Gadolinium hydroxide coprecipitation system for the separation-preconcentration of some heavy metals

被引:26
作者
Soylak, Mustafa [1 ]
Balgunes, Hikmet [1 ]
机构
[1] Erciyes Univ, Fac Art & Sci, Dept Chem, TR-38039 Kayseri, Turkey
关键词
preconcentration; coprecipitation; gadolinium hydroxide; heavy metal;
D O I
10.1016/j.jhazmat.2007.11.096
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A simple and new procedure for the determination of trace amounts of lead(II), cobalt(II), manganese(II) and copper(II) is described, that combines atomic absorption spectrometry-gadolinium hydroxide coprecipitation. One milliliter of 1% gadolinium(III) solution was added to each sample; the pH was then adjusted to 11 in order to collect trace heavy metals on gadolinium(III) hydroxide. The precipitate was separated by centrifugation and dissolved in 1 mL of 1 mol L-1 HNO3. The influences of analytical parameters including amount of gadolinium(III), sample volume, etc. were investigated on the recoveries of analyte ions. The effects of concomitant ions were also examined. The recoveries of the analyte ions were greater than 95. The detection limits for the analyte elements based on 3 sigma (n = 20) were in the range of 0.52-12.0 mu g L-1. The method was applied to the determination of analytes in real samples and good results were obtained (relative standard deviations <10%, recoveries >95%). (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:595 / 600
页数:6
相关论文
共 34 条
[1]   SIMULTANEOUS MULTIELEMENT DETERMINATION OF TRACE-METALS USING 10 ML OF SEAWATER BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH GALLIUM COPRECIPITATION AND MICROSAMPLING TECHNIQUE [J].
AKAGI, T ;
HARAGUCHI, H .
ANALYTICAL CHEMISTRY, 1990, 62 (01) :81-85
[2]   Determination of cadmium in spring water by graphite-furnace atomic absorption spectrometry after coprecipitation with ytterbium hydroxide [J].
Atsumi, K ;
Minami, T ;
Ueda, J .
ANALYTICAL SCIENCES, 2005, 21 (06) :647-649
[3]   Removal of cadmium (II) from aqueous solution using pure smectite and Lewatite S 100: the effect of time and metal concentration [J].
Bedoui, K. ;
Bekri-Abbes, I. ;
Srasra, E. .
DESALINATION, 2008, 223 (1-3) :269-273
[4]   Near infrared spectroscopy:: A novel technique for classifying and characterizing polysulfone membranes [J].
Conesa, A. ;
Gumi, T. ;
Coello, J. ;
Palet, C. .
JOURNAL OF MEMBRANE SCIENCE, 2007, 300 (1-2) :122-130
[5]  
Dalman Ö, 2007, TURK J CHEM, V31, P631
[6]  
DOGRU M, 2005, RECENT ADV MULTIDISC, P865
[7]   Coprecipitation of Cu(II), Ni(II), Fe(III), Cd(II), Pb(II), and Co(II) in wastewater, sediment, and metallic zinc samples with HMDTC-HMA for flame atomic absorption spectrometric determination [J].
Elçi, L ;
Soylak, M ;
Özcan, B .
ANALYTICAL LETTERS, 2003, 36 (05) :987-999
[8]   Applying magnesium hydroxide coprecipitation method for trace analysis to dialysis concentrate [J].
Elci, L ;
Saracoglu, S .
TALANTA, 1998, 46 (06) :1305-1310
[9]   Selected elements in fly agaric Amanita muscaria [J].
Falandysz, J. ;
Kunito, T. ;
Kubota, R. ;
Lipka, K. ;
Mazur, A. ;
Falandysz, Justyna J. ;
Tanabe, S. .
JOURNAL OF ENVIRONMENTAL SCIENCE AND HEALTH PART A-TOXIC/HAZARDOUS SUBSTANCES & ENVIRONMENTAL ENGINEERING, 2007, 42 (11) :1615-1623
[10]  
FAN J, TALANTA IN PRESS