RF atomic magnetometer array with over 40 dB interference suppression using electron spin resonance

被引:11
作者
Cooper, Robert J. [1 ]
Prescott, David W. [1 ]
Lee, Garrett J. [1 ]
Sauer, Karen L. [1 ]
机构
[1] George Mason Univ, Dept Phys & Astron, Fairfax, VA 22030 USA
基金
美国国家科学基金会;
关键词
Electron spin resonance; Atomic magnetometer array; RF magnetic field mapping; Unshielded detection; Interference rejection; Low-field magnetic resonance; NUCLEAR-QUADRUPOLE RESONANCE; MAGNETOCARDIOGRAPHY MEASUREMENTS; MAGNETOENCEPHALOGRAPHY; FIELD; EXPLOSIVES; MRI;
D O I
10.1016/j.jmr.2018.08.007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An unshielded array of Rb-87 atomic magnetometers, operating close to 1 MHz, is used to attenuate interference by 42-48 dB. A sensitivity of 15 fT/root Hz to a local source of signal is retained. In addition, a 2D spectroscopic technique, in which the magnetometers are repeatedly pumped and data acquired between pump times, enables a synchronously generated signal to be distinguished from an interfering signal very close in frequency; the timing and signal mimics what would be observed in a magnetic resonance echo train. Combining the interference rejection and the 2D spectroscopy techniques, a 100 ft local signal is differentiated from a 20 pT interference signal operating only 1 Hz away. A phase-encoded reference signal is used to calibrate the magnetometers in real time in the presence of interference. Key to the strong interference rejection is the accurate calibration of the reference signal across the array, obtained through electron spin resonance measurements. This calibration is found to be sensitive to atomic polarization, RF pulse duration, and direction of the excitation. The experimental parameters required for an accurate and robust calibration are discussed. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:36 / 46
页数:11
相关论文
共 60 条
[1]  
Abragam A., 1983, The principles of nuclear magnetism, V32
[2]  
Alem O., 2011, THESIS
[3]   Magnetic field imaging with microfabricated optically-pumped magnetometers [J].
Alem, Orang ;
Mhaskar, Rahul ;
Jimenez-Martinez, Ricardo ;
Sheng, Dong ;
LeBlanc, John ;
Trahms, Lutz ;
Sander, Tilmann ;
Kitching, John ;
Knappe, Svenja .
OPTICS EXPRESS, 2017, 25 (07) :7849-7858
[4]   Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers [J].
Alem, Orang ;
Sander, Tilmann H. ;
Mhaskar, Rahul ;
LeBlanc, John ;
Eswaran, Hari ;
Steinhoff, Uwe ;
Okada, Yoshio ;
Kitching, John ;
Trahms, Lutz ;
Knappe, Svenja .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (12) :4797-4811
[5]   High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation [J].
Allred, JC ;
Lyman, RN ;
Kornack, TW ;
Romalis, MV .
PHYSICAL REVIEW LETTERS, 2002, 89 (13) :130801-130801
[6]   Theory of spin-exchange optical pumping of 3He and 129Xe [J].
Appelt, S ;
Baranga, AB ;
Erickson, CJ ;
Romalis, MV ;
Young, AR ;
Happer, W .
PHYSICAL REVIEW A, 1998, 58 (02) :1412-1439
[7]   The Emerging Field of Medicines Authentication by Nuclear Quadrupole Resonance Spectroscopy [J].
Barras, J. ;
Althoefer, K. ;
Rowe, M. D. ;
Poplett, I. J. ;
Smith, J. A. S. .
APPLIED MAGNETIC RESONANCE, 2012, 43 (04) :511-529
[8]   Cesium coherent population trapping magnetometer for cardiosignal detection in an unshielded environment [J].
Belfi, J. ;
Bevilacqua, G. ;
Biancalana, V. ;
Cartaleva, S. ;
Dancheva, Y. ;
Moi, L. .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2007, 24 (09) :2357-2362
[9]   All-optical magnetometry for NMR detection in a micro-Tesla field and unshielded environment [J].
Bevilacqua, G. ;
Biancalana, V. ;
Dancheva, Y. ;
Moi, L. .
JOURNAL OF MAGNETIC RESONANCE, 2009, 201 (02) :222-229
[10]  
Bevilacqua G., 2016, APPL PHYS B, V122, P1