Radiation dose in 320-detector-row CT coronary angiography: Prospective ECG triggering combined with multi-segment reconstruction

被引:1
作者
Yang, Ching-Ching [2 ]
Wu, Jay [3 ]
Tsai, Chia-Jung [1 ]
Chen, Chien-Chuan [1 ,4 ]
Hung, Chien-Fu [4 ]
Wu, Tung-Hsin [1 ]
机构
[1] Natl Yang Ming Univ, Dept Biomed Imaging & Radiol Sci, Taipei 112, Taiwan
[2] Tzu Chi Coll Technol, Dept Radiol Technol, Hualien, Taiwan
[3] China Med Univ, Dept Biomed Imaging & Radiol Sci, Taichung, Taiwan
[4] Chang Gung Mem Hosp Linkou, Dept Med Imaging & Intervent, Tao Yuan 333, Taiwan
关键词
320-Detector-row coronary CT angiography; Radiation dose; ECG pulsing window width; Heart rate; CONE-BEAM CT; COMPUTED-TOMOGRAPHY; DIAGNOSTIC-ACCURACY; IMAGE QUALITY; EXPOSURE; SCANNER; ROW; MM;
D O I
10.1016/j.radmeas.2011.07.003
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Background: The aim of this study was to investigate patient doses in prospective electrocardiogram (ECG)-triggered CT coronary angiography (CTCA) combined with multi-segment reconstruction on a 320-detector-row CT. Methods: CTCA data acquired with prospective ECG (pECG) triggering at 0-100% (pECG(100%)). 30-80% (pECG(50%)), 70-80% (pECG(10%)) of the R R interval and reconstructed using mono-, two- and three-segment reconstruction were investigated. Effective doses were estimated by using LiF-TLDs placed at several organ sites in an Alderson-Rando phantom. Results: With pECG(100%). the estimation of effective dose of data reconstructed using mono-segment (pECG(100%_1S)) reconstruction was 10.01 +/- 0.56 mSv. For data acquired using pECG(50%), the effective doses were 6.16 +/- 0.12, 9.92 +/- 0.37 and 13.51 +/- 0.17 mSv in mono-segment (pECG(50%_1S)). two-segment (pECG(50%_2S)) and three-segment (pECG(50%_3S)) reconstruction, respectively. The effective dose of data acquired with pECG(10%) and reconstructed using mono-segment (pECG(10%_1S)) reconstruction was 3.61 +/- 0.07 mSv. We observed a difference of around 7.46% between effective doses estimated using TLD-phantom measurement and CT dose index (CTDI) obtained from the scanner. Conclusion: For patients with low and intermediate heart rate, radiation exposure could be reduced by 38.6% or more by narrowing pulsing window width. Although slightly higher radiation dose was observed in multi-segment reconstruction, this method can be used in high heart rate patients to provide data of high temporal resolution without increasing radiation exposure when it is combined with prospective ECG triggering. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2060 / 2064
页数:5
相关论文
共 15 条
  • [1] Diagnostic accuracy of 320-row multidetector computed tomography coronary angiography in the non-invasive evaluation of significant coronary artery disease
    de Graaf, Fleur R.
    Schuijf, Joanne D.
    van Velzen, Joella E.
    Kroft, Lucia J.
    de Roos, Albert
    Reiber, Johannes H. C.
    Boersma, Eric
    Schalij, Martin J.
    Spano, Fabrizio
    Jukema, J. Wouter
    van der Wall, Ernst E.
    Bax, Jeroen J.
    [J]. EUROPEAN HEART JOURNAL, 2010, 31 (15) : 1908 - 1915
  • [2] Noninvasive Coronary Angiography by 320-Row Computed Tomography With Lower Radiation Exposure and Maintained Diagnostic Accuracy Comparison of Results With Cardiac Catheterization in a Head-to-Head Pilot Investigation
    Dewey, Marc
    Zimmermann, Elke
    Deissenrieder, Florian
    Laule, Michael
    Duebel, Hans-Peter
    Schlattmann, Peter
    Knebel, Fabian
    Rutsch, Wolfgang
    Hamm, Bernd
    [J]. CIRCULATION, 2009, 120 (10) : 867 - 875
  • [3] Dose Exposure of Patients Undergoing Comprehensive Stroke Imaging by Multidetector-Row CT: Comparison of 320-Detector Row and 64-Detector Row CT Scanners
    Diekmann, S.
    Siebert, E.
    Juran, R.
    Roll, M.
    Deeg, W.
    Bauknecht, H. -C.
    Diekmann, F.
    Klingebiel, R.
    Bohner, G.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (06) : 1003 - 1009
  • [4] Chernobyl: consequences of the catastrophe for people and the environment
    Elojeimy, Saeed
    Tipnis, Sameer
    Huda, Walter
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (01) : 43 - 49
  • [5] Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner
    Geleijns, J.
    Artells, M. Salvado
    Bruin, P. Wde
    Matter, R.
    Muramatsu, Y.
    McNitt-Gray, M. F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (10) : 3141 - 3159
  • [6] Prospective ECG-triggered sequential scan protocol for coronary dual-source CT angiography: initial experience
    Hein, Franziska
    Meyer, Tanja
    Hadamitzky, Martin
    Bischoff, Bernhard
    Will, Albrecht
    Hendrich, Eva
    Martinoff, Stefan
    Schoemig, Albert
    Hausleiter, Joerg
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2009, 25 : 231 - 239
  • [7] Does two-segment image reconstruction at 64-section CT coronary angiography improve image quality and diagnostic accuracy?
    Herzog, Christopher
    Nguyen, Shaun A.
    Savino, Giancarlo
    Zwerner, Peter L.
    Doll, Josh
    Nielsen, Christopher D.
    Flohr, Thomas G.
    Vogl, Thomas J.
    Costello, Philip
    Schoepf, U. Joseph
    [J]. RADIOLOGY, 2007, 244 (01) : 121 - 129
  • [8] Computing effective dose in cardiac CT
    Huda, Walter
    Tipnis, Sameer
    Sterzik, Alexander
    Schoepf, U. Joseph
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (13) : 3675 - 3684
  • [9] *ICRP, 2007, ANN ICRP, V37, P62
  • [10] Mono- versus bisegment reconstruction algorithms for dual-source computed tomography coronary angiography
    Leschka, Sebastian
    Alkadhi, Hatem
    Stolzmann, Paul
    Schmid, Florian T.
    Leschka, Simon C.
    Scheffel, Hans
    Stinn, Bjoern
    Flohr, Thomas G.
    Marincek, Borut
    Wildermuth, Simon
    [J]. INVESTIGATIVE RADIOLOGY, 2008, 43 (10) : 703 - 711