Optimization of image quality and patient dose in radiographs of paediatric extremities using direct digital radiography

被引:15
|
作者
Jones, A. [1 ,2 ]
Ansell, C. [3 ]
Jerrom, C. [1 ]
Honey, I. D. [1 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Dept Med Phys, London, England
[2] Western Sydney Local Hlth Dist, Dept Med Phys, Sydney, NSW, Australia
[3] Evelina London Childrens Hosp, Guys & St Thomas NHS Fdn Trust, Dept Radiol, London, England
关键词
COMPUTED RADIOGRAPHY; SCREEN-FILM; FLAT-PANEL; REDUCTION; FILTRATION; FRACTURES; DETECTOR; NOISE; CHEST;
D O I
10.1259/bjr.20140660
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this study was to evaluate the effect of beamquality on the image quality (IQ) of ankle radiographs of paediatric patients in the age range of 0-1 year whilst maintaining constant effective dose (ED). Methods: Lateral ankle radiographs of an infant foot phantom were taken at a range of tube potentials (40.0-64.5 kV(p)) with and without 0.1-mm copper (Cu) filtration using a Trixell Pixium 4600 detector (Trixell, Morains, France). ED to the patient was computed for the default exposure parameters using PCXMC v. 2.0 and was fixed for other beam qualities by modulating the tube current-time product. The contrast-to-noise ratio (CNR) was measured between the tibia and adjacent soft tissue. The IQ of the phantom images was assessed by three radiologists and a reporting radiographer. Four IQ criteria were defined each with a scale of 1-3, giving a maximum score of 12. Finally, a service audit of clinical images at the default and optimum beam qualities was undertaken. Results: The measured CNR for the 40kVp/no Cu image was 12.0 compared with 7.6 for the default mode (55 0.1mm Cu). An improvement in the clinical IQ scores was also apparent at this lower beam quality. Conclusion: Lowering tube potential and removing filtration improved the clinical IQ of paediatric ankle radiographs in this age range. Advances in knowledge: There are currently no UK guidelines on exposure protocols for paediatric imaging using direct digital radiography. A lower beam quality will produce better IQ with no additional dose penalty for infant extremity imaging.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Image quality and dose differences caused by vendor-specific image processing of neonatal radiographs
    Sensakovic, William F.
    O'Dell, M. Cody
    Letter, Haley
    Kohler, Nathan
    Rop, Baiywo
    Cook, Jane
    Logsdon, Gregory
    Varich, Laura
    PEDIATRIC RADIOLOGY, 2016, 46 (11) : 1606 - 1613
  • [42] Reduction of patient dose in digital mammography: Simulation of low-dose image using computed radiography system and flat panel detector system
    Saito, Yuki
    Sakai, Maki
    Fujita, Naotoshi
    Kodera, Yoshie
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [43] DIAGNOSTIC IMAGE QUALITY AND DOSE REDUCTION USING NIOBIUM FILTRATION FOR CEPHALOMETRIC RADIOGRAPHY
    WAKOH, M
    FARMAN, AG
    KELLY, MS
    KUROYANAGI, K
    DENTOMAXILLOFACIAL RADIOLOGY, 1993, 22 (04) : 189 - 194
  • [44] Optimisation of direct digital chest radiography using Cu filtration
    Ekpo, Ernest U.
    Hoban, Alishja C.
    McEntee, Mark F.
    RADIOGRAPHY, 2014, 20 (04) : 346 - 350
  • [45] Chest Computed Radiography Imaging Parameters for Nodule Detection: A Comparison of Image Quality and Entrance Skin Dose
    Sun, Zhonghua
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2014, 4 (04) : 587 - 592
  • [46] Muscle and bone dose in paediatric limb digital radiography: a Monte Carlo evaluation
    Matyagin, Yuri, V
    McRobbie, Donald W.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (01) : 79 - 92
  • [47] Assessment of image quality and radiation dose in some models of digital radiography systems - A Pehamed FLUORAD A plus D phantom study
    Kaheni, H.
    Chaparian, A.
    Nafisi-Moghadam, R.
    Hamzian, N.
    RADIOPROTECTION, 2018, 53 (02) : 139 - 144
  • [48] Screen-film versus digital radiography of sacroiliac joints: evaluation of image quality and dose to patients
    Jablanovic, Dragoslav
    Ciraj-Bjelac, Olivera
    Damjanov, Nemanja
    Seric, Srdjan
    Radak-Perovic, Marija
    Arandjic, Danijela
    Maksimovic, Ruzica
    RADIATION PROTECTION DOSIMETRY, 2013, 155 (01) : 88 - 95
  • [49] Optimization of Exposure Factors and Image Quality for Computed Radiography
    Khotle, T.
    De Vos, H.
    Herbst, C. P.
    Rae, W. I. D.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 251 - 254
  • [50] Developing a novel phantom for image quality evaluation in digital radiography
    Yalcin, Asena
    Olgar, Turan
    RADIATION PHYSICS AND CHEMISTRY, 2024, 222