Risk Factors for Earlier Reexposure of Glaucoma Drainage Devices

被引:9
作者
Thompson, Atalie C. [1 ]
Manjunath, Varsha [1 ]
Muir, Kelly W. [1 ,2 ]
机构
[1] Duke Univ Hosp, Dept Ophthalmol, Durham, NC USA
[2] Durham VA Ctr Hlth Serv Res Primary Care, Durham, NC USA
关键词
glaucoma complications; glaucoma drainage device exposure; scleral patch graft; race; TRABECULECTOMY TVT; TUBE COVERAGE; FOLLOW-UP; EXPOSURE; OUTCOMES; IMPLANT; SURGERY; COMPLICATIONS; GRAFT;
D O I
10.1097/IJG.0000000000000821
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:The purpose of this study was to investigate factors associated with a second exposure of a glaucoma drainage device (GDD) following repair of an initial GDD exposure.Materials and Methods:This IRB-approved retrospective cohort study examined the incidence of a second exposure of a GDD following initial repair for exposure. Logistic regression was performed to assess the relationship between demographic and clinical characteristics and a second exposure of the GDD. Kaplan-Meier survival curves were plotted and Cox regression was performed to examine factors impacting the time to a second GDD exposure.Results:Ninety-four eyes of subjects that underwent initial revision for GDD exposure were reviewed. Approximately 44% (N=41/94) of subjects underwent surgical revision for a second exposure. Factors associated with reexposure in multivariate logistic regression included caucasian race (odds ratio, 2.99; P=0.02) and use of a nonscleral patch graft (odds ratio, 2.93; P=0.019). Time from revision of the initial exposure to reexposure was significantly shorter for those with a nonscleral patch graft (hazard ratio, 2.23; P=0.01) and caucasian race (hazard ratio, 2.08; P=0.04).Conclusions:Caucasian race and use of a nonscleral patch graft during revision surgery was associated with a higher risk of experiencing a sooner reexposure of the GDD following revision of an initial exposure. Future studies should examine whether particular graft materials increase the risk of GDD reexposure.
引用
收藏
页码:1155 / 1160
页数:6
相关论文
共 25 条
[1]   Risk factors of implant exposure outside the conjunctiva after Ahmed glaucoma valve implantation [J].
Byun, Yong Soo ;
Lee, Na Young ;
Park, Chan Kee .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2009, 53 (02) :114-119
[2]   Risk factors for tube exposure as a late complication of glaucoma drainage implant surgery [J].
Chaku, Meenakshi ;
Netland, Peter A. ;
Ishida, Kyoko ;
Rhee, Douglas J. .
CLINICAL OPHTHALMOLOGY, 2016, 10 :547-553
[3]   Late endophthalmitis associated with glaucoma drainage implants [J].
Gedde, SJ ;
Scott, IU ;
Tabandeh, H ;
Luu, KKM ;
Budenz, DL ;
Greenfield, DS ;
Flynn, HW .
OPHTHALMOLOGY, 2001, 108 (07) :1323-1327
[4]   Postoperative Complications in the Tube Versus Trabeculectomy (TVT) Study During Five Years of Follow-up [J].
Gedde, Steven J. ;
Herndon, Leon W. ;
Brandt, James D. ;
Budenz, Donald L. ;
Feuer, William J. ;
Schiffman, Joyce C. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (05) :804-814
[5]   Treatment Outcomes in the Tube Versus Trabeculectomy (TVT) Study After Five Years of Follow-up [J].
Gedde, Steven J. ;
Schiffman, Joyce C. ;
Feuer, William J. ;
Herndon, Leon W. ;
Brandt, James D. ;
Budenz, Donald L. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (05) :789-803
[6]   Conjunctival Complications Related to Ahmed Glaucoma Valve Insertion [J].
Geffen, Noa ;
Buys, Yvonne M. ;
Smith, Michael ;
Anraku, Ayako ;
Alasbali, Tariq ;
Rachmiel, Ronny ;
Trope, Graham E. .
JOURNAL OF GLAUCOMA, 2014, 23 (02) :109-114
[7]   The DEDUCE Guided Query tool: Providing simplified access to clinical data for research and quality improvement [J].
Horvath, Monica M. ;
Winfield, Stephanie ;
Evans, Steve ;
Slopek, Steve ;
Shang, Howard ;
Ferranti, Jeffrey .
JOURNAL OF BIOMEDICAL INFORMATICS, 2011, 44 (02) :266-276
[8]   Aqueous Shunt Exposure: A Retrospective Review of Repair Outcome [J].
Huddleston, Stephen M. ;
Feldman, Robert M. ;
Budenz, Donald L. ;
Bell, Nicholas P. ;
Lee, David A. ;
Chuang, Alice Z. ;
Mankiewicz, Kimberly A. ;
Koval, Michael S. ;
Elizabeth Truong ;
Moraczewski, Alexei .
JOURNAL OF GLAUCOMA, 2013, 22 (06) :433-438
[9]   Risk Factors for Tube Shunt Exposure: A Matched Case-Control Study [J].
Koval, Michael S. ;
El Sayyad, Fouad F. ;
Bell, Nicholas P. ;
Chuang, Alice Z. ;
Lee, David A. ;
Hypes, Stephen M. ;
Grover, Davinder S. ;
Baker, Laura A. ;
Huddleston, Stephen M. ;
Budenz, Donald L. ;
Feldman, Robert M. .
JOURNAL OF OPHTHALMOLOGY, 2013, 2013 :1-5
[10]   Intermediate-term outcomes of 350-mm2 Baerveldt glaucoma implants [J].
Krishna, R ;
Godfrey, DG ;
Budenz, DL ;
Escalona-Camaaño, E ;
Gedde, SJ ;
Greenfield, DS ;
Feuer, W ;
Scott, IU .
OPHTHALMOLOGY, 2001, 108 (03) :621-626