Design of Dual-Beam Spectrometer in Spectrophotometer for Colorimetry

被引:0
作者
Liu Yi-xuan [1 ,2 ]
Yan Chang-xiang [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, Changchun 130033, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
关键词
Optical design; Spectrophotometer for colorimetry; Dual-beam spectrometer; Dual-beam spectrophotometer; Portable; Double-slit; CZERNY-TURNER SPECTROMETER; ASTIGMATISM;
D O I
10.3964/j.issn.1000-0593(2015)07-2060-05
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Spectrophotometers for colorimetry are usually composed of two independent and identical spectrometers. In order to reduce the volume of spectrophotometer for colorimetry, a design method of double-beam spectrometer is put forward. A traditional spectrometer is modified so that a new spectrometer can realize the function of double spectrometers, which is especially suitable for portable instruments. One slit is replaced by the double-slit, than two beams of spectrum can be detected. The working principle and design requirement of double-beam spectrometer are described. A spectrometer of portable spectrophotometer is designed by this method. A toroidal imaging mirror is used for the Czerny-Turner double-beam spectrometer in this paper, which can better correct astigmatism, and prevent the dual-beam spectral crosstalk. The results demonstrate that the double-beam spectrometer designed by this method meets the design specifications, with the spectral resolution less than 10 nm, the spectral length of 9. 12 mm, and the volume of 57 mmX54 mmX23 mm, and without the dual-beam spectral overlap in the detector either. Comparing with a traditional spectrophotometer, the modified spectrophotoMeter uses a set of double-beam spectrometer instead of two sets of spectrometers, which can greatly reduce the volume. This design method can be specially applied in portable spectrophotometers, also can be widely applied in other double-beam spectrophotometers, which offers a new idea for the design of dual-beam spectrophotometers.
引用
收藏
页码:2060 / 2064
页数:5
相关论文
共 10 条
[1]   Highly efficient and aberration-corrected spectrometer for advanced Raman spectroscopy [J].
Futamata, M ;
Takenouchi, T ;
Katakura, K .
APPLIED OPTICS, 2002, 41 (22) :4655-4665
[2]   Industrial colour measurement - the state of the art [J].
Hirschler, R ;
Gay, J .
AIC: 9TH CONGRESS OF THE INTERNATIONAL COLOUR ASSOCIATION, 2002, 4421 :773-776
[3]  
HU Wei-jie, 1998, CHINESE J SCI INSTRU, V19, P407
[4]  
Imura K, 2005, U. S. Patent, Patent No. 6917429
[5]   Broadband astigmatism-corrected Czerny-Turner spectrometer [J].
Lee, Kye-Sung ;
Thompson, Kevin P. ;
Rolland, Jannick P. .
OPTICS EXPRESS, 2010, 18 (22) :23378-23384
[6]  
LIU Kang, 2013, ACTA PHOTONICA SINIC, V10, P1218
[7]  
Rinke G., 1993, U. S. Patent, Patent No. 5251007
[8]   CORRECTING FOR ASTIGMATISM IN CZERNY-TURNER SPECTROMETER AND SPECTROGRAPH [J].
SHAFER, AB .
APPLIED OPTICS, 1967, 6 (01) :159-&
[9]  
[薛庆生 Xue Qingsheng], 2009, [光学学报, Acta Optica Sinica], V29, P35
[10]   Aberration-corrected Czerny-Turner imaging spectrometer with a wide spectral region [J].
Xue, Qingsheng ;
Wang, Shurong ;
Lu, Fengqin .
APPLIED OPTICS, 2009, 48 (01) :11-16