Translational and Brownian motion in laser-Doppler flowmetry of large tissue volumes

被引:17
作者
Binzoni, T [1 ]
Leung, TS
Seghier, ML
Delpy, DT
机构
[1] Univ Geneva, Fac Med, Dept Physiol, Geneva, Switzerland
[2] Univ Hosp, Dept Radiol, Geneva, Switzerland
[3] UCL, Dept Med Phys & Bioengn, London, England
关键词
D O I
10.1088/0031-9155/49/24/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study reports the derivation of a precise mathematical relationship existing between the different p-moments of the power spectrum of the photoelectric current, obtained from a laser-Doppler flowmeter (LDF), and the red blood cell speed. The main purpose is that both the Brownian (defining the 'biological zero') and the translational movements are taken into account, clarifying in this way what the exact contribution of each contribution of each is based on the quasi-elastic scattering theory and holds for multiple scattering (i.e. measurements in large tissue volumes and/or very high red blood cell concentration). The paper also discusses why experimentally there exists a range in which the relationship between the first moment of the power spectrum and the average red blood cells speed may be considered as `linear' and what are the physiological determinants that can result in nonlinearity. A correct way to subtract the biological zero from the LDF data is also proposed. The findings should help in the design of improved LDF instruments and in the interpretation of experimental data.
引用
收藏
页码:5445 / 5458
页数:14
相关论文
共 33 条
[1]   INVIVO EVALUATION OF SIGNAL PROCESSORS FOR LASER DOPPLER TISSUE FLOWMETERS [J].
AHN, H ;
JOHANSSON, K ;
LUNDGREN, O ;
NILSSON, GE .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1987, 25 (02) :207-211
[2]  
[Anonymous], 1992, Laser Pioneers
[3]   Non-invasive laser Doppler perfusion measurements of large tissue volumes and human skeletal muscle blood RMS velocity [J].
Binzoni, T ;
Leung, TS ;
Boggett, D ;
Delpy, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15) :2527-2549
[4]  
BINZONI T, 2002, MAGN RESON MATER PHY, V14, P74
[5]  
Boas D. A., 1996, THESIS U PENNSYLVANI
[6]   MODEL FOR LASER DOPPLER MEASUREMENTS OF BLOOD-FLOW IN TISSUE [J].
BONNER, R ;
NOSSAL, R .
APPLIED OPTICS, 1981, 20 (12) :2097-2107
[7]  
Bonner R. F., 1981, SCATTERING TECHNIQUE, P685
[8]   Laser Doppler, speckle and related techniques for blood perfusion mapping and imaging [J].
Briers, JD .
PHYSIOLOGICAL MEASUREMENT, 2001, 22 (04) :R35-R66
[9]  
Cummins H. Z., 1970, PROGRESS OPTICS, V8, P133
[10]   OBSERVATIONS OF A LASER DOPPLER FLOWMETER OUTPUT MADE USING A CALIBRATION STANDARD [J].
FAIRS, SLE .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (04) :404-406