Digital chest radiography with a solid-state flat-panel X-ray detector Contrast-detail evaluation with processed images printed on film hard copy

被引:42
|
作者
Chotas, HG [1 ]
Ravin, CE [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Thorac Imaging Res Div, Durham, NC 27710 USA
关键词
experimental study; lung; radiography; digital; flat panel; thorax;
D O I
10.1148/radiology.218.3.r01mr27679
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate and compare human observer performance in a contrast-detail test by using postprocessed hard-copy images from a digital chest radiography system and conventional screen-film radiographs. MATERIALS AND METHODS: The digital radiography system is based on a large- area flat-panel x-ray detector with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Images of a contrast-detail phantom were acquired at two exposure levels by using two standard thoracic screen-film systems and the digital system at matched dose. By using images of the phantom processed with standard chest image postprocessing techniques, a four-alternative forced-choice observer perception study was performed, and the number of detectable test signals (disk-shaped objects 0.3-4.0 mm in diameter) was determined for each image type. RESULTS: On average, observers detected more test signals on digital images than on screen-film radiographs at all diameters up to 2.0 mm and an equivalent number at larger diameters. Test signals with lower inherent subject contrast were detected more readily on digital images than on screen-film images, even when x-ray exposure levels for the digital system were reduced by 20%. CONCLUSION: Observer performance in a contrast-detail detection task can be improved by using images acquired with the flat-panel digital chest radiography system as compared with those acquired with state-of-the-art screen-film combinations.
引用
收藏
页码:679 / 682
页数:4
相关论文
共 39 条
  • [1] Flat-panel digital mammography system: Contrast-detail comparison between screen-film radiographs and hard-copy images
    Suryanarayanan, S
    Karellas, A
    Vedantham, S
    Ved, H
    Baker, SP
    D'Orsi, CJ
    RADIOLOGY, 2002, 225 (03) : 801 - 807
  • [2] Contrast-detail phantom study for X-ray spectrum optimization regarding chest radiography using a cesium iodide -: Amorphous silicon flat-panel detector
    Hamer, OW
    Völk, M
    Zorger, N
    Borisch, I
    Büittner, R
    Feuerbach, S
    Strotzer, M
    INVESTIGATIVE RADIOLOGY, 2004, 39 (10) : 610 - 618
  • [3] High resolution flat-panel X-ray detector for digital radiography
    Park, JK
    Kang, SS
    Cha, BY
    Kim, JY
    Kim, SY
    Choi, HK
    Nam, SH
    Healthcom 2005: 7th International Workshop on Enterprise Networking and Computing in Healthcare Industry, Proceedings, 2005, : 359 - 363
  • [4] Experimental evaluation of a-Se flat-panel X-ray detector for digital radiography
    Kim, JY
    Park, JK
    Choi, JY
    Cha, BY
    Kang, SS
    An, SH
    Nam, SH
    Choi, HK
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4048 - 4052
  • [5] Experimental evaluation of a-Se flat-panel X-ray detector for digital radiography
    Choi, JY
    Park, JK
    Lee, DG
    Kang, SS
    Nam, SH
    NEW APPLICATIONS FOR WIDE-BANDGAP SEMICONDUCTORS, 2003, 764 : 265 - 270
  • [6] Improved contrast-detail detection with flat-panel digital radiography: Comparison with computed radiography and a conventional film-screen system
    Erb, G
    Aretz, K
    Hahmann, M
    Radeleff, BA
    Hansmann, J
    Kauffmann, GW
    RADIOLOGY, 2000, 217 : 164 - 164
  • [7] Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems - A contrast-detail phantom study
    Rong, XJJ
    Shaw, CC
    Liu, XM
    Lemacks, MR
    Thompson, SK
    MEDICAL PHYSICS, 2001, 28 (11) : 2328 - 2335
  • [8] Experimental evaluation of a novel CdZnTe flat-panel X-ray detector for digital radiography and fluoroscopy
    Tokuda, S
    Adachi, S
    Sato, T
    Yoshimuta, T
    Nagata, H
    Uehara, K
    Izumi, Y
    Teranuma, O
    Yamada, S
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 140 - 147
  • [9] Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography
    Svenja P. Hennigs
    Marietta Garmer
    Horst J. Jaeger
    Reinhard Classen
    Andreas Jacobs
    Hans M. Gissler
    Andreas Christmann
    Klaus Mathias
    European Radiology, 2001, 11 : 1688 - 1696
  • [10] Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography
    Hennigs, SP
    Garmer, M
    Jaeger, HJ
    Classen, R
    Jacobs, A
    Gissler, HM
    Christmann, A
    Mathias, K
    EUROPEAN RADIOLOGY, 2001, 11 (09) : 1688 - 1696