Comparison of swept-source versus spectral-domain optical coherence tomography angiography for detection of macular neovascularization

被引:4
作者
Lentzsch, Anna [1 ,2 ]
Schoellhorn, Laura [1 ,2 ]
Schnorr, Christel [1 ,2 ]
Siggel, Robert [1 ,2 ,3 ]
Liakopoulos, Sandra [1 ,2 ,4 ]
机构
[1] Cologne Image Reading Ctr, Fac Med, Dept Ophthalmol, Cologne, Germany
[2] Univ Hosp Cologne, Cologne, Germany
[3] Univ Witten Herdecke, Helios Univ Hosp Wuppertal, Dept Ophthalmol, Wuppertal, Germany
[4] Goethe Univ, Dept Ophthalmol, Frankfurt, Germany
关键词
Optical coherence tomography angiography; Macular neovascularization; Swept-source OCTA; Spectral-domain OCTA; OCT-ANGIOGRAPHY; CHOROIDAL NEOVASCULARIZATION; IMAGE ARTIFACTS; QUANTITATION;
D O I
10.1007/s00417-021-05229-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare swept-source (SS) versus spectral-domain (SD) optical coherence tomography angiography (OCTA) for the detection of macular neovascularization (MNV). Methods In this prospective cohort study, 72 eyes of 54 patients with subretinal hyperreflective material (SHRM) and/or pigment epithelial detachment (PED) on OCT possibly corresponding to MNV in at least one eye were included. OCTA scans were acquired using two devices, the PLEX Elite 9000 SS-OCTA and the Spectralis SD-OCTA. Fluorescein angiography (FA) was used as reference. Two graders independently evaluated en face OCTA images using a preset slab as well as a manually modified slab, followed by a combination of en face and cross-sectional OCTA. Results Sensitivity (specificity) for the automated slabs was 51.7% (93.0%) for SS-OCTA versus 58.6% (95.3%) for SD-OCTA. Manual modification of segmentation increased sensitivity to 79.3% for SS-OCTA but not for SD-OCTA (58.6%). The combination of en face OCTA with cross-sectional OCTA reached highest sensitivity values (SS-OCTA: 82.8%, SD-OCTA: 86.2%), and lowest number of cases with discrepancies between SS-OCTA and SD-OCTA (4.2%). Fleiss kappa as measure of concordance between FA, SS-OCTA, and SD-OCTA was 0.56 for the automated slabs, 0.60 for the manual slabs, and 0.73 (good agreement) for the combination of en face OCTA with cross-sectional OCTA. Concordance to FA was moderate for the automated slabs and good for manual slabs and combination with cross-sectional OCTA of both devices. Conclusion Both devices reached comparable results regarding the detection of MNV on OCTA. Sensitivity for MNV detection and agreement between devices was best when evaluating a combination of en face and cross-sectional OCTA.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 21 条
  • [11] The Diagnostic Accuracy of Optical Coherence Tomography Angiography for Neovascular Age-Related Macular Degeneration: A Comparison with Fundus Fluorescein Angiography
    Gong, Jingwen
    Yu, Suqin
    Gong, Yuanyuan
    Wang, Fenghua
    Sun, Xiaodong
    [J]. JOURNAL OF OPHTHALMOLOGY, 2016, 2016
  • [12] A Comparison Between Optical Coherence Tomography Angiography and Fluorescein Angiography for the Imaging of Type 1 Neovascularization
    Inoue, Maiko
    Jung, Jesse J.
    Balaratnasingam, Chandrakumar
    Dansingani, Kunal K.
    Dhrami-Gavazi, Elona
    Suzuki, Mihoko
    de Carlo, Talisa E.
    Shahlaee, Abtin
    Klufas, Michael A.
    El Maftouhi, Adil
    Duker, Jay S.
    Ho, Allen C.
    Quaranta-El Maftouhi, Maddalena
    Sarraf, David
    Freund, K. Bailey
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (09) : OCT314 - OCT323
  • [13] Accurate OCT-angiography Interpretation - Detection and Exclusion of Artifacts
    Lang, Gabriele E.
    Enders, Christian
    Loidl, Max
    Lang, Gerhard K.
    Werner, Jens Ulrich
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2017, 234 (09) : 1109 - 1118
  • [14] Choroidal Neovascularization Analyzed on Ultrahigh-Speed Swept-Source Optical Coherence Tomography Angiography Compared to Spectral-Domain Optical Coherence Tomography Angiography
    Novais, Eduardo A.
    Adhi, Mehreen
    Moult, Eric M.
    Louzada, Ricardo N.
    Cole, Emily D.
    Husvogt, Lennart
    Lee, Byungkun
    Dang, Sabin
    Regatieri, Caio V. S.
    Witkin, Andre J.
    Baumal, Caroline R.
    Hornegger, Joachim
    Jayaraman, Vijaysekhar
    Fujimoto, James G.
    Duker, Jay S.
    Waheed, Nadia K.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 164 : 80 - 88
  • [15] CHOROIDAL NEOVASCULAR AREA AND VESSEL DENSITY COMPARISON BETWEEN TWO SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY DEVICES
    Ohayon, Avi
    Sacconi, Riccardo
    Semoun, Oudy
    Corbelli, Eleonora
    Souied, Eric H.
    Querques, Giuseppe
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2020, 40 (03): : 521 - 528
  • [16] Comparison of Automated versus Manually Modified OCT Angiography En Face Slabs for Detection of Choroidal Neovascularization
    Siggel, Robert
    Spital, Christel
    Lentzsch, Anna
    Liakopoulos, Sandra
    [J]. OPHTHALMOLOGY RETINA, 2020, 4 (05): : 471 - 480
  • [17] Optical coherence tomography angiography
    Spaide, Richard F.
    Fujimoto, James G.
    Waheed, Nadia K.
    Sadda, Srinivas R.
    Staurenghi, Giovanni
    [J]. PROGRESS IN RETINAL AND EYE RESEARCH, 2018, 64 : 1 - 55
  • [18] IMAGE ARTIFACTS IN OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY
    Spaide, Richard F.
    Fujimoto, James G.
    Waheed, Nadia K.
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2015, 35 (11): : 2163 - 2180
  • [19] Retinal Vascular Layers Imaged by Fluorescein Angiography and Optical Coherence Tomography Angiography
    Spaide, Richard F.
    Klancnik, James M., Jr.
    Cooney, Michael J.
    [J]. JAMA OPHTHALMOLOGY, 2015, 133 (01) : 45 - 50
  • [20] Zhang QQ, 2017, OPHTHALMOL RETINA, V1, P124, DOI 10.1016/j.oret.2016.08.005