Evaluation of Common Nasal Cannulas in Neonatal Noninvasive Ventilation (NIV) Using a Novel Neonatal Nasal Model

被引:3
作者
Borg, Ulf [1 ,3 ]
Aviano, Jeffrey [2 ]
Ginani, Milan [2 ]
Li, Kun [2 ]
机构
[1] Medtronic, Dept Med Sci, Resp Intervent, Boulder, CO USA
[2] Medtronic, Dept Res & Dev, Resp Intervent, Carlsbad, CA USA
[3] Medtronic, Dept Med Sci, Resp Intervent, 6135 Gunbarrel Ave, Boulder, CO 80301 USA
关键词
non-invasive ventilation; nasal cannula; airway pressure; ventilator; model; 3D print; POSITIVE-PRESSURE VENTILATION; PEDIATRIC AIRWAY; INFANTS; CPAP;
D O I
10.2147/MDER.S374418
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Non-invasive ventilation (NIV) may reduce intubation rates and be especially beneficial in the care of preterm infants, in addition to other care modalities. Currently, ventilators do not display the pressure at the nares but the pressure in the ventilator tubing system. There are several nasal cannulas available for use to deliver NIV. The purpose of this study was to compare the inspiratory pressure on the ventilator to the measured pressure delivered at the nares using three cannula brands (Manufacturer A, Fisher &Patients and Methods: This bench study utilized a 3D printed nasal model, including nares in multiple sizes to accommodate all nasal prongs studied. The nasal adaptors were connected to neonatal test lungs, to simulate patient breathing. Multiple sizes of nasal cannulas from the three manufacturers were tested for inspiratory vs delivered pressure at the patient side of the cannula, using eight combinations of ventilator settings. Each nasal cannula was tested on six Puritan BennettTM 980 ventilators.Results: The measured delivered pressure at the nares was consistently lower than the clinician-set inspiratory pressure. Across all ventilator settings, 7 of the 11 cannulas delivered significantly less pressure at the nares compared to the inspiratory ventilator pressure (p < 0.01). For each cannula, as inspiratory pressure increased, the difference between delivered and inspiratory pressures also increased. The cannula from Manufacturer B consistently demonstrated the greatest differences between set inspiratory and delivered pressures for each ventilator setting.Conclusion: This study demonstrated substantial differences between ventilator inspiratory pressure and measured delivered pressure, which may have clinical significance. Being unaware of the actual airway pressure delivered to the patient may lead to erroneous adjustments to the level of ventilator pressure, which may be especially consequential to those with delicate and developing respiratory systems.
引用
收藏
页码:307 / 315
页数:9
相关论文
共 22 条
[1]   Synchronized nasal intermittent positive pressure ventilation (SNIPPV) decreases work of breathing (WOB) in premature infants with respiratory distress syndrome (RDS) compared to nasal continuous positive airway pressure (NCPAP) [J].
Aghai, Zubair H. ;
Saslow, Judy G. ;
Nakhla, Tarek ;
Milcarek, Barry ;
Hart, James ;
Lawrysh-Plunkett, Robyn ;
Stahl, Gary ;
Habib, Robert H. ;
Pyon, Kee H. .
PEDIATRIC PULMONOLOGY, 2006, 41 (09) :875-881
[2]   Physiology of non-invasive respiratory support [J].
Alexiou, Stamatia ;
Panitch, Howard B. .
SEMINARS IN FETAL & NEONATAL MEDICINE, 2016, 21 (03) :174-180
[3]   Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated Model [J].
Bailes, Stephanie A. ;
Firestone, Kimberly S. ;
Dunn, Diane K. ;
McNinch, Neil L. ;
Brown, Miraides F. ;
Volsko, Teresa A. .
RESPIRATORY CARE, 2016, 61 (03) :333-339
[4]   Good Publication Practice for Communicating Company-Sponsored Medical Research: GPP3 [J].
Battisti, Wendy P. ;
Wager, Elizabeth ;
Baltzer, Lise ;
Bridges, Dan ;
Cairns, Angela ;
Carswell, Christopher I. ;
Citrome, Leslie ;
Gurr, James A. ;
Mooney, LaVerne A. ;
Moore, B. Jane ;
Pena, Teresa ;
Sanes-Miller, Carol H. ;
Veitch, Keith ;
Woolley, Karen L. ;
Yarker, Yvonne E. .
ANNALS OF INTERNAL MEDICINE, 2015, 163 (06) :461-+
[5]   Synchronized Nasal Intermittent Positive-Pressure Ventilation and Neonatal Outcomes [J].
Bhandari, Vineet ;
Finer, Neil N. ;
Ehrenkranz, Richard A. ;
Saha, Shampa ;
Das, Abhik ;
Walsh, Michele C. ;
Engle, William A. ;
Van Meurs, Krisa P. .
PEDIATRICS, 2009, 124 (02) :517-526
[6]   Nitric Oxide Delivery by Neonatal Noninvasive Respiratory Support Devices [J].
DiBlasi, Robert M. ;
Dupras, Donna ;
Kearney, Christine ;
Costa, Eddie, Jr. ;
Griebel, Jeffrey L. .
RESPIRATORY CARE, 2015, 60 (02) :219-230
[7]   Measured CPAP in a Noninvasive Pediatric Airway and Lung Model [J].
Fernandes, Neil D. ;
Chung, Esther ;
Salt, Michael D. ;
Ejiofor, Beverly ;
Carroll, Ryan W. ;
Kacmarek, Robert M. .
RESPIRATORY CARE, 2021, 66 (01) :87-94
[8]   Is synchronised NIPPV more effective than NIPPV and NCPAP in treating apnoea of prematurity (AOP)? A randomised cross-over trial [J].
Gizzi, Camilla ;
Montecchia, Francesco ;
Panetta, Valentina ;
Castellano, Chiara ;
Mariani, Chiara ;
Campelli, Maristella ;
Papoff, Paola ;
Moretti, Corrado ;
Agostino, Rocco .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2015, 100 (01) :F17-F23
[9]   The Miracles of Surfactant: Less Invasive Surfactant Administration, Nebulization, and Carrier of Topical Drugs [J].
Haertel, Christoph ;
Glaser, Kirsten ;
Speer, Christian P. .
NEONATOLOGY, 2021, 118 (02) :225-234
[10]   Endotracheal Intubation in Neonates: A Prospective Study of Adverse Safety Events in 162 Infants [J].
Hatch, L. Dupree ;
Grubb, Peter H. ;
Lea, Amanda S. ;
Walsh, William F. ;
Markham, Melinda H. ;
Whitney, Gina M. ;
Slaughter, James C. ;
Stark, Ann R. ;
Ely, E. Wesley .
JOURNAL OF PEDIATRICS, 2016, 168 :62-+