Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery

被引:27
作者
Cheung, Hoiyin [1 ]
Dong, Quan [1 ]
Dong, Rong [1 ]
Yu, Buwei [1 ]
机构
[1] Shanghai Jiao Tong Univ, Ruijin Hosp, Dept Anesthesiol, Sch Med, Shanghai 200025, Peoples R China
关键词
Measurement techniques; Cardiac output; Thermodilution; Non-invasiveness; Bioreactance; CRITICALLY-ILL PATIENTS; PULSE CONTOUR; CATHETER; BIOREACTANCE; BIOIMPEDANCE; CARE;
D O I
10.1007/s00540-014-1938-z
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
This observational study was designed to evaluate the clinical value of cardiac output (CO) obtained via bioreactance (NICOM (TM)) as compared with values of CO obtained via thermodilution (using pulmonary artery catheter, Vigilance (TM)) and the thoracic bioimpedance (BioZ.com (TM)), in patients undergoing off-pump coronary artery bypass surgery. Fifty American Society of Anesthesiologists physical status I-III patients, aged 38-81 years, scheduled for off-pump coronary artery bypass surgery were enrolled in this study. CO data (NCO, BCO, PCO) were recorded during the operative period at ten time points after stable hemodynamic conditions were achieved. The equation of the relationship between the PCO and NCO is PCO = 0.945 x NCO + 0.328 (r = 0.77), and that of PCO and BCO is PCO = 0.965 x BCO + 0.729 (r = 0.63). Furthermore, no statistical difference was found between PCO versus NCO (mean (SD): 4.4 (1.1) versus 4.4 (0.9), p = 0.431). A significant correlation was found between PCO and NCO (r = 0.77, p < 0.001). Correlation was also found between PCO and BCO (r = 0.63, p < 0.001). The NICOM device is a safe, convenient, and reliable device for measuring continuous non-invasive cardiac output and cardiac index, and the trends of change in CO during the surgery are similar between NICOM and PAC.
引用
收藏
页码:416 / 420
页数:5
相关论文
共 22 条
[1]   Fluid responsiveness predicted by noninvasive Bioreactance-based passive leg raise test [J].
Benomar, Brahim ;
Ouattara, Alexandre ;
Estagnasie, Philippe ;
Brusset, Alain ;
Squara, Pierre .
INTENSIVE CARE MEDICINE, 2010, 36 (11) :1875-1881
[2]  
BOLDT J, 1994, CRIT CARE MED, V22, P1913
[3]   The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[4]   A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output [J].
Critchley, Lester A. ;
Lee, Anna ;
Ho, Anthony M. -H. .
ANESTHESIA AND ANALGESIA, 2010, 111 (05) :1180-1192
[5]   An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery [J].
de Wilde, R. B. P. ;
Schreuder, J. J. ;
van den Berg, P. C. M. ;
Jansen, J. R. C. .
ANAESTHESIA, 2007, 62 (08) :760-768
[6]   Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method [J].
Engoren, M ;
Barbee, D .
AMERICAN JOURNAL OF CRITICAL CARE, 2005, 14 (01) :40-45
[7]  
Guyatt G, 1991, J Intensive Care Med, V6, P91
[8]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[9]   Simultaneous comparison of thoracic bioimpedance and arterial pulse waveform-derived cardiac output with thermodilution measurement [J].
Hirschl, MM ;
Kittler, H ;
Woisetschläger, C ;
Siostrzonek, P ;
Staudinger, T ;
Kofler, J ;
Oschatz, E ;
Bur, A ;
Gwechenberger, M ;
Laggner, AN .
CRITICAL CARE MEDICINE, 2000, 28 (06) :1798-1802
[10]   Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance [J].
Keren, Hanan ;
Burkhoff, Daniel ;
Squara, Pierre .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2007, 293 (01) :H583-H589