The use of a gold disc microelectrode for the determination of copper in human sweat

被引:19
作者
de Souza, Ana P. R. [1 ]
Lima, Alex S. [1 ]
Salles, Maiara O. [1 ]
Nascimento, Angerson N. [1 ]
Bertotti, Mauro [1 ]
机构
[1] Univ Sao Paulo, Inst Quim, BR-05508900 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
Microelectrode; Sweat; Copper; Stripping analysis; ANODIC-STRIPPING VOLTAMMETRY; CYSTIC-FIBROSIS; TRACE-METALS; MERCURY; ELECTRODE; ZINC; LEAD; DIAGNOSIS; EXERCISE; LACTATE;
D O I
10.1016/j.talanta.2010.09.001
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel approach of using a gold disc microelectrode to analyze sweat samples for copper ions by anodic square wave stripping voltammetry (SW stripping voltammetry) is described Sweat was collected from the lower back of four subjects after physical exercise and the sample volume required for the determinations was 100 mu L. Under the optimized conditions the calibration plot was linear over the range 1-100 mu mol L-1 Cu(II) with a limit of detection of 0 25 mu mol L-1 The precision was evaluated by carrying out five replicate measurements in a 1 mu mol L-1 Cu(II) solution and the standard deviation was found to be 1 5% Measurements were performed by inserting the microelectrode into sweat drops and Cu(II) concentrations in the analyzed samples ranged from 09 to 28 mu mol L-1 Values obtained by the proposed voltammetric method agreed well with those found using graphite furnace atomic absorption spectroscopy (GFAAS) (C) 2010 Elsevier B V All rights reserved
引用
收藏
页码:167 / 170
页数:4
相关论文
共 34 条
[1]  
ABCD A, 1987, ANALYST, V112, P199
[2]  
[Anonymous], 1985, HDB CHEM EQUILIBRIA
[3]  
Baldo MA, 1997, ANALYST, V122, P1, DOI 10.1039/a605370j
[4]   Sweat testing for cystic fibrosis: standards of performance in Australasia [J].
Coakley, John ;
Scott, Sue ;
Mackay, Richard ;
Greaves, Ronda ;
Jolly, Lisa ;
Massie, John ;
Mishra, Avantika ;
Bransden, Anna ;
Doery, James C. G. ;
Chiriano, Angela ;
Robins, Heather .
ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 :332-337
[5]  
COHN JR, 1978, ANN CLIN LAB SCI, V8, P270
[6]   Development of an anodic stripping voltammetric assay, using a disposable mercury-free screen-printed carbon electrode, for the determination of zinc in human sweat [J].
Crew, A. ;
Cowell, D. C. ;
Hart, J. P. .
TALANTA, 2008, 75 (05) :1221-1226
[7]   Sweat as an Alternative Matrix for Amphetamines and MethylenEdioxy Derivatives Analysis [J].
De Martinis, Bruno S. .
CURRENT PHARMACEUTICAL ANALYSIS, 2008, 4 (04) :274-278
[8]   Identification of promising antigenic components in latent fingermark residues [J].
Drapel, Valerie ;
Becue, Andy ;
Champod, Christophe ;
Margot, Pierre .
FORENSIC SCIENCE INTERNATIONAL, 2009, 184 (1-3) :47-53
[9]   Sweat lactate response between males with high and low aerobic fitness [J].
Green, JM ;
Pritchett, RC ;
Crews, TR ;
McLester, JR ;
Tucker, DC .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2004, 91 (01) :1-6
[10]   Generation of micrometer-sized patterns for microanalytical applications using a laser direct-write method and microcontact printing [J].
Grzybowski, BA ;
Haag, R ;
Bowden, N ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (22) :4645-4652