Quantitative Comparison of Tandem Mass Spectra Obtained on Various Instruments

被引:33
作者
Bazso, Fanni Laura [1 ]
Ozohanics, Oliver [2 ]
Schlosser, Gitta [3 ]
Ludanyi, Krisztina [4 ]
Vekey, Karoly [1 ,5 ]
Drahos, Laszlo [1 ,2 ]
机构
[1] Hungarian Acad Sci, MS Prote Res Grp, Res Ctr Nat Sci, Magyar Tudsok Krt 2, H-1117 Budapest, Hungary
[2] Hungarian Acad Sci, MTA TTK NAP MS Neuroprote Res Grp B, Res Ctr Nat Sci, Magyar Tudsok Krt 2, H-1117 Budapest, Hungary
[3] Eotvos Lorand Univ, Hungarian Acad Sci, MTA ELTE Res Grp Peptide Chem, H-1117 Budapest, Hungary
[4] Semmelweis Univ, Dept Pharmaceut, Hogyes E St 7-9, H-1092 Budapest, Hungary
[5] Hungarian Acad Sci, Core Technol Ctr, Res Ctr Nat Sci, Magyar Tudosok Krt 2, H-1117 Budapest, Hungary
关键词
Tandem mass spectrometry; Similarity; Instrument comparison; SPECTROMETRY; IDENTIFICATION; SIMILARITY; ALGORITHMS; LIBRARIES; COMPOUND; IONS;
D O I
10.1007/s13361-016-1408-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The similarity between two tandem mass spectra, which were measured on different instruments, was compared quantitatively using the similarity index (SI), defined as the dot product of the square root of peak intensities in the respective spectra. This function was found to be useful for comparing energy-dependent tandem mass spectra obtained on various instruments. Spectral comparisons show the similarity index in a 2D "heat map", indicating which collision energy combinations result in similar spectra, and how good this agreement is. The results and methodology can be used in the pharma industry to design experiments and equipment well suited for good reproducibility. We suggest that to get good long-term reproducibility, it is best to adjust the collision energy to yield a spectrum very similar to a reference spectrum. It is likely to yield better results than using the same tuning file, which, for example, does not take into account that contamination of the ion source due to extended use may influence instrument tuning. The methodology may be used to characterize energy dependence on various instrument types, to optimize instrumentation, and to study the influence or correlation between various experimental parameters.
引用
收藏
页码:1357 / 1365
页数:9
相关论文
共 24 条
[1]  
Collette C, 1998, RAPID COMMUN MASS SP, V12, P1673, DOI 10.1002/(SICI)1097-0231(19981130)12:22<1673::AID-RCM385>3.0.CO
[2]  
2-A
[3]   COMPUTER METHODS IN ANALYTICAL MASS SPECTROMETRY - IDENTIFICATION OF AN UNKNOWN COMPOUND IN A CATALOG [J].
CRAWFORD, LR ;
MORRISON, JD .
ANALYTICAL CHEMISTRY, 1968, 40 (10) :1464-&
[4]   Correction of precursor and product ion relative abundances in order to standardize CID spectra and improve Ecom50 accuracy for non-targeted metabolomics [J].
Dubey, Ritvik ;
Hill, Dennis W. ;
Lai, Steven ;
Chen, Ming-Hui ;
Grant, David F. .
METABOLOMICS, 2015, 11 (03) :753-763
[5]   Extending the coverage of spectral libraries: A neighbor-based approach to predicting intensities of peptide fragmentation spectra [J].
Ji, Chao ;
Arnold, Randy J. ;
Sokoloski, Kevin J. ;
Hardy, Richard W. ;
Tang, Haixu ;
Radivojac, Predrag .
PROTEOMICS, 2013, 13 (05) :756-765
[6]   A tandem mass spectrometry-based method to assess the architectural purity of synthetic polymers: a case of a cyclic polylactide obtained by click chemistry [J].
Josse, Thomas ;
De Winter, Julien ;
Dubois, Philippe ;
Coulembier, Olivier ;
Gerbaux, Pascal ;
Memboeuf, Antony .
POLYMER CHEMISTRY, 2015, 6 (01) :64-69
[7]  
Lambert JP, 2013, NAT METHODS, V10, P1239, DOI [10.1038/NMETH.2702, 10.1038/nmeth.2702]
[8]  
Li Q., 2010, INT J PROTEOMICS, V2010, P15
[9]   Methods for peptide identification by spectral comparison [J].
Liu, Jian ;
Bell, Alexander W. ;
Bergeron, John J. M. ;
Yanofsky, Corey M. ;
Carrillo, Brian ;
Beaudrie, Christian E. H. ;
Kearney, Robert E. .
PROTEOME SCIENCE, 2007, 5 (1)
[10]   Precision proteomics: The case for high resolution and high mass accuracy [J].
Mann, Matthias ;
Kelleher, Neil L. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2008, 105 (47) :18132-18138