Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams

被引:28
作者
Akbas, Ugur [1 ]
Kesen, Nazmiye Donmez [1 ]
Koksal, Canan [1 ]
Bilge, Hatice [1 ]
机构
[1] Istanbul Univ, Inst Oncol, Div Med Phys, Istanbul, Turkey
关键词
RADIOCHROMIC FILM; DOSIMETRY; RADIOTHERAPY; CALIBRATION; WATER;
D O I
10.1155/2016/8361028
中图分类号
O412 [相对论、场论]; O572.2 [粒子物理学];
学科分类号
摘要
The aim of the study was to investigate surface and buildup region doses for 6MV and 15MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6MV and 15MV photon beams at 100 cm source-detector distance for 5 x 5, 10 x 10, and 20 x 20 cm(2) field sizes and 0 degrees, 30 degrees, 60 degrees, and 80 degrees beam angles. Thesurface doses using 6 MVphoton beams for 10 x 10 cm(2) field sizewere found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15MV photon beams for 10 x 10 cm(2) field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.
引用
收藏
页数:10
相关论文
共 28 条
  • [1] AIRD EGA, 1996, BRIT J RADIOLOGY S, V25, P1
  • [2] Surface dose and build-up region measurements with wedge filters for 6 and 18 MV photon beams
    Bilge, Hatice
    Ozbek, Nurdan
    Okutan, Murat
    Cakir, Aydin
    Acar, Hilal
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2010, 28 (02): : 110 - 116
  • [3] Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams
    Bilge, Hatice
    Cakir, Aydin
    Okutan, Murat
    Acar, Hilal
    [J]. PHYSICA MEDICA, 2009, 25 (02) : 101 - 104
  • [4] Extrapolated surface dose measurements with radiochromic film
    Butson, MJ
    Yu, PKN
    Metcalfe, PE
    [J]. MEDICAL PHYSICS, 1999, 26 (03) : 485 - 488
  • [5] A new radiotherapy surface dose detector: The MOSFET
    Butson, MJ
    Rozenfeld, A
    Mathur, JN
    Carolan, M
    Wong, TPY
    Metcalfe, PE
    [J]. MEDICAL PHYSICS, 1996, 23 (05) : 655 - 658
  • [6] Radiochromic film for medical radiation dosimetry
    Butson, MJ
    Yu, PKN
    Cheung, T
    Metcalfe, P
    [J]. MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) : 61 - 120
  • [7] Water equivalent properties of materials commonly used in proton dosimetry
    de Vera, Pablo
    Abril, Isabel
    Garcia-Molina, Rafael
    [J]. APPLIED RADIATION AND ISOTOPES, 2014, 83 : 122 - 127
  • [8] Accurate skin dose measurements using radiochromic film in clinical applications
    Devic, S
    Seuntjens, J
    Abdel-Rahman, W
    Evans, M
    Olivares, M
    Podgorsak, EB
    Vuong, T
    Soares, CG
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1116 - 1124
  • [9] The effects of thermoplastic masks used for immobilization of patients on surface and build-up for megavoltage radiation
    Erkan, H. Elvan
    Kemikler, Gonul
    [J]. TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2013, 28 (03): : 112 - 119
  • [10] MEASUREMENT OF DOSE IN THE BUILDUP REGION USING FIXED-SEPARATION PLANE-PARALLEL IONIZATION CHAMBERS
    GERBI, BJ
    KHAN, FM
    [J]. MEDICAL PHYSICS, 1990, 17 (01) : 17 - 26