A Monolithically Integrated Phase-Sensitive Optical Sensor for Frequency-Domain NIR Spectroscopy

被引:23
作者
Yun, Ruida [1 ]
Joyner, Valencia M. [1 ]
机构
[1] Tufts Univ, Dept Elect & Comp Engn, Medford, MA 02155 USA
基金
美国国家科学基金会;
关键词
Amplitude detector; analog signal processing; CMOS; near-infrared spectroscopy; optical sensor; phase detector; transimpedance amplifier; NEAR-INFRARED SPECTROSCOPY; TOMOGRAPHY; AMPLITUDE; DETECTOR; SYSTEM; TIME;
D O I
10.1109/JSEN.2010.2044502
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents design and measurement results of a fully integrated optical sensor for phase and amplitude detection of RF modulated optical signals up to 110 MHz in the near-infrared (NIR) region (650-850 nm) for use in frequency-domain spectroscopy instruments. The sensor consists of an NIR-sensitive photodetector monolithically integrated with a front-end analog amplifier and signal processing circuitry for amplitude and phase detection in an unmodified complementary metal oxide semiconductor (CMOS) process. A high-gain, low-noise differential transimpedance amplifier (TIA) is implemented to amplify the photocurrent signal. Amplitude and phase resolution are evaluated with a 690 nm laser diode modulated at 100 MHz. The amplitude response exhibits 2.2 mV/mu W resolution with 0.4% linearity. The measured amplitude output noise is 72 mu V. The proposed phase detector detects 0 degrees-360 degrees phase difference with a measured average phase resolution of 4.8 mV/degree and 255 mu V output noise. The sensor is implemented in a 180 nm CMOS technology and consumes 23.4 mW from a 1.8 V supply voltage.
引用
收藏
页码:1234 / 1242
页数:9
相关论文
共 33 条
[1]  
[Anonymous], 1993, AM NAT STAND SAF US
[2]  
Baker R. J., 2007, CMOS CIRCUIT DESIGN
[3]   Blood flow and oxygen consumption with near-infrared spectroscopy and venous occlusion: spatial maps and the effect of time and pressure of inflation [J].
Casavola, C ;
Paunescu, LA ;
Fantini, S ;
Gratton, E .
JOURNAL OF BIOMEDICAL OPTICS, 2000, 5 (03) :269-276
[4]   Integrated color pixels in 0.18-μm complementary metal oxide semiconductor technology [J].
Catrysse, PB ;
Wandell, BA .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2003, 20 (12) :2293-2306
[5]   Phase measurement of light absorption and scatter in human tissue [J].
Chance, B ;
Cope, M ;
Gratton, E ;
Ramanujam, N ;
Tromberg, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (10) :3457-3481
[6]   CMOS-based phase fluorometric oxygen sensor system [J].
Chodavarapu, Vamsy P. ;
Shubin, Daniil O. ;
Bukowski, Rachel M. ;
Titus, Albert H. ;
Cartwright, Alexander N. ;
Bright, Frank V. .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2007, 54 (01) :111-118
[7]   Multichannel time-resolved tissue oximeter for functional imaging of the brain [J].
Contini, D ;
Torricelli, A ;
Pifferi, A ;
Spinelli, L ;
Taroni, P ;
Quaresima, V ;
Ferrari, M ;
Cubeddu, R .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2006, 55 (01) :85-90
[8]   Three-dimensional diffuse optical tomography in the parallel plane transmission geometry: Evaluation of a hybrid frequency domain/continuous wave clinical system for breast imaging [J].
Culver, JP ;
Choe, R ;
Holboke, MJ ;
Zubkov, L ;
Durduran, T ;
Slemp, A ;
Ntziachristos, V ;
Chance, B ;
Yodh, AG .
MEDICAL PHYSICS, 2003, 30 (02) :235-247
[9]  
Fantini S, 2000, PROCEEDINGS OF INTER-INSTITUTE WORKSHOP ON IN VIVO OPTICAL IMAGING AT THE NIH, P111
[10]   Frequency-domain techniques enhance optical mammography: Initial clinical results [J].
Franceschini, MA ;
Moesta, KT ;
Fantini, S ;
Gaida, G ;
Gratton, E ;
Jess, H ;
Mantulin, WW ;
Seeber, M ;
Schlag, PM ;
Kaschke, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1997, 94 (12) :6468-6473