Analytical efficiency curve for coaxial germanium detectors

被引:8
作者
Svec, A. [1 ]
机构
[1] Slovak Inst Metrol, Ctr Ionizing Radiat, Bratislava 84255, Slovakia
关键词
radionuclide activity; gamma-ray spectrometry; efficiency curve;
D O I
10.1016/j.apradiso.2008.02.070
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A new semi-empirical analytical expression of gamma-ray spectrometer efficiency curves has been proposed and a method for their parameters determination has been developed. The complete curve consists of two parts, first being the basic detector characterization curve of low number of parameters and second one being a corrective curve, which adjusts the final results to the required and achievable accuracy. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:786 / 791
页数:6
相关论文
共 49 条
  • [21] A new and general efficiency calibration methodology for Ge detectors using LabSOCS
    Barba-Lobo, A.
    RADIATION MEASUREMENTS, 2025, 181
  • [22] Monte Carlo calculation of the efficiency calibration curve and coincidence-summing corrections in low-level gamma-ray spectrometry using well-type HPGe detectors
    Laborie, JM
    Le Petit, G
    Abt, D
    Girard, M
    APPLIED RADIATION AND ISOTOPES, 2000, 53 (1-2) : 57 - 62
  • [23] MCNP5 modelling of HPGe detectors for efficiency evaluation in -ray spectrometry
    Maurotto, A.
    Rizzo, S.
    Tomarchio, E.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2009, 164 (5-6): : 302 - 306
  • [24] The use of efficiency curve approach for predicting the classification performance of teetered bed separator
    Ozcan, Ozgur
    SEPARATION SCIENCE AND TECHNOLOGY, 2021, 56 (15) : 2667 - 2679
  • [25] CREPT-MCNP code for efficiency calibration of HPGe detectors with the representative point method
    Saegusa, Jun
    APPLIED RADIATION AND ISOTOPES, 2008, 66 (6-7) : 774 - 779
  • [26] Accurate calculation of total efficiency of Ge well-type detectors suitable for efficiency calibration using common standard sources
    Sima, O
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 450 (01) : 98 - 108
  • [27] Study of equivalent efficiency calibration method for radioactive gas source measured on HPGe detectors
    Yu, Gongshuo
    Li, Xuesong
    Bai, Tao
    He, Xiaobing
    Fan, Jinlong
    Shang, Jianbo
    Shi, Quanlin
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2020, 57 (05) : 599 - 606
  • [28] Efficiency transfer in the calibration of a coaxial p-type HpGe detector using the Monte Carlo method
    Vargas, MJ
    Díaz, NC
    Sánchez, DP
    APPLIED RADIATION AND ISOTOPES, 2003, 58 (06) : 707 - 712
  • [29] Analytical approach for radioactivity correlation of disc sources with HPGe detector efficiency
    Challan, Mohsen B.
    El-Taher, A.
    APPLIED RADIATION AND ISOTOPES, 2014, 85 : 23 - 27
  • [30] Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique
    Gouda, M. M.
    Hamzawy, A.
    Badawi, M. S.
    El-Khatib, A. M.
    Thabet, A. A.
    Abbas, M. I.
    INDIAN JOURNAL OF PHYSICS, 2016, 90 (02) : 201 - 210