Dosimetry of the I-Plant Model 3500 iodine-125 brachytherapy source

被引:25
作者
Duggan, DM [1 ]
Johnson, BL [1 ]
机构
[1] Vanderbilt Univ, Med Ctr, Dept Radiat Oncol, Nashville, TN 37232 USA
关键词
dosimetry; I-125; Task Group 43; brachytherapy;
D O I
10.1118/1.1357456
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
I-125 brachytherapy sources have been widely used for interstitial implants for a number of years in several tumor sites, especially the prostate. The design of the new I-Plant Model 3500 iodine source is novel, yet its characteristics are similar to those of two existing designs, Model 6711 and the Symmetra. Dosimetry parameters (including dose rate constant, radial dose function, and anisotropy function, as defined by AAPM Task Group 43) were measured with LiF thermoluminescent dosimeters in water-equivalent plastic phantoms. The dose rate constant was found by direct comparison of calibrated I-Plant Model 3500 and Model 6711 seeds in a solid water phantom, to be 1.01 (cGy/h)/U. The radial dose function and anisotropy function are similar to those of the Model 6711 and Symmetra seeds. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:661 / 670
页数:10
相关论文
共 50 条
  • [41] ECHOGRAPHIC FOLLOW UP OF MALIGNANT MELANOMA OF THE CHOROID AFTER BRACHYTHERAPY WITH IODINE-125
    Abraham-Marin, M.
    Cortes-Luna, C.
    Morales-Canton, V.
    Velasco-Barona, C.
    Moragrega-Adame, E.
    ACTA CLINICA CROATICA, 2012, 51 : 79 - 81
  • [42] Five-year potency preservation after iodine-125 prostate brachytherapy
    Shuichi Nishimura
    Atsunori Yorozu
    Toshio Ohashi
    Masanori Sakayori
    Yasuto Yagi
    Toru Nishiyama
    Shiro Saito
    Yutaka Shiraishi
    Kayo Yoshida
    Kazuhito Toya
    Naoyuki Shigematsu
    International Journal of Clinical Oncology, 2014, 19 : 940 - 945
  • [43] Five-year potency preservation after iodine-125 prostate brachytherapy
    Nishimura, Shuichi
    Yorozu, Atsunori
    Ohashi, Toshio
    Sakayori, Masanori
    Yagi, Yasuto
    Nishiyama, Toru
    Saito, Shiro
    Shiraishi, Yutaka
    Yoshida, Kayo
    Toya, Kazuhito
    Shigematsu, Naoyuki
    INTERNATIONAL JOURNAL OF CLINICAL ONCOLOGY, 2014, 19 (05) : 940 - 945
  • [44] Iodine-125 brachytherapy for a desmoid tumor in the abdomen-thorax: A retrospective study
    Chuang, He
    Xiao-Han, Huang
    Li, Liu
    Dong-Yuan, Li
    Liang-Shan, Li
    Ting-Yuan, Li
    Qing-Hua, Liang
    He-Nan, Liu
    Wei, Li
    Xue-Quan, Huang
    BRACHYTHERAPY, 2023, 22 (06) : 851 - 857
  • [45] Prostatic length predicts functional outcomes after iodine-125 prostate brachytherapy
    Pal, Raj P.
    Bhatt, Jaimin R.
    Khan, Masood A.
    Duggleby, Stuart
    Camilleri, Philip
    Bell, C. Richard
    Elwell, Christine
    Kunkler, Roger B.
    BRACHYTHERAPY, 2011, 10 (02) : 107 - 116
  • [46] Outcomes following iodine-125 prostate brachytherapy with or without neoadjuvant androgen deprivation
    Ohashi, Toshio
    Yorozu, Atsunori
    Saito, Shiro
    Momma, Tetsuo
    Toya, Kazuhito
    Nishiyama, Toru
    Yamashita, Shoji
    Shiraishi, Yutaka
    Shigematsu, Naoyuki
    RADIOTHERAPY AND ONCOLOGY, 2013, 109 (02) : 241 - 245
  • [47] Glaucoma After Iodine-125 Brachytherapy for Uveal Melanoma: Incidence and Risk Factors
    Kim, Eun-Ah
    Salazar, Diana
    McCannel, Colin A.
    Kamrava, Mitchell
    Demanes, David J.
    Lamb, James
    Caprioli, Joseph
    McCannel, Tara A.
    JOURNAL OF GLAUCOMA, 2020, 29 (01) : 1 - 10
  • [48] Perianal synovial sarcoma treated postoperatively with Iodine-125 brachytherapy: Technical details
    Tovey, Susan
    Stannard, Clare
    Maree, Genie
    Hiscock, Robin
    Thomson, Charlie
    Panieri, Eugenio
    Goldberg, Paul
    BRACHYTHERAPY, 2017, 16 (03) : 565 - 571
  • [49] Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source
    Meigooni, AS
    Yoe-Sein, MM
    Al-Otoom, AY
    Sowards, KT
    APPLIED RADIATION AND ISOTOPES, 2002, 56 (04) : 589 - 599
  • [50] Tolerance of the aorta using intraoperative iodine-125 interstitial brachytherapy in cancer of the lung
    Trombetta, Mark G.
    Colonias, Athanasios
    Makishi, Daryl
    Keenan, Robert
    Werts, E. Day
    Landreneau, Rodney
    Parda, David S.
    BRACHYTHERAPY, 2008, 7 (01) : 50 - 54