A phantom based study on the effect of subject positioning on morphometric X-ray absorptiometry using the Lunar EXPERT-XL

被引:4
|
作者
Thorpe, JA
Steel, SA
Langton, CM
机构
[1] Royal Hull Hosp, NHS Trust, Ctr Metab Bone Dis, Hull HU3 2RW, N Humberside, England
[2] Univ Hull, Hull Royal Infirm, Hull HU3 2RW, N Humberside, England
关键词
D O I
10.1259/bjr.71.851.10434910
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Morphometric X-ray absorptiometry (MXA) relies on accurate measurement of vertical dimensions of vertebrae from a lateral perspective. Deviations resulting from scoliotic curvature or poor patient positioning produce distortions of visible vertebral dimensions and may lead to analysis error. This study utilized a phantom developed at this centre to assess the effect of vertebral malalignment on the accuracy of the MXA technique on the Lunar Expert-XL. Measured vertebral heights were found to be consistently underestimated by an average of 3.7%. Precision ranged from 0.79% for anterior height measurement to 1.03% for middle height measurement. Vertebral malalignment was investigated as the effect of rotation around the anteroposterior, lateral and superoinferior axes. Rotation around the lateral axis produced little discernible effect. However, superoinferior axial rotation showed a change of more than two standard deviations in the mid/posterior ratios of biconcave vertebrae at comparatively small angles of rotation. Anteroposterior axial rotation produced an increase in observed height at small angles of rotation, and a rapid decrease in vertebral height as rotation increased. The results suggest that whilst kyphosis or lordosis of up to at least 5.8 degrees has a minimal effect on MXA, scoliosis of 4.6 degrees or above produces a distinctive effect on the defining crush height ratios.
引用
收藏
页码:1153 / 1161
页数:9
相关论文
共 50 条
  • [41] Effect Of Hydration Status On Body Composition Assessment Using Dual Energy X-ray Absorptiometry And Ultrasound
    Cutrufello, Paul T.
    Landram, Michael J.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05) : 260 - 260
  • [42] Localization Accuracy of a Novel Patient Positioning System with Four X-Ray Tubes and Flat Panel Detectors: A Phantom Study
    Tanabe, S.
    Utsunomiya, S.
    Umetsu, O.
    Sasage, T.
    Takatou, H.
    Kunii, M.
    Amaki, A.
    Kawaguchi, G.
    Aoyama, H.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [43] Effect Of Hydration Status On Body Composition Assessment Using Dual Energy X-ray Absorptiometry And Ultrasound
    Cutrufello, Paul T.
    Landram, Michael J.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2015, 47 (05) : 260 - 260
  • [44] Bipolar disorder and adiposity: a study using whole body dual energy X-ray absorptiometry scans
    Williams, Lana J.
    Pasco, Julie A.
    Jacka, Felice N.
    Henry, Margaret J.
    Dodd, Seetal
    Nicholson, Geoffrey C.
    Kotowicz, Mark A.
    Berk, Michael
    ACTA NEUROPSYCHIATRICA, 2011, 23 (05) : 219 - 223
  • [45] Morphometric x-ray absorptiometry of the spine: Identification of prevalent vertebral deformity in Polish adult population of Lodz region - Part of the EPOLOS study
    Skowronska-Jozwiak, E.
    Pludowski, P.
    Karczmarewicz, E.
    Lewinski, A.
    Lorenc, R.
    OSTEOPOROSIS INTERNATIONAL, 2006, 17 : S195 - S195
  • [46] Background Subtraction From X-Ray Projection Images Using Prior Information: A Phantom Study
    Zhang, J.
    Lasio, G.
    Yi, B.
    Yu, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [47] Effect of Endodontic Cement on Bone Mineral Density Using Serial Dual-energy X-ray Absorptiometry
    Saghiri, Mohammad Ali
    Orangi, Jafar
    Tanideh, Nader
    Janghorban, Kamal
    Sheibani, Nader
    JOURNAL OF ENDODONTICS, 2014, 40 (05) : 648 - 651
  • [48] Reproducibility of pediatric whole body bone and body composition measures by dual-energy X-ray absorptiometry using the GE lunar prodigy
    Margulies, L
    Horlick, M
    Thornton, JC
    Wang, J
    Ioannidou, E
    Heymsfield, SB
    JOURNAL OF CLINICAL DENSITOMETRY, 2005, 8 (03) : 298 - 304
  • [49] Effect of total knee replacement on skeletal muscle mass measurements using dual energy X-ray absorptiometry
    Jang, Jae Young
    Kim, Miji
    Lee, Daehyun
    Won, Chang Won
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [50] Effect of total knee replacement on skeletal muscle mass measurements using dual energy X-ray absorptiometry
    Jae Young Jang
    Miji Kim
    Daehyun Lee
    Chang Won Won
    Scientific Reports, 13