A strategy for a fast (ca. 20 min), specific, electrochemical immunoassay for the cardiac biomarker creatine kinase (CK) and the human cytokine interleukin 10 (IL10) has been developed in this paper. The polyaniline modified gold surface formed from electrochemical reduction of diazonium salt supplies a solid substrate to link the activated carboxylic acid groups from the antibodies, which were labelled with ferrocene. The direct electrochemistry of ferrocene allows the analysis of protein markers with good sensitivity. The creatine kinase sensor demonstrates limit of detection of 0.5 pg mL-1 in a physiological Krebs-Henseleit solution. The anti-IL10 antibody retained fluorescence activity after further coupling to ferrocene and covalent immobilization on to a gold electrode, showing a linear detection range for IL-10 from 0.001 ng?mL-1 to 50 ng?mL-1 in PBS. We attribute the high sensitivity to the well-controlled modified surface which results in endon antibodies that can specifically capture the antigen with ease.
引用
收藏
页码:264 / 272
页数:9
相关论文
共 66 条
[1]   Signal generation at an electrochemical immunosensor via the direct oxidation of an electroactive label [J].
Akram, M ;
Stuart, MC ;
Wong, DKY .
ELECTROANALYSIS, 2006, 18 (03) :237-246
[2]   Oxytocin protects rat heart against ischemia reperfusion injury via pathway involving mitochondrial ATP-dependent potassium channel [J].
Alizadeh, Ali Mohammad ;
Faghihi, Mandieh ;
Sadeghipour, Harnid Reza ;
MohammadGhasemi, Fahimeh ;
Imani, Alireza ;
Houshmand, Fariba ;
Khorid, Vahid .
PEPTIDES, 2010, 31 (07) :1341-1345
[3]   Ultrafast voltammetry for probing interfacial electron transfer in molecular wires [J].
Amatore, Christian ;
Maisonhaute, Emmanuel ;
Schollhorn, Bernd ;
Wadhawan, Jay .
CHEMPHYSCHEM, 2007, 8 (09) :1321-1329
[4]  
[Anonymous], RAPIDCARD INSTATEST
[5]   Hybrid coating on steel: ZnNi electrodeposition and surface modification with organothiols and diazonium salts [J].
Berger, Francois ;
Delhalle, Joseph ;
Mekhalif, Zineb .
ELECTROCHIMICA ACTA, 2008, 53 (06) :2852-2861
[6]  
Brett C.M.A., 1998, ELECTROANAL
[7]  
Cheah L.-T., UNPUB
[8]   Microfluidic perfusion system for maintaining viable heart tissue with real-time electrochemical monitoring of reactive oxygen species [J].
Cheah, Lih-Tyng ;
Dou, Yue-Hua ;
Seymour, Anne-Marie L. ;
Dyer, Charlotte E. ;
Haswell, Stephen J. ;
Wadhawan, Jay D. ;
Greenman, John .
LAB ON A CHIP, 2010, 10 (20) :2720-2726
[9]   Ultrasensitive magnetic biosensor for homogeneous immunoassay [J].
Chemla, YR ;
Crossman, HL ;
Poon, Y ;
McDermott, R ;
Stevens, R ;
Alper, MD ;
Clarke, J .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2000, 97 (26) :14268-14272
[10]   Controlling antibody orientation on charged self-assembled monolayers [J].
Chen, SF ;
Liu, LY ;
Zhou, J ;
Jiang, SY .
LANGMUIR, 2003, 19 (07) :2859-2864