The aim of the study was to investigate surface and buildup region doses for 6MV and 15MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6MV and 15MV photon beams at 100 cm source-detector distance for 5 x 5, 10 x 10, and 20 x 20 cm(2) field sizes and 0 degrees, 30 degrees, 60 degrees, and 80 degrees beam angles. Thesurface doses using 6 MVphoton beams for 10 x 10 cm(2) field sizewere found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15MV photon beams for 10 x 10 cm(2) field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.
引用
收藏
页数:10
相关论文
共 28 条
  • [11] International Commission on Radiation Units and Measurements, 1985, 39 ICRU
  • [12] Characterization of MOSkin detector for in vivo skin dose measurement during megavoltage radiotherapy
    Jong, Wei Loong
    Wong, Jeannie Hsiu Ding
    Ung, Ngie Min
    Ng, Kwan Hoong
    Ho, Gwo Fuang
    Cutajar, Dean L.
    Rosenfeld, Anatoly B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05): : 120 - 132
  • [13] Khan F.M., 2010, The physics of radiation therapy
  • [14] ENTRANCE AND EXIT DOSE REGIONS FOR A CLINAC-2100C
    KLEIN, EE
    PURDY, JA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02): : 429 - 435
  • [15] Skin dose measurement by using ultra-thin TLDs
    Lin, JP
    Chu, TC
    Lin, SY
    Liu, MT
    [J]. APPLIED RADIATION AND ISOTOPES, 2001, 55 (03) : 383 - 391
  • [16] RADIOCHROMIC PLASTIC FILMS FOR ACCURATE MEASUREMENT OF RADIATION ABSORBED DOSE AND DOSE DISTRIBUTIONS
    MCLAUGHLIN, WL
    MILLER, A
    FIDAN, S
    PEJTERSEN, K
    PEDERSEN, WB
    [J]. RADIATION PHYSICS AND CHEMISTRY, 1977, 10 (02): : 119 - 127
  • [17] DETERMINATION OF BUILDUP REGION OVER-RESPONSE CORRECTIONS FOR A MARKUS-TYPE CHAMBER
    MELLENBERG, DE
    [J]. MEDICAL PHYSICS, 1990, 17 (06) : 1041 - 1044
  • [18] In vivo verification of superficial dose for head and neck treatments using intensity-modulated techniques
    Qi, Zhen-Yu
    Deng, Xiao-Wu
    Huang, Shao-Min
    Zhang, Li
    He, Zhi-Chun
    Li, X. Allen
    Kwan, Ian
    Lerch, Michael
    Cutajar, Dean
    Metcalfe, Peter
    Rosenfeld, Anatoly
    [J]. MEDICAL PHYSICS, 2009, 36 (01) : 59 - 70
  • [19] Angular Dependence of the MOSFET Dosimeter and its Impact on In vivo Surface Dose Measurement in Breast Cancer Treatment
    Qin, Songbing
    Chen, Ting
    Wang, Lili
    Tu, Yu
    Yue, Ning
    Zhou, Juying
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (04) : 345 - 352
  • [20] Measurement of radiotherapy x-ray skin dose on a chest wall phantom
    Quach, KY
    Morales, J
    Butson, MJ
    Rosenfeld, AB
    Metcalfe, PE