Prostate cancer: 1.5 T endo-coil dynamic contrast-enhanced MRI and MR spectroscopy-correlation with prostate biopsy and prostatectomy histopathological data

被引:14
作者
Chabanova, Elizaveta [1 ]
Balslev, Ingegerd [2 ]
Logager, Vibeke [1 ]
Hansen, Alastair [2 ]
Jakobsen, Henrik [3 ]
Kromann-Andersen, Bjarne [3 ]
Norgaard, Nis [3 ]
Horn, Thomas [2 ]
Thomsen, Henrik S. [1 ]
机构
[1] Copenhagen Univ Hosp, Dept Diagnost Radiol, DK-2730 Herlev, Denmark
[2] Copenhagen Univ Hosp, Dept Pathol, DK-2730 Herlev, Denmark
[3] Copenhagen Univ Hosp, Dept Urol, DK-2730 Herlev, Denmark
关键词
Prostate cancer; DCEMRI; SCI; Prostatectomy; Histopathology; PERIPHERAL ZONE; LOCALIZATION; ANTIGEN; TRANSITION; DIAGNOSIS; ACCURACY; TISSUE; ARRAY;
D O I
10.1016/j.ejrad.2010.07.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate diagnostic accuracy of detection of prostate cancer by magnetic resonance: to evaluate the performance of T2WI, DCEMRI and CSI and to correlate the results with biopsy and radical prostatectomy histopathological data. Materials and methods: 43 patients, scheduled for radical prostatectomy, underwent prostate MR examination. Prostate cancer was identified by transrectal ultrasonographically (TRUS) guided sextant biopsy. MR examination was performed at 1.5T with an endorectal MR coil. Cancer localisation was performed on sextant-basis - for comparison between TRUS biopsy, MR techniques and histopathological findings on prostatectomy specimens. Results: Prostate cancer was identified in all 43 patients by combination of the three MR techniques. The detection of prostate cancer on sextant-basis showed sensitivity and specificity: 50% and 91% for TRUS, 72% and 55% for T2WI, 49% and 69% for DCEMRI, and 46% and 78% for CSI. Conclusion: T2WI, DCEMRI and CSI in combination can identify prostate cancer. Further development of MR technologies for these MR methods is necessary to improve the detection of the prostate cancer. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:292 / 296
页数:5
相关论文
共 20 条
[1]   Transition zone prostate cancers: Features, detection, localization, and staging at endorectal MR imaging [J].
Akin, Oguz ;
Sala, Evis ;
Moskowitz, Chaya S. ;
Kuroiwa, Kentaro ;
Ishill, Nicole M. ;
Pucar, Darko ;
Scardino, Peter T. ;
Hricak, Hedvig .
RADIOLOGY, 2006, 239 (03) :784-792
[2]  
Alzoni R, 2007, EUR J RADIOL, V63, P335
[3]   Prostate cancer detection: Comparison of T2-weighted imaging, diffusion-weighted imaging, proton magnetic resonance spectroscopic imaging, and the three techniques combined [J].
Chen, M. ;
Dang, H. -D. ;
Wang, J. -Y. ;
Zhou, C. ;
Li, S. -Y. ;
Wang, W. -C. ;
Zhao, W. -F. ;
Yang, Z. -H. ;
Zhong, C. -Y. ;
Li, G. -Z. .
ACTA RADIOLOGICA, 2008, 49 (05) :602-610
[4]   Diagnosis of prostate cancer in patients with an elevated prostate-specific antigen level: Role of endorectal MRI and MR spectroscopic Imaging [J].
Costouros, Nick G. ;
Coakley, Fergus V. ;
Westphalen, Antonio C. ;
Qayyum, Aliya ;
Yeh, Benjamin M. ;
Joe, Bonnie N. ;
Kurhanewicz, John .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (03) :812-816
[5]   Discrimination of prostate cancer from normal peripheral zone and central gland tissue by using dynamic contrast-enhanced MR imaging [J].
Engelbrecht, MR ;
Huisman, HJ ;
Laheij, RJF ;
Jager, GJ ;
van Leenders, GJLH ;
Hulsbergen-Van De Kaa, CA ;
de la Rosette, JJMCH ;
Blickman, JG ;
Barentsz, JO .
RADIOLOGY, 2003, 229 (01) :248-254
[6]   Prostate cancer localization with dynamic contrast-enhanced MR imaging and proton MR spectroscopic imaging [J].
Fuetterer, Jurgen J. ;
Heijmink, Stijn W. T. P. J. ;
Scheenen, Tom W. J. ;
Veltman, Jeroen ;
Huisman, Henkjan J. ;
Vos, Pieter ;
Hulsbergen-Van de Kaa, Christina A. ;
Witjes, J. Alfred ;
Krabbe, Paul F. M. ;
Heerschap, Arend ;
Barentsz, Jelle O. .
RADIOLOGY, 2006, 241 (02) :449-458
[7]   Prostate cancer: comparison of local staging accuracy of pelvic phased-array coil alone versus integrated endorectal-pelvic phased-array coils - Local staging accuracy of prostate cancer using endorectal coil MR imaging [J].
Futterer, Jurgen J. ;
Engelbrecht, Marc R. ;
Jager, Gerrit J. ;
Hartman, Robert P. ;
King, Bernard F. ;
Hulsbergen-Van de Kaa, Christina A. ;
Witjes, J. Alfred ;
Barentsz, Jelle O. .
EUROPEAN RADIOLOGY, 2007, 17 (04) :1055-1065
[8]   Prostate cancer:: Body-array versus endorectal coil MR imaging at 3 T -: Comparison of image quality, localization, and staging performance [J].
Heijmink, Stijn W. T. P. J. ;
Futterer, Jurgen J. ;
Hambrock, Thomas ;
Takahashi, Satoru ;
Scheenen, Tom W. J. ;
Huisman, Henkjan J. ;
Hulsbergen-Van de Kaa, Christina A. ;
Knipscheer, Ben C. ;
Kiemeney, Lambertus A. L. M. ;
Witjes, J. Alfred ;
Barentsz, Jelle O. .
RADIOLOGY, 2007, 244 (01) :184-195
[9]   Imaging prostate cancer: A multidisciplinary perspective [J].
Hricak, Hedvig ;
Choyke, Peter L. ;
Eberhardt, Steven C. ;
Leibel, Steven A. ;
Scardino, Peter T. .
RADIOLOGY, 2007, 243 (01) :28-53
[10]  
Kenney J. F., 1962, Mathematics of Statistics, V3rd, P252