A phantom based study on the effect of subject positioning on morphometric X-ray absorptiometry using the Lunar EXPERT-XL

被引:4
|
作者
Thorpe, JA
Steel, SA
Langton, CM
机构
[1] Royal Hull Hosp, NHS Trust, Ctr Metab Bone Dis, Hull HU3 2RW, N Humberside, England
[2] Univ Hull, Hull Royal Infirm, Hull HU3 2RW, N Humberside, England
来源
BRITISH JOURNAL OF RADIOLOGY | 1998年 / 71卷 / 851期
关键词
D O I
10.1259/bjr.71.851.10434910
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Morphometric X-ray absorptiometry (MXA) relies on accurate measurement of vertical dimensions of vertebrae from a lateral perspective. Deviations resulting from scoliotic curvature or poor patient positioning produce distortions of visible vertebral dimensions and may lead to analysis error. This study utilized a phantom developed at this centre to assess the effect of vertebral malalignment on the accuracy of the MXA technique on the Lunar Expert-XL. Measured vertebral heights were found to be consistently underestimated by an average of 3.7%. Precision ranged from 0.79% for anterior height measurement to 1.03% for middle height measurement. Vertebral malalignment was investigated as the effect of rotation around the anteroposterior, lateral and superoinferior axes. Rotation around the lateral axis produced little discernible effect. However, superoinferior axial rotation showed a change of more than two standard deviations in the mid/posterior ratios of biconcave vertebrae at comparatively small angles of rotation. Anteroposterior axial rotation produced an increase in observed height at small angles of rotation, and a rapid decrease in vertebral height as rotation increased. The results suggest that whilst kyphosis or lordosis of up to at least 5.8 degrees has a minimal effect on MXA, scoliosis of 4.6 degrees or above produces a distinctive effect on the defining crush height ratios.
引用
收藏
页码:1153 / 1161
页数:9
相关论文
共 50 条
  • [1] Development of a phantom for morphometric X-ray absorptiometry
    Rea, JA
    Blake, GM
    Fogelman, I
    BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (880): : 341 - 350
  • [2] Development and evaluation of a phantom for morphometric X-ray absorptiometry
    Steel, SA
    Thorpe, JA
    Walker, R
    Howey, S
    Langton, CM
    OSTEOPOROSIS INTERNATIONAL, 1999, 9 (01) : 38 - 44
  • [3] Development and Evaluation of a Phantom for Morphometric X-ray Absorptiometry
    S. A. Steel
    J. A. Thorpe
    R. Walker
    S. Howey
    C. M. Langton
    Osteoporosis International, 1999, 9 : 38 - 44
  • [4] Bone mineral density of the wind using dual X-ray absorptiometry (Lunar-Expert).
    Diessel, E
    Alenfeld, F
    Gowin, W
    Mueller, C
    Felsenberg, D
    JOURNAL OF BONE AND MINERAL RESEARCH, 1996, 11 : S580 - S580
  • [5] The effect of tissue depth and scan mode on bone mineral density using a Lunar Expert-XL fan beam densitometer
    Goodby, A
    Steel, SA
    OSTEOPOROSIS INTERNATIONAL, 2000, 11 : S44 - S44
  • [6] Visual identification of vertebral fractures in osteoporosis using morphometric X-ray absorptiometry
    Ferrar, L
    Jiang, G
    Eastell, R
    Peel, NFA
    JOURNAL OF BONE AND MINERAL RESEARCH, 2003, 18 (05) : 933 - 938
  • [7] Vertebral morphometry by morphometric radiography (MR) and morphometric x-ray absorptiometry (MXA) using the same reference population
    Ferrar, L
    Barrington, NA
    Jiang, G
    Eastell, R
    JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (09) : P38 - P38
  • [8] Vertebral morphometry: Repeat scan precision using the lunar Expert-XL and the hologic 4500A. A study for the 'WISDOM' RCT of hormone replacement therapy
    Crabtree, N
    Wright, J
    Walgrove, A
    Rea, J
    Hanratty, L
    Lunt, M
    Fogelman, I
    Palmer, R
    Vickers, M
    Compston, JE
    Reeve, J
    OSTEOPOROSIS INTERNATIONAL, 2000, 11 (06) : 537 - 543
  • [9] Vertebral Morphometry: Repeat Scan Precision Using the Lunar Expert-XL and the Hologic 4500A. A Study for the “WISDOM” RCT of Hormone Replacement Therapy
    N. Crabtree
    J. Wright
    A. Walgrove
    J. Rea
    L. Hanratty
    M. Lunt
    I. Fogelman
    R. Palmer
    M. Vickers
    J. E. Compston
    J. Reeve
    Osteoporosis International, 2000, 11 : 537 - 543
  • [10] Vertebral morphometry by morphometric radiography (MR) and morphometric X-ray absorptiometry (MXA) using the same reference population.
    Ferrar, L
    Barrington, NA
    Jiang, G
    Eastell, R
    JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 : P306 - P306