PET imaging for treatment verification of ion therapy: Implementation and experience at GSI Darmstadt and MGH Boston

被引:59
作者
Parodi, Katia [1 ]
Bortfeld, Thomas [2 ]
Enghardt, Wolfgang [3 ,4 ]
Fiedler, Fine [3 ]
Knopf, Antje [2 ]
Paganetti, Harald [2 ]
Pawelke, Joerg [3 ]
Shakirin, Georgy [3 ]
Shih, Helen [2 ]
机构
[1] Heidelberg Ion Beam Therapy Ctr, D-69120 Heidelberg, Germany
[2] Massachusetts Gen Hosp, Boston, MA 02114 USA
[3] Forschungszentrum Dresden Rossendorf, D-01314 Dresden, Germany
[4] Tech Univ Dresden, D-01307 Dresden, Germany
关键词
positron emission tomography; proton therapy; carbon ion therapy;
D O I
10.1016/j.nima.2008.03.075
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Ion beams offer the possibility of improved conformation of the dose delivered to the tumor with better sparing of surrounding tissue and critical structures in comparison to conventional photon and electron external radiation treatment modalities. Full clinical exploitation of this advantage can benefit from in vivo confirmation of the actual beam delivery and, in particular, of the ion range in the patient. During irradiation, positron emitters like (15)O (half-life T(1/2) approximate to 2 min) and (11)C (T(1/2) approximate to 20 min) are formed in nuclear interactions between the ions and the tissue. Detection of this transient radioactivity via positron emission tomography (PET) and comparison with the expectation based on the prescribed beam application may serve as an in vivo, non-invasive range validation method of the whole treatment planning and delivery chain. For technical implementation, PET imaging during irradiation (in-beam) requires the development of customized, limited angle detectors with data acquisition synchronized with the beam delivery. Alternatively, commercial PET or PET/CT scanners in close proximity to the treatment site enable detection of the residual activation from long-lived emitters shortly after treatment (offline). This paper reviews two clinical examples using a dedicated in-beam PET scanner for verification of carbon ion therapy at GSI Darmstadt, Germany, as well as a commercial offline PET/CT tomograph for post-radiation imaging of proton treatments at Massachusetts General Hospital, Boston, USA. Challenges as well as pros and cons of the two imaging approaches in dependence of the different ion type and beam delivery system are discussed. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:282 / 286
页数:5
相关论文
共 17 条
[1]  
[Anonymous], 2005, CERN YELLOW REPORTS
[2]   On the detector arrangement for in-beam PET for hadron therapy monitoring [J].
Crespo, Paulo ;
Shakirin, Georgy ;
Enghardt, Wolfgang .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (09) :2143-2163
[3]   Dose quantification from in-beam positron emission tomography [J].
Enghardt, W ;
Parodi, K ;
Crespo, P ;
Fiedler, F ;
Pawelke, J ;
Pönisch, F .
RADIOTHERAPY AND ONCOLOGY, 2004, 73 :S96-S98
[4]   Charged hadron tumour therapy monitoring by means of PET [J].
Enghardt, W ;
Crespo, P ;
Fiedler, F ;
Hinz, R ;
Parodi, K ;
Pawelke, J ;
Pönisch, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2) :284-288
[5]  
FASSO A, 2003, P CHEP C E CONF
[6]  
Fiedler F, 2006, IEEE T NUCL SCI, V53, P2252, DOI 10.1109/TNS.2006.877854
[7]   Quantitative assessment of the accuracy of proton beam range verification with PET/CT [J].
Knopf', A. ;
Parodi, K. ;
Paganetti, H. ;
Cascio, E. ;
Bonab, A. ;
Bortfeld, T. .
MEDICAL PHYSICS, 2007, 34 (06) :2508-2508
[8]   Clinical CT-based calculations of dose and positron emitter distributions in proton therapy using the FLUKA Monte Carlo code [J].
Parodi, K. ;
Ferrari, A. ;
Sommerer, F. ;
Paganetti, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12) :3369-3387
[9]   Experimental study on the feasibility of in-beam PET for accurate monitoring of proton therapy [J].
Parodi, K ;
Pönisch, F ;
Enghardt, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) :778-786
[10]   Random coincidences during in-beam PET measurements at microbunched therapeutic ion beams [J].
Parodi, K ;
Crespo, P ;
Eickhoff, H ;
Haberer, T ;
Pawelke, J ;
Schardt, D ;
Enghardt, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 545 (1-2) :446-458