Influence of sharpening filters on the detection of root fractures using low-dose cone-beam computed tomography

被引:5
作者
Gonzalez, Alexandra Robles [1 ]
Tosoni, Guilherme Monteiro [2 ]
Freitas, Deborah Queiroz [1 ]
Oliveira, Matheus L. [1 ]
机构
[1] Univ Estadual Campinas, Piracicaba Dent Sch FOP UNICAMP, Dept Oral Diag, Div Oral Radiol, Av Limeira 901, BR-13414903 Piracicaba, SP, Brazil
[2] Sao Paulo State Univ FOAr UNESP, Araraquara Dent Sch, Dept Oral Diag & Surg, Div Oral Radiol, Araraquara, SP, Brazil
关键词
Cone-beam computed tomography; Diagnostic imaging; Tooth fractures; Radiographic image enhancement; TUBE CURRENT; DIAGNOSIS; RADIOGRAPHY; ENHANCEMENT; TEETH;
D O I
10.1007/s00784-022-04444-7
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives To evaluate the influence of sharpening filters in the detection of root fractures using low-dose cone-beam computed tomography (CBCT). Materials and methods Eighty-four CBCT volumes acquired at three mA levels of 28 teeth inserted in the dental socket of dry human skull were selected from a previous study. The teeth were divided into four groups according to the presence and absence of root fracture and endodontic filling. Five radiologists evaluated all CBCT volumes for the presence of root fracture with and without the application of "Sharpen 1x" and "Sharpen 2x" filters in OnDemand3D software. Area under the ROC curve (AUC), sensitivity, specificity, and inter- and intra-observer concordance were calculated and compared (alpha = 0.05). Results Sharpening filters did not lead to significant differences in AUC, sensitivity, and specificity at the three mA levels tested (p > 0.05), regardless of the presence of endodontic filling (p > 0.05). However, the significant reduction of AUC observed in CBCT volumes at 4 mA without filter (p < 0.05) ceased to exist after the application of filters (p > 0.05). Sensitivity and specificity ranged from low and moderate. Conclusions The use of sharpening filters can be recommended in CBCT volumes at 4 mA for root fracture detection for leading to the same performance as those at 6.3 and 10 mA. The presence of endodontic filling material did not influence the action of filters in the diagnosis of root fracture.
引用
收藏
页码:4797 / 4803
页数:7
相关论文
共 32 条
[1]   Comparison of intraoral radiography and cone-beam computed tomography for the detection of horizontal root fractures: an in vitro study [J].
Avsever, Hakan ;
Gunduz, Kaan ;
Orhan, Kaan ;
Uzun, Ismail ;
Ozmen, Bilal ;
Egrioglu, Erol ;
Midilli, Muhammed .
CLINICAL ORAL INVESTIGATIONS, 2014, 18 (01) :285-292
[2]   A comparison of cone beam computed tomography and periapical radiography for the detection of vertical root fractures in nonendodontically treated teeth [J].
Brady, E. ;
Mannocci, F. ;
Brown, J. ;
Wilson, R. ;
Patel, S. .
INTERNATIONAL ENDODONTIC JOURNAL, 2014, 47 (08) :735-746
[3]   Cone beam CT optimisation for detection of vertical root fracture with metal in the field of view or the exomass [J].
Candemil, Amanda P. ;
Salmon, Benjamin ;
Vasconcelos, Karla F. ;
Oenning, Anne C. ;
Jacobs, Reinhilde ;
Freitas, Deborah Q. ;
Haiter-Neto, Francisco ;
Mangione, Francesca ;
Oliveira, Matheus L. .
SCIENTIFIC REPORTS, 2021, 11 (01)
[4]   Cone-beam Computed Tomography for Detecting Vertical Root Fractures in Endodontically Treated Teeth: A Systematic Review [J].
Chang, Edwin ;
Lam, Ernest ;
Shah, Prakesh ;
Azarpazhooh, Amir .
JOURNAL OF ENDODONTICS, 2016, 42 (02) :177-185
[5]   Image enhancement of digital periapical radiographs according to diagnostic tasks [J].
Choi, Jin-Woo ;
Han, Won-Jeong ;
Kim, Eun-Kyung .
IMAGING SCIENCE IN DENTISTRY, 2014, 44 (01) :31-35
[6]   Influence of Cone-Beam Computed Tomography Enhancement Filters on Diagnosis of Simulated External Root Resorption [J].
de Azevedo Vaz, Sergio Lins ;
Vasconcelos, Taruska Ventorini ;
Neves, Frederico Sampaio ;
de Freitas, Deborah Queiroz ;
Haiter-Neto, Francisco .
JOURNAL OF ENDODONTICS, 2012, 38 (03) :305-308
[7]   Diagnosis of Mesiodistal Vertical Root Fractures in Teeth with Metal Posts: Influence of Applying Filters in Cone-beam Computed Tomography Images at Different Resolutions [J].
De Martin e Silva, Debora ;
Campos, Celso Neiva ;
Pires Carvalho, Antonio Carlos ;
Devito, Karina Lopes .
JOURNAL OF ENDODONTICS, 2018, 44 (03) :470-474
[8]  
de Sousa Emerson-Tavares, 2017, J Clin Exp Dent, V9, pe516, DOI 10.4317/jced.53496
[9]   Accuracy of High-resolution Small-volume Cone-Beam Computed Tomography in the Diagnosis of Vertical Root Fracture: An In Vivo Analysis [J].
Dias, Debora R. ;
Iwaki, Lilian C., V ;
de Oliveira, Alexandre C. A. ;
Martinhao, Felipe S. ;
Rossi, Robson M. ;
Araujo, Mauricio G. ;
Hayacibara, Roberto M. .
JOURNAL OF ENDODONTICS, 2020, 46 (08) :1059-1066
[10]   Detection of Simulated Vertical Root Fractures: Which Cone-beam Computed Tomographic System Is the Most Accurate? [J].
Elsaltani, Mohamed Hussein ;
Farid, Mary Medhat ;
Ashmawy, Mostafa Saad Eldin .
JOURNAL OF ENDODONTICS, 2016, 42 (06) :972-977