Shading correction for on-board cone-beam CT in radiation therapy using planning MDCT images

被引:127
作者
Niu, Tianye [1 ,2 ]
Sun, Mingshan [3 ]
Star-Lack, Josh [3 ]
Gao, Hewei [4 ]
Fan, Qiyong [1 ,2 ]
Zhu, Lei [1 ,2 ]
机构
[1] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Nucl & Radiol Engn Program, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Med Phys Program, Atlanta, GA 30332 USA
[3] Varian Med Syst, Ginzton Technol Ctr, Mountain View, CA 94043 USA
[4] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
关键词
shading correction; scatter correction; radiation therapy; cone-beam CT; X-RAY SCATTER; DOSE CALCULATION ACCURACY; COMPUTED-TOMOGRAPHY; CORRECTION ALGORITHM; PRIMARY MODULATION; SPREAD FUNCTION; RECONSTRUCTION; REGISTRATION; RADIOTHERAPY; ARRAY;
D O I
10.1118/1.3483260
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Applications of cone-beam CT (CBCT) to image-guided radiation therapy (IGRT) are hampered by shading artifacts in the reconstructed images. These artifacts are mainly due to scatter contamination in the projections but also can result from uncorrected beam hardening effects as well as nonlinearities in responses of the amorphous silicon flat panel detectors. While currently, CBCT is mainly used to provide patient geometry information for treatment setup, more demanding applications requiring high-quality CBCT images are under investigation. To tackle these challenges, many CBCT correction algorithms have been proposed; yet, a standard approach still remains unclear. In this work, we propose a shading correction method for CBCT that addresses artifacts from low-frequency projection errors. The method is consistent with the current workflow of radiation therapy. Methods: With much smaller inherent scatter signals and more accurate detectors, diagnostic multidetector CT (MDCT) provides high quality CT images that are routinely used for radiation treatment planning. Using the MDCT image as "free" prior information, we first estimate the primary projections in the CBCT scan via forward projection of the spatially registered MDCT data. Since most of the CBCT shading artifacts stem from low-frequency errors in the projections such as scatter, these errors can be accurately estimated by low-pass filtering the difference between the estimated and raw CBCT projections. The error estimates are then subtracted from the raw CBCT projections. Our method is distinct from other published correction methods that use the MDCT image as a prior because it is projection-based and uses limited patient anatomical information from the MDCT image. The merit of CBCT-based treatment monitoring is therefore retained. Results: The proposed method is evaluated using two phantom studies on tabletop systems. On the Catphan(C)600 phantom, our approach reduces the reconstruction error from 348 Hounsfield unit (HU) without correction to 4 HU around the object center after correction, and from 375 HU to 17 HU in the high-contrast regions. In the selected regions of interest (ROIs), the average image contrast is increased by a factor of 3.3. When noise suppression is implemented, the proposed correction substantially improves the contrast-to-noise ratio (CNR) and therefore the visibility of low-contrast objects, as seen in a more challenging pelvis phantom study. Besides a significant improvement in image uniformity, a low-contrast object of similar to 25 HU, which is otherwise buried in the shading artifacts, can be clearly identified after the proposed correction due to a CNR increase of 3.1. Compared to a kernel-based scatter correction method coupled with an analytical beam hardening correction, our approach also shows an overall improved performance with some residual artifacts. Conclusions: By providing effective shading correction, our approach has the potential to improve the accuracy of more advanced CBCT-based clinical applications for IGRT, such as tumor delineation and dose calculation. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3483260]
引用
收藏
页码:5395 / 5406
页数:12
相关论文
共 41 条
[1]   CHARACTERIZATION OF THE POINT SPREAD FUNCTION AND MODULATION TRANSFER-FUNCTION OF SCATTERED RADIATION USING A DIGITAL IMAGING-SYSTEM [J].
BOONE, JM ;
ARNOLD, BA ;
SEIBERT, JA .
MEDICAL PHYSICS, 1986, 13 (02) :254-256
[2]   AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[3]   Inclusion of organ deformation in dose calculations [J].
Brock, KK ;
McShan, DL ;
Ten Haken, RK ;
Hollister, SJ ;
Dawson, LA ;
Balter, JM .
MEDICAL PHYSICS, 2003, 30 (03) :290-295
[4]   Accelerated simulation of cone beam X-ray scatter projections [J].
Colijn, AP ;
Beekman, FJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (05) :584-590
[5]   Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines [J].
Coselmon, MM ;
Balter, JM ;
McShan, DL ;
Kessler, ML .
MEDICAL PHYSICS, 2004, 31 (11) :2942-2948
[6]   Formulating adaptive radiation therapy (ART) treatment planning into a closed-loop control framework [J].
de la Zerda, Adam ;
Armbruster, Benjamin ;
Xing, Lei .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4137-4153
[7]   Magnitude and effects of x-ray scatter in a 256-slice CT scanner [J].
Endo, Masahiro ;
Mori, Shinichiro ;
Tsunoo, Takanori ;
Miyazaki, Hiroaki .
MEDICAL PHYSICS, 2006, 33 (09) :3359-3368
[8]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[9]   Scatter correction method for x-ray CT using primary modulation: Phantom studies [J].
Gao, Hewei ;
Fahrig, Rebecca ;
Bennett, N. Robert ;
Sun, Mingshan ;
Star-Lack, Josh ;
Zhu, Lei .
MEDICAL PHYSICS, 2010, 37 (02) :934-946
[10]   Automated registration of large deformations for adaptive radiation therapy of prostate cancer [J].
Godley, Andrew ;
Ahunbay, Ergun ;
Peng, Cheng ;
Li, X. Allen .
MEDICAL PHYSICS, 2009, 36 (04) :1433-1441