High-performance liquid chromatographic analysis of chlorhexidine in saliva after mouthrinsing

被引:17
作者
Tsuchiya, H [1 ]
Miyazaki, T [1 ]
Ohmoto, S [1 ]
机构
[1] Asahi Univ, Sch Dent, Dept Dent Pharmacol, Hozumi, Gifu 5010296, Japan
关键词
chlorhexidine; high-performance liquid chromatography; quantitative analysis; retention; saliva;
D O I
10.1159/000016510
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
A high-performance liquid chromatographic method was developed to quantify chlorhexidine in human saliva. After addition of an internal standard and acidic-deproteinization, saliva samples were chromatographed by the reversed-phase ion-pair system using pentanesulfonate as a counterion and the eluates were detected by a diode array detector. Under optimal conditions, the baseline separation of analytes was achieved within 4.5 min without interference from components in saliva matrix. The method showed high selectivity to salivary chlorhexidine, quantitative range (50.0 ng/ml-50.0 mu g/ml), recovery (96.0-97.8%) and analytical precision (intra- and interassay CVs within 0.41%), permitting the effective application to both saliva and aqueous solutions. Chlorhexidine was found in saliva at microgram per milliliter levels for at least 8 h after mouthrinsing with 10 ml of an aqueous solution of chlorhexidine (0.1 or 1.0 mg/ml) for 1 min, The concentrations of salivary chlorhexidine were reduced by ingesting the acidic beverage and food, but not by ingesting the neutral beverage. The adsorption experiments in artificial saliva revealed that chlorhexidine was significantly adsorbed to saliva-coated hydroxyapatite, buccal epithelial cells and mucin. The proposed method will be a useful tool to study salivary chlorhexidine after oral application and its retention in the oral cavity.
引用
收藏
页码:156 / 163
页数:8
相关论文
共 50 条
[21]   Column-switching high-performance liquid chromatography of ofloxacin in human saliva and correlation of ofloxacin level in saliva and serum [J].
Ohkubo, T ;
Suno, M ;
Kudo, M ;
Uno, T ;
Sugawara, K .
THERAPEUTIC DRUG MONITORING, 1996, 18 (05) :598-603
[22]   A HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE QUANTITATIVE ENANTIOSELECTIVE ANALYSIS OF MEFLOQUINE STEREOISOMERS [J].
QIU, YH ;
KITAMURA, S ;
GUILLORY, JK .
PHARMACEUTICAL RESEARCH, 1992, 9 (12) :1640-1643
[23]   BENZOYLATION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETECTION OF SOLASODINE [J].
DREWES, FE ;
VANSTADEN, J ;
DREWES, SE .
PHYTOCHEMICAL ANALYSIS, 1992, 3 (02) :85-87
[24]   A High-Performance Liquid Chromatographic Method for the Determination of Meropenem in Serum [J].
Dincel, Demet ;
Sagirli, Olcay ;
Topcu, Gulacti .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2020, 58 (02) :144-150
[25]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION OF HISTIDINE OLIGOPEPTIDES [J].
UEDA, J ;
HANAKI, A ;
YASUHARA, T ;
NAKAJIMA, T ;
SOFUKU, S .
ANALYTICAL SCIENCES, 1992, 8 (04) :487-489
[26]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF AMBROXOL IN PHARMACEUTICALS [J].
BRIZZI, V ;
PASETTI, U .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (01) :107-109
[27]   A high-performance liquid chromatographic method for determination of praziquantel in plasma [J].
Hanpitakpong, W ;
Banmairuroi, V ;
Kamanikom, B ;
Choemung, A ;
Na-Bangchang, K .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 36 (04) :871-876
[28]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OFLOXAXIN IN SERUM [J].
OHKUBO, T ;
KUDO, M ;
SUGAWARA, K .
ANALYTICAL SCIENCES, 1991, 7 (05) :741-743
[29]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF METHOTREXATE IN PLASMA [J].
ALKAYSI, HN ;
GHARAIBEH, AM ;
SALEM, MAS .
THERAPEUTIC DRUG MONITORING, 1990, 12 (02) :191-194
[30]   Quantitative Analysis and Chromatographic Fingerprinting of WeiChangShu Tablet Using High-Performance Liquid Chromatography Combined with Chemometric Methods [J].
Weiwei Zhao ;
Fengrun Sun ;
Junfeng Cui ;
Yan Ren ;
Mengan Liu ;
Feng Zhao .
Chromatographia, 2018, 81 :1293-1304