Photoplethysmography sampling frequency: pilot assessment of how low can we go to analyze pulse rate variability with reliability?

被引:89
作者
Choi, A. [1 ]
Shin, H. [2 ]
机构
[1] Gachon Univ, Dept Software, Seongnam, South Korea
[2] Chonnam Natl Univ, Dept Biomed Engn, Yeosu, South Korea
基金
新加坡国家研究基金会;
关键词
heart rate variability; photoplethysmography; pulse rate variability; sampling frequency; HEART-RATE-VARIABILITY; SMARTPHONE;
D O I
10.1088/1361-6579/aa5efa
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Pulse rate variability (PRV) analysis appears as the first alternative to heart rate variability analysis for wearable devices; however, there is a constraint on computational load and energy consumption for the limited system resources available to the devices. Considering that adjustment of the sampling frequency is one of the strategies for reducing computational load and power consumption, this study aimed to investigate the influence of sampling frequency (f(s)) on PRV analysis and to find the minimum sampling frequency while maintaining reliability. We generated 5000, 2500, 1000, 500, 250, 100, 50, 25, 20, 15, 10, 5 Hz down-sampled photoplethysmograms from 10 kHz-sampled PPGs and derived time-and frequency-domain variables of the PRV. These included AVNN, SDNN, SDSD, RMSSD, NN50, pNN50, total power, VLF, LF, HF, LF/HF, nLF and nHF for each down-sampled signal. Derived variables were compared with heart rate variability of the 10 kHz-sampled electrocardiograms, and then statistically investigated using one-way ANOVA test and Bland-Altman analysis. As a result, significant differences (P < 0.05) were found for SDNN, SDSD, RMSSD, NN50, pNN50, TP, HF, LF/HF, nLF and nHF, but not for AVNN, VLF and LF. Based on the post hoc tests, it was found that the NN50 and pNN50, SDSD and RMSSD, LF/HF and nHF, SDNN, TP and nLF analysis had significant differences at f(s) <= 20 Hz, f(s) <= 15 Hz, f(s) <= 10 Hz; f(s) = 5 Hz, respectively. In other words, a significant difference was not seen for any variable if the f(s) was greater than 25 Hz. Consequently, our pilot study suggests that analysis of variability in the time and frequency domain from pulse rate obtained through PPG may be potentially as reliable as that derived from the analysis of the electrocardiogram, provided that f(s) >= 25 Hz sampling frequency is used.
引用
收藏
页码:586 / 600
页数:15
相关论文
共 24 条
  • [11] Assessment of pulse rate variability by the method of pulse frequency demodulation
    Hayano, Junichiro
    Barros, Allan Kardec
    Kamiya, Atsunori
    Ohte, Nobuyuki
    Yasuma, Fumihiko
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2005, 4 (1)
  • [12] What is the adequate sampling interval of the ECG signal for heart rate variability analysis in the time domain?
    Hejjel, L
    Roth, E
    [J]. PHYSIOLOGICAL MEASUREMENT, 2004, 25 (06) : 1405 - 1411
  • [13] Investigating a smartphone imaging unit for photoplethysmography
    Jonathan, E.
    Leahy, Martin
    [J]. PHYSIOLOGICAL MEASUREMENT, 2010, 31 (11) : N79 - N83
  • [14] Using a Smartphone to Measure Heart Rate Changes during Relived Happiness and Anger
    Lakens, Daniel
    [J]. IEEE TRANSACTIONS ON AFFECTIVE COMPUTING, 2013, 4 (02) : 238 - 241
  • [15] A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects
    Lu, G.
    Yang, F.
    Taylor, J.A.
    Stein, J.F.
    [J]. Journal of Medical Engineering and Technology, 2009, 33 (08) : 634 - 641
  • [16] Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information?
    Lu S.
    Zhao H.
    Ju K.
    Shin K.
    Lee M.
    Shelley K.
    Chon K.H.
    [J]. Journal of Clinical Monitoring and Computing, 2008, 22 (1) : 23 - 29
  • [17] Mahdiani S, 2015, IEEE ENG MED BIO, P5948, DOI 10.1109/EMBC.2015.7319746
  • [18] Deriving heart period variability from blood pressure waveforms
    McKinley, PS
    Shapiro, PA
    Bagiella, E
    Myers, MM
    De Meersman, RE
    Grant, I
    Sloan, RP
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2003, 95 (04) : 1431 - 1438
  • [19] Towards Measuring Stress with Smartphones and Wearable Devices During Workday and Sleep
    Muaremi A.
    Arnrich B.
    Tröster G.
    [J]. BioNanoScience, 2013, 3 (02) : 172 - 183
  • [20] A Novel Method to Detect Heart Beat Rate Using a Mobile Phone
    Pelegris, P.
    Banitsas, K.
    Orbach, T.
    Marias, K.
    [J]. 2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 5488 - 5491