Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms

被引:91
作者
Followill, David S. [1 ]
Evans, DeeAnn Radford [1 ]
Cherry, Christopher [1 ]
Molineu, Andrea [1 ]
Fisher, Gary [1 ]
Hanson, William F. [1 ]
Lbbott, Geoffrey S. [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
quality assurance; anthropomorphic phantoms;
D O I
10.1118/1.2737158
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Radiological Physics Center (RPC) developed two heterogeneous anthropomorphic quality assurance phantoms for use in verifying the accuracy of radiation delivery: one for intensity-modulated radiation therapy (IMRT) to the pelvis and the other for stereotactic body radiation therapy (SBRT) to the thorax. The purpose of this study was to describe the design and development of these two phantoms and to demonstrate the reproducibility of measurements generated with them. The phantoms were built to simulate actual patient anatomy. They are lightweight and water-fillable, and they contain imageable targets and organs at risk of radiation exposure that are of similar densities to their human counterparts. Dosimetry inserts accommodate radiochromic film for relative dosimetry and thermoluminesent dosimetry capsules for absolute dosimetry. As a part of the commissioning process, each phantom was imaged, treatment plans were developed, and radiation was delivered at least three times. Under these controlled irradiation conditions, the reproducibility of dose delivery to the target TLD in the pelvis and thorax phantoms was 3% and 0.5%, respectively. The reproducibility of radiation-field localization was less than 2.5 mm for both phantoms. Using these anthropomorphic phantoms, pelvic IMRT and thoracic SBRT radiation treatments can be verified with a high level of precision. These phantoms can be used to effectively credential institutions for participation in specific NCI-sponsored clinical trials (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:2070 / 2076
页数:7
相关论文
共 28 条
[1]  
AGUIRRE JF, 2002, STANDARDS CODES PRAC, V2, P191
[2]   Evaluation of doses delivered by SBRT to the lung of an anthropomorphic thorax phantom [J].
Alvarez, P ;
Molineu, A ;
Hernandez, N ;
Followill, D ;
Ibbott, G .
MEDICAL PHYSICS, 2005, 32 (06) :2043-2043
[3]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[4]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[5]   Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method [J].
Butson, MJ ;
Elferink, R ;
Cheung, T ;
Yu, PKN ;
Stokes, M ;
Quach, KY ;
Metcalfe, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :N143-N149
[6]   Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium [J].
Carrasco, P ;
Jornet, N ;
Duch, MA ;
Weber, L ;
Ginjaume, M ;
Eudaldo, T ;
Jurado, D ;
Ruiz, A ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (10) :2899-2911
[7]  
CHERRY CP, 2000, P 2000 WORLD C MED P
[8]   The evaluation of several commercial IMRT treatment planning systems heterogeneity dose calculation algorithms using an anthropomorphic thorax phantom [J].
Davidson, S ;
Followill, D ;
Ibbott, G ;
Prado, K .
MEDICAL PHYSICS, 2005, 32 (06) :1988-1988
[9]  
Fisher G, 2003, MED PHYS, V30, P1507
[10]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829