Dose comparison of classical 2-plane DSA and 3D rotational angiography for the assessment of intracranial aneurysms

被引:15
作者
Guberina, N. [1 ]
Lechel, U. [2 ]
Forsting, M. [1 ]
Moenninghoff, C. [1 ]
Dietrich, U. [1 ]
Ringelstein, A. [1 ]
机构
[1] Univ Hosp Essen, Inst Diagnost & Intervent Radiol & Neuroradiol, Hufelandstr 55, D-45147 Essen, Germany
[2] Fed Off Radiat Protect, Dept Med & Occupat Radiat Protect, Neuherberg, Germany
关键词
Effective dose; 2-plane DSA; 3D rotational angiography; Experimental phantom study; Dose measurement; DIGITAL-SUBTRACTION-ANGIOGRAPHY; SUBARACHNOID HEMORRHAGE; ENDOVASCULAR TREATMENT; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; PATIENT; NEURORADIOLOGY; 2D;
D O I
10.1007/s00234-016-1671-4
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this experimental phantom study was to compare radiation doses imparted to patients undergoing classical two-plane digital subtraction angiography (2-plane DSA) and 3D rotational angiography in interventional neuroradiology. Thermoluminescence dosimeter (TLD) measurements were performed at an anthropomorphic phantom using a digital interventional angiography system. Two-plane DSA included posterior/anterior (PA) and lateral (LAT) projections (frame-rate, 7.6 frames (PA) and 9.8 frames (LAT) for a scan time of approximately 8 s; image intensifier 27 cm (PA) and 25 cm (LAT)). For 3D rotational angiography, 122 images were acquired from one single image run with the imaging system rotating 240A degrees around the phantom's head (image intensifier 37 cm). Effective dose was 0.4 mSv for 2-plane DSA compared to 0.1 mSv for 3D rotational angiography. Organ doses were approximately two to five times higher for classical 2-plane technique compared to the 3D rotational angiography, respectively: brain (11.4 vs. 2.4 mSv), eye lens (4.5 vs.1 mSv), salivary glands (7 vs. 1,7 mSv), oral mucosa (2.7 vs.0.9 mSv), thyroid (0.5 vs. 0.2 mSv), thymus (0.2 vs. 0.05 mSv), bone marrow within imaged region (1 vs. 0.2 mSv), oesophagus (0.07 vs. 0.03 mSv), endotracheal system (2.6 vs. 0.7 mSv) and skeletal components in the imaged region (0.7 vs. 0.2 mSv). Three-dimensional rotational angiography clearly reduces radiation doses compared to the classical 2-plane technique. Replacement of additional 2-plane DSA projections with 3D rotational angiography will lead to a remarkable decrease in patient radiation dose, without loss of image quality. Thus, we recommend routine application of 3D rotational angiography, in particular for diagnostic assessment of aneurysm morphology.
引用
收藏
页码:673 / 678
页数:6
相关论文
共 19 条
  • [1] Abe T, 2002, AM J NEURORADIOL, V23, P686
  • [2] Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology procedures
    Alexander, M. D.
    Oliff, M. C.
    Olorunsola, O. G.
    Brus-Ramer, M.
    Nickoloff, E. L.
    Meyers, P. M.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2010, 2 (01) : 6 - 10
  • [3] Intracranial aneurysms: Clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment
    Anxionnat, R
    Bracard, S
    Ducrocq, X
    Trousset, Y
    Launay, L
    Kerrien, E
    Braun, M
    Vaillant, R
    Scomazzoni, F
    Lebedinsky, A
    Picard, L
    [J]. RADIOLOGY, 2001, 218 (03) : 799 - 808
  • [4] Patient dose from 3D rotational neurovascular studies
    Bridcut, R. R.
    Murphy, E.
    Workman, A.
    Flynn, P.
    Winder, R. J.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (953) : 362 - 366
  • [5] Comparison of 2D Digital Subtraction Angiography and 3D Rotational Angiography in the Evaluation of Dome-to-Neck Ratio
    Brinjikji, W.
    Cloft, H.
    Lanzino, G.
    Kallmes, D. F.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (04) : 831 - 834
  • [6] Cristy M., 1987, ORNLTM8381V1
  • [7] Radiation exposure with 3D rotational angiography of the skull
    Gosch, D.
    Kurze, W.
    Deckert, F.
    Schulz, T.
    Patz, A.
    Kahn, T.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (09): : 880 - 885
  • [8] Hirai T, 2003, AM J NEURORADIOL, V24, P1067
  • [9] Angiogram-negative subarachnoid hemorrhage in the era of three dimensional rotational angiography
    Ishihara, Hideyuki
    Kato, Shoichi
    Akimura, Tatsuo
    Suehiro, Eiichi
    Oku, Takayuki
    Suzuki, Michiyasu
    [J]. JOURNAL OF CLINICAL NEUROSCIENCE, 2007, 14 (03) : 252 - 255
  • [10] Kiyosue H, 2002, AM J NEURORADIOL, V23, P1206