High resolution intravascular MRI and MRS by using a catheter receiver coil

被引:140
作者
Atalar, E [1 ]
Bottomley, PA [1 ]
Ocali, O [1 ]
Correia, LCL [1 ]
Kelemen, MD [1 ]
Lima, JAC [1 ]
Zerhouni, EA [1 ]
机构
[1] JOHNS HOPKINS UNIV HOSP,SCH MED,DIV CARDIOL,BALTIMORE,MD 21287
关键词
intravascular MRI; MR catheter coil; atherosclerosis; MRI of arterial wall;
D O I
10.1002/mrm.1910360415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Potentially important diagnostic information about atherosclerosis can be obtained by using magnetic resonance imaging and spectroscopy techniques, Because critical vessels such as the aorta, coronary arteries, and renal arteries are not near the surface of the body, surface coils are not adequate to increase the data quality to desired levels, A few catheter MR receiver coil designs have been proposed for imaging the walls of large blood vessels such as the aorta. These coils have limited longitudinal coverage and they are too thick to be placed into small vessels. A flexible, long and narrow receiver coil that can be placed on the tip of a catheter and will enable multi-slice high resolution imaging of small vessels has been developed, The authors describe the theory of the coil design technique, derive formulae for the signal-to-noise ratio characteristics of the coil, and show examples of high resolution cross-sectional images from isolated human aortas acquired by using this catheter coil. In addition, high resolution in vivo rabbit aorta images were obtained as well as a set of spatially resolved chemical shift spectra from a dog circumflex coronary artery.
引用
收藏
页码:596 / 605
页数:10
相关论文
共 28 条
[1]  
*AIN PLAST INC, 1993, CAT 93
[2]   MAGNETIC-RESONANCE-IMAGING OF LIPID DEPOSITS IN HUMAN ATHEROMA VIA A STIMULATED-ECHO DIFFUSION-WEIGHTED TECHNIQUE [J].
ALTBACH, MI ;
MATTINGLY, MA ;
BROWN, MF ;
GMITRO, AF .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :319-326
[3]   EVALUATION OF ATHEROSCLEROTIC LESIONS USING NMR MICROIMAGING [J].
ASDENTE, M ;
PAVESI, L ;
ORESTE, PL ;
COLOMBO, A ;
KUHN, W ;
TREMOLI, E .
ATHEROSCLEROSIS, 1990, 80 (03) :243-253
[4]  
CHIABRERA A, 1985, NATO ASI SERIES A, V97
[5]   REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE [J].
DUMOULIN, CL ;
SOUZA, SP ;
DARROW, RD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :411-415
[6]   ELECTRONIC DECOUPLING OF SURFACE-COIL RECEIVERS FOR NMR IMAGING AND SPECTROSCOPY [J].
EDELSTEIN, WA ;
HARDY, CJ ;
MUELLER, OM .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (01) :156-161
[7]  
HUA J, 1990, P SMRM 9 ANN M, P1335
[8]   INTRAVASCULAR (CATHETER) NMR RECEIVER PROBE - PRELIMINARY DESIGN ANALYSIS AND APPLICATION TO CANINE ILIOFEMORAL IMAGING [J].
HURST, GC ;
HUA, JM ;
DUERK, JL ;
COHEN, AM .
MAGNETIC RESONANCE IN MEDICINE, 1992, 24 (02) :343-357
[9]  
Kandarpa K, 1993, J Vasc Interv Radiol, V4, P419, DOI 10.1016/S1051-0443(93)71891-6
[10]   INVIVO P-31 NUCLEAR MAGNETIC-RESONANCE MEASUREMENTS IN CANINE HEART USING A CATHETER-COIL [J].
KANTOR, HL ;
BRIGGS, RW ;
BALABAN, RS .
CIRCULATION RESEARCH, 1984, 55 (02) :261-266