Automatic Implantable Cardioverter Defibrillator in Electroconvulsive Therapy

被引:0
作者
Bryson, Ethan O. [1 ]
Popeo, Dennis M. [2 ]
Briggs, Mimi C. [2 ]
Pasculli, Rosa M. [2 ]
Kellner, Charles H. [2 ]
机构
[1] Mt Sinai Sch Med, Dept Anesthesiol, New York, NY USA
[2] Mt Sinai Sch Med, Dept Psychiat, New York, NY USA
关键词
ECT; implanted defibrillator; deactivation;
D O I
10.1097/YCT.0000000000000173
中图分类号
B84 [心理学]; C [社会科学总论]; Q98 [人类学];
学科分类号
03 ; 0303 ; 030303 ; 04 ; 0402 ;
摘要
As the number of patients with implantable cardiac devices increases so too does the frequency with which these individuals present for electroconvulsive therapy (ECT). The rationale for deactivating an automatic implantable cardioverter defibrillator before ECT has been made based on the concern that artifacts generated during treatment could be interpreted as a treatable rhythm by the internal device, resulting in a discharge. We believe that the risk of inappropriate discharge during ECT is very low and outweighed by the considerable benefit of an active device being able to more quickly treat a malignant dysrhythmia.
引用
收藏
页码:E22 / E22
页数:1
相关论文
共 1 条
[1]   Cardiac Rhythm Management Devices and Electroconvulsive Therapy A Critical Review Apropos of a Depressed Patient With a Pacemaker [J].
Kokras, Nikolaos ;
Politis, Antonios M. ;
Zervas, Iannis M. ;
Pappa, Dimitra ;
Markatou, Maria ;
Katirtzoglou, Evgenia ;
Papadimitriou, George N. .
JOURNAL OF ECT, 2011, 27 (03) :214-220