Cone-beam imaging with tilted rotation axis: Method and performance evaluation

被引:15
作者
Zhao, Chumin [1 ]
Herbst, Magdalena [2 ]
Vogt, Sebastian [2 ]
Ritschl, Ludwig [2 ]
Kappler, Steffen [2 ]
Siewerdsen, Jeffrey H. [1 ,3 ]
Zbijewski, Wojciech [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Siemens Healthineers, D-91301 Forchheim, Germany
[3] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21287 USA
关键词
cervical spine; cone-beam CT; extremity; robotic x-ray systems; scan orbit optimization; tilted orbits; COMPUTED-TOMOGRAPHY CBCT; CT; SYSTEM; RECONSTRUCTION; DETECTOR; CALIBRATION; MULTISLICE; QUALITY; SCANNER; DESIGN;
D O I
10.1002/mp.14209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The recently introduced robotic x-ray systems provide the flexibility to acquire cone-beam computed tomography (CBCT) data using customized, application-specific source-detector trajectories. We exploit this capability to mitigate the effects of x-ray scatter and noise in CBCT imaging of weight-bearing foot and cervical spine (C-spine) using scan orbits with a tilted rotation axis. Methods We used an advanced CBCT simulator implementing accurate models of x-ray scatter, primary attenuation, and noise to investigate the effects of the orbital tilt angle in upright foot and C-spine imaging. The system model was parameterized using a laboratory version of a three-dimensional (3D) robotic x-ray system (Multitom RAX, Siemens Healthineers). We considered a generalized tilted axis scan configuration, where the detector remained parallel to patient's long body axis during the acquisition, but the elevation of source and detector was changing. A modified Feldkamp-Davis-Kress (FDK) algorithm was developed for reconstruction in this configuration, which departs from the FDK assumption of a detector that is perpendicular to the scan plane. The simulated foot scans involved source-detector distance (SDD) of 1386 mm, orbital tilt angles ranging 10 degrees to 40 degrees, and 400 views at 1 mAs/view and 0.5 degrees increment; the C-spine scans involved -25 degrees to -45 degrees tilt angles, SDD of 1090 mm, and 202 views at 1.3 mAs and 1 degrees increment The imaging performance was assessed by projection-domain measurements of the scatter-to-primary ratio (SPR) and by reconstruction-domain measurements of contrast, noise and generalized contrast-to-noise ratio (gCNR, accounting for both image noise and background nonuniformity) of the metatarsals (foot imaging) and cervical vertebrae (spine imaging). The effects of scatter correction were also compared for horizontal and tilted scans using an ideal Monte Carlo (MC)-based scatter correction and a frame-by-frame mean scatter correction. Results The proposed modified FDK, involving projection resampling, mitigated streak artifacts caused by the misalignment between the filtering direction and the detector rows. For foot imaging (no grids), an optimized 20 degrees tilted orbit reduced the maximum SPR from similar to 1.5 in a horizontal scan to <0.5. The gCNR of the second metatarsal was enhanced twofold compared to a horizontal orbit. For the C-spine (with vertical grids), imaging with a tilted orbit avoided highly attenuating x-ray paths through the lower cervical vertebrae and shoulders. A -35 degrees tilted orbit yielded improved image quality and visualization of the lower cervical spine: the SPR of lower cervical vertebrae was reduced from similar to 10 (horizontal orbit) to <6 (tilted orbit), and the gCNR for C5-C7 increased by a factor of 2. Furthermore, tilted orbits showed potential benefits over horizontal orbits by enabling scatter correction with a simple frame-by-frame mean correction without substantial increase in noise-induced artifacts after the correction. Conclusions Tilted scan trajectories, enabled by the emerging robotic x-ray system technology, were optimized for CBCT imaging of foot and cervical spine using an advanced simulation framework. The results demonstrated the potential advantages of tilted axis orbits in mitigation of scatter artifacts and improving contrast-to-noise ratio in CBCT reconstructions.
引用
收藏
页码:3305 / 3320
页数:16
相关论文
共 52 条
  • [1] Cone Beam Computed Tomography (CBCT) in the Field of Interventional Oncology of the Liver
    Bapst, Blanche
    Lagadec, Matthieu
    Breguet, Romain
    Vilgrain, Valerie
    Ronot, Maxime
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2016, 39 (01) : 8 - 20
  • [2] Initial Assessment of a Prototype 3D Cone-Beam Computed Tomography System for Imaging of the Lumbar Spine, Evaluating Human Cadaveric Specimens in the Upright Position
    Benz, Robyn Melanie
    Harder, Dorothee
    Amsler, Felix
    Voigt, Johannes
    Fieselmann, Andreas
    Falkowski, Anna Lucja
    Stieltjes, Bram
    Hirschmann, Anna
    [J]. INVESTIGATIVE RADIOLOGY, 2018, 53 (12) : 714 - 719
  • [3] Initial evaluation of image performance of a 3-D x-ray system: phantom-based comparison of 3-D tomography with conventional computed tomography
    Benz, Robyn Melanie
    Garcia, Meritxell Alzamora
    Amsler, Felix
    Voigt, Johannes
    Fieselmann, Andreas
    Falkowski, Anna Lucja
    Stieltjes, Bram
    Hirschmann, Anna
    [J]. JOURNAL OF MEDICAL IMAGING, 2018, 5 (01)
  • [4] Boas FE., 2012, Imaging Med, V4, P229, DOI [10.2217/iim.12.13, DOI 10.2217/IIM.12.13]
  • [5] Computed tomography for imaging the breast
    Boone, John M.
    Kwan, Alex L. C.
    Yang, Kai
    Burkett, George W.
    Lindfors, Karen K.
    Nelson, Thomas R.
    [J]. JOURNAL OF MAMMARY GLAND BIOLOGY AND NEOPLASIA, 2006, 11 (02) : 103 - 111
  • [6] A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy
    Boswell, SA
    Jeraj, R
    Ruchala, KJ
    Olivera, GH
    Jaradat, HA
    James, JA
    Gutierrez, A
    Pearson, D
    Frank, G
    Mackie, TR
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1630 - 1639
  • [7] Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities
    Cao, Qian
    Sisniega, Alejandro
    Brehler, Michael
    Stayman, J. Webster
    Yorkston, John
    Siewerdsen, Jeffrey H.
    Zbijewski, Wojciech
    [J]. MEDICAL PHYSICS, 2018, 45 (01) : 114 - 130
  • [8] Task-driven source-detector trajectories in cone-beam computed tomography: II. Application to neuroradiology
    Capostagno, Sarah
    Stayman, J. Webster
    Jacobson, Matthew
    Ehtiati, Tina
    Weiss, Clifford R.
    Siewerdsen, Jeffrey H.
    [J]. JOURNAL OF MEDICAL IMAGING, 2019, 6 (02)
  • [9] Accurate technique for complete geometric calibration of cone-beam computed tomography systems
    Cho, YB
    Moseley, DJ
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2005, 32 (04) : 968 - 983
  • [10] Accuracy of a new intraoperative cone beam CT imaging technique (Artis zeego II) compared to postoperative CT scan for assessment of pedicle screws placement and breaches detection
    Cordemans, Virginie
    Kaminski, Ludovic
    Banse, Xavier
    Francq, Bernard G.
    Cartiaux, Olivier
    [J]. EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2906 - 2916