three-dimensional, (3D) maximum intensity projection (MIP) images. Materials and Methods: The prescan procedure of STIR DWEPI was modified using the data,from 30 volunteers. During each exam, an optimized center frequency (CF) was used to minimize the slice offsets in. consecutive scan stations. Prescan time was reduced,from. 50 seconds to 20 seconds with improved station profile. Total scan time was 30 minutes for five stations and 1.2 in coverage. A total of 30 patients with histologically-proven malignant, disease were scanned under the final protocol using a built-in body coil. The image quality and the degree of background body signal suppression were assessed. Results: Free-breathing WB-DWI was 100% successfully performed in all patients, without slice misregistration, fat contamination, significant distortion, or nominiformity The reconstructed 3D-MIP images were adequate to depict malignant lesions in all 30 patients. The results of WB-DWI were found to be comparable to those of single-photon emission computed tomography,, (SPECT), and positron emission tomography (PET) Conclusion: Stable and high resolution WB-DWI is feasible using the technical improvements described in this study. WB-DWI might have important clinical value for the detection of primary and metastatic, malignancies within the wholebody The potential for diagnosis and therapeutic assessment of tumors should be further assessed in a larger patient cohort.
引用
收藏
页码:1139 / 1144
页数:6
相关论文
共 14 条
[1]  
Chan YL, 1997, CHINESE MED J-PEKING, V110, P485
[2]   Diffusion-weighted MRI of soft tissue tumours [J].
Einarsdóttir, H ;
Karlsson, M ;
Wejde, J ;
Bauer, HCF .
EUROPEAN RADIOLOGY, 2004, 14 (06) :959-963
[3]   Differentiation of clinically benign and malignant breast lesions using diffusion-weighted imaging [J].
Guo, Y ;
Cai, YQ ;
Cai, ZL ;
Gao, YG ;
An, NY ;
Ma, L ;
Mahankali, S ;
Gao, JH .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (02) :172-178
[4]   Whole-body MRI for staging patients with head and neck squamous cell carcinoma [J].
Herborn, CU ;
Unkel, C ;
Vogt, FM ;
Massing, S ;
Lauenstein, TC ;
Neumann, A .
ACTA OTO-LARYNGOLOGICA, 2005, 125 (11) :1224-1229
[5]   Diffusion-weighted single-shot echoplanar MR imaging for liver disease [J].
Kim, T ;
Murakami, T ;
Takahashi, S ;
Hori, M ;
Tsuda, K ;
Nakamura, H .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (02) :393-398
[6]   Comparison of whole-body STIR-MRI and 99mTc-methylene-diphosphonate scintigraphy in children with suspected multifocal bone lesions [J].
Mentzel, HJ ;
Kentouche, K ;
Sauner, D ;
Fleischmann, C ;
Vogt, S ;
Gottschild, D ;
Zintl, F ;
Kaiser, WA .
EUROPEAN RADIOLOGY, 2004, 14 (12) :2297-2302
[7]  
MURO I, 2006, P 14 ANN M ISMRM SEA
[8]  
Nakanishi Katsuyuki, 2005, Magn Reson Med Sci, V4, P11, DOI 10.2463/mrms.4.11
[9]  
SANDRA R, 2006, EUR RADIOL, V16, P1054
[10]  
Schmidt GR, 2004, RADIOLOGE, V44, P1079, DOI 10.1007/s00117-004-1128-3