Dosimetry of the I-Plant Model 3500 iodine-125 brachytherapy source

被引:25
|
作者
Duggan, DM [1 ]
Johnson, BL [1 ]
机构
[1] Vanderbilt Univ, Med Ctr, Dept Radiat Oncol, Nashville, TN 37232 USA
关键词
dosimetry; I-125; Task Group 43; brachytherapy;
D O I
10.1118/1.1357456
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
I-125 brachytherapy sources have been widely used for interstitial implants for a number of years in several tumor sites, especially the prostate. The design of the new I-Plant Model 3500 iodine source is novel, yet its characteristics are similar to those of two existing designs, Model 6711 and the Symmetra. Dosimetry parameters (including dose rate constant, radial dose function, and anisotropy function, as defined by AAPM Task Group 43) were measured with LiF thermoluminescent dosimeters in water-equivalent plastic phantoms. The dose rate constant was found by direct comparison of calibrated I-Plant Model 3500 and Model 6711 seeds in a solid water phantom, to be 1.01 (cGy/h)/U. The radial dose function and anisotropy function are similar to those of the Model 6711 and Symmetra seeds. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:661 / 670
页数:10
相关论文
共 50 条
  • [1] Comprehensive Monte Carlo calculations of AAPM task group report no. 43 dosimetry parameters for the model 3500 I-Plant 125I brachytherapy source
    Rivard, MJ
    APPLIED RADIATION AND ISOTOPES, 2002, 57 (03) : 381 - 389
  • [2] Experimental determination of the dosimetric characteristics for the I-Plant™ model 3600 125I brachytherapy source
    Tries, MA
    Mancevice, CG
    Munro, JJ
    MEDICAL PHYSICS, 2005, 32 (06) : 1469 - 1472
  • [3] Model 3500 125I brachytherapy source dosimetric characterization
    Wallace, RE
    APPLIED RADIATION AND ISOTOPES, 2002, 56 (04) : 581 - 587
  • [4] Report on the dosimetry of a new design 125Iodine brachytherapy source
    Wallace, RE
    Fan, JJ
    MEDICAL PHYSICS, 1999, 26 (09) : 1925 - 1931
  • [5] Evaluation of a new brachytherapy iodine-125 source by AAPM TG43 formalism
    Wallace, RE
    Fan, JJ
    MEDICAL PHYSICS, 1998, 25 (11) : 2190 - 2196
  • [6] Dosimetric characterization of a new 125Iodine brachytherapy source, model I125-SL
    Wallace, RE
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 376 - 379
  • [7] Iodine-125 brachytherapy for brain tumours - a review
    Schwarz, Silke B.
    Thon, Niklas
    Nikolajek, Katharina
    Niyazi, Maximilian
    Tonn, Joerg-Christian
    Belka, Claus
    Kreth, Friedrich-Wilhelm
    RADIATION ONCOLOGY, 2012, 7
  • [8] Rectal dose constraints for salvage iodine-125 prostate brachytherapy
    Peters, Max
    Hoekstra, Carel J.
    van Zyp, Jochem R. N. van der Voort
    Westendorp, Hendrik
    van de Pol, Sandrine M. G.
    Moerland, Marinus A.
    Maenhout, Metha
    Kattevilder, Rob
    van Vulpen, Marco
    BRACHYTHERAPY, 2016, 15 (01) : 85 - 93
  • [9] Empirical dosimetric characterization of model I125-SL 125Iodine brachytherapy source in phantom
    Wallace, RE
    MEDICAL PHYSICS, 2000, 27 (12) : 2796 - 2802
  • [10] Urethral and bladder dosimetry of total and focal salvage Iodine-125 prostate brachytherapy: Late toxicity and dose constraints
    Peters, Max
    van Zyp, Jochem van der Voort
    Hoekstra, Carel
    Westendorp, Hendrik
    van de Pol, Sandrine
    Moerland, Marinus
    Maenhout, Metha
    Kattevilder, Rob
    van Vulpen, Marco
    RADIOTHERAPY AND ONCOLOGY, 2015, 117 (02) : 262 - 269