Cone-beam imaging with tilted rotation axis: Method and performance evaluation

被引:15
作者
Zhao, Chumin [1 ]
Herbst, Magdalena [2 ]
Vogt, Sebastian [2 ]
Ritschl, Ludwig [2 ]
Kappler, Steffen [2 ]
Siewerdsen, Jeffrey H. [1 ,3 ]
Zbijewski, Wojciech [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Siemens Healthineers, D-91301 Forchheim, Germany
[3] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21287 USA
关键词
cervical spine; cone-beam CT; extremity; robotic x-ray systems; scan orbit optimization; tilted orbits; COMPUTED-TOMOGRAPHY CBCT; CT; SYSTEM; RECONSTRUCTION; DETECTOR; CALIBRATION; MULTISLICE; QUALITY; SCANNER; DESIGN;
D O I
10.1002/mp.14209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The recently introduced robotic x-ray systems provide the flexibility to acquire cone-beam computed tomography (CBCT) data using customized, application-specific source-detector trajectories. We exploit this capability to mitigate the effects of x-ray scatter and noise in CBCT imaging of weight-bearing foot and cervical spine (C-spine) using scan orbits with a tilted rotation axis. Methods We used an advanced CBCT simulator implementing accurate models of x-ray scatter, primary attenuation, and noise to investigate the effects of the orbital tilt angle in upright foot and C-spine imaging. The system model was parameterized using a laboratory version of a three-dimensional (3D) robotic x-ray system (Multitom RAX, Siemens Healthineers). We considered a generalized tilted axis scan configuration, where the detector remained parallel to patient's long body axis during the acquisition, but the elevation of source and detector was changing. A modified Feldkamp-Davis-Kress (FDK) algorithm was developed for reconstruction in this configuration, which departs from the FDK assumption of a detector that is perpendicular to the scan plane. The simulated foot scans involved source-detector distance (SDD) of 1386 mm, orbital tilt angles ranging 10 degrees to 40 degrees, and 400 views at 1 mAs/view and 0.5 degrees increment; the C-spine scans involved -25 degrees to -45 degrees tilt angles, SDD of 1090 mm, and 202 views at 1.3 mAs and 1 degrees increment The imaging performance was assessed by projection-domain measurements of the scatter-to-primary ratio (SPR) and by reconstruction-domain measurements of contrast, noise and generalized contrast-to-noise ratio (gCNR, accounting for both image noise and background nonuniformity) of the metatarsals (foot imaging) and cervical vertebrae (spine imaging). The effects of scatter correction were also compared for horizontal and tilted scans using an ideal Monte Carlo (MC)-based scatter correction and a frame-by-frame mean scatter correction. Results The proposed modified FDK, involving projection resampling, mitigated streak artifacts caused by the misalignment between the filtering direction and the detector rows. For foot imaging (no grids), an optimized 20 degrees tilted orbit reduced the maximum SPR from similar to 1.5 in a horizontal scan to <0.5. The gCNR of the second metatarsal was enhanced twofold compared to a horizontal orbit. For the C-spine (with vertical grids), imaging with a tilted orbit avoided highly attenuating x-ray paths through the lower cervical vertebrae and shoulders. A -35 degrees tilted orbit yielded improved image quality and visualization of the lower cervical spine: the SPR of lower cervical vertebrae was reduced from similar to 10 (horizontal orbit) to <6 (tilted orbit), and the gCNR for C5-C7 increased by a factor of 2. Furthermore, tilted orbits showed potential benefits over horizontal orbits by enabling scatter correction with a simple frame-by-frame mean correction without substantial increase in noise-induced artifacts after the correction. Conclusions Tilted scan trajectories, enabled by the emerging robotic x-ray system technology, were optimized for CBCT imaging of foot and cervical spine using an advanced simulation framework. The results demonstrated the potential advantages of tilted axis orbits in mitigation of scatter artifacts and improving contrast-to-noise ratio in CBCT reconstructions.
引用
收藏
页码:3305 / 3320
页数:16
相关论文
共 52 条
  • [11] X-RAY TRANSMISSION FORMULA FOR ANTISCATTER GRIDS
    DAY, GJ
    DANCE, DR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12) : 1429 - 1433
  • [12] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [13] Twin robotic x-ray system for 2D radiographic and 3D cone-beam CT imaging
    Fieselmann, Andreas
    Steinbrener, Jan
    Jerebko, Anna K.
    Voigt, Johannes M.
    Scholz, Rosemarie
    Ritschl, Ludwig
    Mertelmeier, Thomas
    [J]. MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [14] Frank E.D., 2013, Merrill's Atlas of Radiographic Positioning and Procedures-E-Book
  • [15] Herbst M, 2019, MED IMAGING PHYS MED, V57
  • [16] Cone-beam CT with flat-panel-detector digital angiography system: Early experience in abdominal interventional procedures
    Hirota, Shozo
    Nakao, Norio
    Yamamoto, Satoshi
    Kobayashi, Kaoru
    Maeda, Hiroaki
    Ishikura, Reiichi
    Miura, Koui
    Sakamoto, Kiyoshi
    Ueda, Ken
    Baba, Rika
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (06) : 1034 - 1038
  • [17] A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance
    Hua, Chiaho
    Yao, Weiguang
    Kidani, Takao
    Tomida, Kazuo
    Ozawa, Saori
    Nishimura, Takenori
    Fujisawa, Tatsuya
    Shinagawa, Ryousuke
    Merchant, Thomas E.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1079)
  • [18] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [19] Generalized multi-dimensional adaptive filtering for conventional and spiral single-slice, multi-slice, and cone-beam CT
    Kachelriess, M
    Watzke, O
    Kalender, WA
    [J]. MEDICAL PHYSICS, 2001, 28 (04) : 475 - 490
  • [20] Dedicated breast CT: Initial clinical experience
    Lindfors, Karen K.
    Boone, John M.
    Nelson, Thomas R.
    Yang, Kai
    Kwan, Alexander L. C.
    Miller, DeWitt F.
    [J]. RADIOLOGY, 2008, 246 (03) : 725 - 733