Critical Reliability Issues of Common Type Alcohol-Based Handrub Dispensers

被引:14
作者
Bansaghi, Szava [1 ]
Soule, Herve [2 ,3 ]
Guitart, Chloe [2 ,3 ]
Pittet, Didier [2 ,3 ]
Haidegger, Tanis [4 ,5 ]
机构
[1] Semmelweis Univ, Dept Epidemiol, Budapest, Hungary
[2] Univ Geneva Hosp, Geneva, Switzerland
[3] Fac Med HUG, Geneva, Switzerland
[4] Obuda Univ, Univ Res & Innovat Ctr EKIK, Budapest, Hungary
[5] Austrian Ctr Med Innovat & Technol ACMIT, Wiener Neustadt, Austria
关键词
Hand hygiene; Handrub volume; Dispenser quality; Alcohol-based handrub; Handrub dosing; Patient safety standards; Hand sanitizer;
D O I
10.1186/s13756-020-00735-4
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background Hand hygiene can only be efficient if the whole hand surface is treated with sufficient alcohol-based handrub (ABHR); therefore, the volume of handrub applied is a critical factor in patient safety. The proper amount of ABHR should be provided by handrub dispensers. The aim of this study was to investigate the dispensing performance of wall-mounted ABHR dispensers commonly employed in hospital settings. Method In a multicenter study, we tested 46 dispensers (22 in laboratory and 24 in clinical environments), measuring dispensed ABHR volume during continuous use and after a period of non-use. The influence of the pumping mechanism, liquid level, ABHR formats, handrub composition, temperature, and atmospheric pressure was investigated. Results A total of 7 out of the 22 investigated dispensers (32%) lost a significant amount of handrub; greater than 30% of the nominal volume after 8 h of non-use, thus frequently dispensing suboptimal volume, as measured in laboratory settings. Key influencing factors were found to be handrub format (gel or liquid), handrub level in the container and type of dispenser. When gel ABHR was used, after 4 h of non-use of the dispensers, the volume of the dispensed amount of ABHR insignificantly changed (97% of the original amount), while it technically decreased to zero in the case of liquid ABHR (1% of the original amount). The liquid level had a medium effect on the dispensed volume in each investigated case; the magnitude of this effect varied widely depending on the dispensing mechanism. When dispensers were in continuous use, they dispensed a cumulated 3 mL of ABHR from two consecutive pushes, while when they were not in use for 1 h, up to 4 consecutive pushes were necessary to provide a total of 3 mL ABHR. Design and production quality were also identified as important contributing factors with respect to the volume dispensed. Data collected in clinical settings confirmed these findings, for multiple types of dispensers. Conclusion All ABHR dispensers should be regularly audited to control the reference volume distributed, with particular attention paid to regular mechanical pump units filled with liquid handrub.
引用
收藏
页数:13
相关论文
共 12 条
[1]   Recommendations and requirements for soap and hand rub dispensers in healthcare facilities [J].
Assadian, Ojan ;
Kramer, Axel ;
Christiansen, Baerbel ;
Exner, Martin ;
Martiny, Heike ;
Sorger, Arno ;
Suchomel, Miranda .
GMS HYGIENE AND INFECTION CONTROL, 2012, 7 (01)
[2]  
ASTM International, 2013, ASTM E1174-13
[3]   Should Alcohol-Based Handrub Use Be Customized to Healthcare Workers' Hand Size? [J].
Bellissimo-Rodrigues, Fernando ;
Soule, Herve ;
Gayet-Ageron, Angele ;
Martin, Yves ;
Pittet, Didier .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2016, 37 (02) :219-221
[4]  
European Committee for Standardization (CEN), 2013, 1500 CEN
[5]   Less and less-influence of volume on hand coverage and bactericidal efficacy in hand disinfection [J].
Kampf, Guenter ;
Ruselack, Sigunde ;
Eggerstedt, Sven ;
Nowak, Nicolas ;
Bashir, Muhammad .
BMC INFECTIOUS DISEASES, 2013, 13
[6]   The importance of evaluating product dispensers when selecting alcohol-based handrubs [J].
Kohan, C ;
Ligi, C ;
Dumigan, DG ;
Boyce, JM .
AMERICAN JOURNAL OF INFECTION CONTROL, 2002, 30 (06) :373-375
[7]   Hand hygiene technique assessment using electronic equipment in 26 Hungarian healthcare institutes [J].
Lehotsky Akos ;
Morvai Julia ;
Szilagyi Laszlo ;
Bansaghi Szava ;
Benko Aliz ;
Haidegger Tamas .
ORVOSI HETILAP, 2017, 158 (29) :1143-1148
[8]   Back to the future: rising to the Semmelweis challenge in hand hygiene [J].
Stewardson, Andrew ;
Allegranzi, Benedetta ;
Sax, Hugo ;
Kilpatrick, Claire ;
Pittet, Didier .
FUTURE MICROBIOLOGY, 2011, 6 (08) :855-876
[9]   A large-scale assessment of hand hygiene quality and the effectiveness of the "WHO 6-steps" [J].
Szilagyi, Laszlo ;
Haidegger, Tamas ;
Lehotsky, Akos ;
Nagy, Melinda ;
Csonka, Erik-Artur ;
Sun, Xiuying ;
Ooi, Kooi Li ;
Fisher, Dale .
BMC INFECTIOUS DISEASES, 2013, 13
[10]   Hand hygiene in hospitals: anatomy of a revolution [J].
Vermeil, T. ;
Peters, A. ;
Kilpatrick, C. ;
Pires, D. ;
Allegranzi, B. ;
Pittet, D. .
JOURNAL OF HOSPITAL INFECTION, 2019, 101 (04) :383-392