Precise sample metering method by coordinated burst action of hydrophobic burst valves applied to dried blood spot collection

被引:3
作者
Vloemans, Dries [1 ]
Van Hileghem, Lorenz [1 ]
Verbist, Wannes [1 ]
Thomas, Debby [2 ]
Dal Dosso, Francesco [1 ]
Lammertyn, Jeroen [1 ]
机构
[1] Katholieke Univ Leuven, Biosensors Grp, Dept Biosyst, Willem Croylaan 42,Box 2428, B-3001 Leuven, Belgium
[2] Katholieke Univ Leuven, Dept Pharmaceut & Pharmacol Sci, Herestr 49,Box 424, B-3000 Leuven, Belgium
关键词
QUANTITATIVE BIOANALYSIS; HEMATOCRIT; INFLIXIMAB; DEVICE; ISSUES; VOLUME; BIAS;
D O I
10.1039/d1lc00422k
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Dried blood spot (DBS) sampling by finger-pricking has recently gained a lot of interest as an alternative sample collection method. The reduced invasiveness, requirement of lower sample volumes and suitability for long-term storage at room temperature make DBS ideal for use in home settings or low-resource environments. However, traditional protocols often suffer from biased analysis data due to variable and not exactly known blood volumes present in the samples. In this work, a novel device has been developed to split-off precisely metered volumes from a blood drop and load them on pre-cut filter paper. Hereto, hydrophobic burst valves (HBV) were developed to temporarily retain a fluid flow, configurable to burst at pressures within a range of 175-600 Pa. By combining HBVs with different burst pressures, a volume metering system was developed to allow parallel metering of multiple pre-defined sample volumes. The system was shown to be accurate and consistent for blood volumes between 5-15 mu L and for hematocrit levels spanning the range of 25-70%. Finally, a point-of-care DBS sampling device was developed combining the self-powered microfluidic SIMPLE technology. To evaluate the system's practical applicability, a validation study in the context of therapeutic drug monitoring of biologicals was performed using adalimumab-spiked blood samples. Microfluidic DBS samples showed good performance compared to the traditional DBS method with improved recovery rates (86% over 62%). This innovative metering system, allowing for parallelization and integration with complex liquid manipulations, will greatly impact the field of robust sampling, sample preparation, storage and analysis at the point-of-care.
引用
收藏
页码:4445 / 4454
页数:10
相关论文
共 47 条
  • [1] Investigation of Different Approaches to Incorporating Internal Standard in DBS Quantitative Bioanalytical Workflows and Their Effect on Nullifying Hematocrit-Based Assay Bias
    Abu-Rabie, Paul
    Denniff, Philip
    Spooner, Neil
    Chowdhry, Babur Z.
    Pullen, Frank S.
    [J]. ANALYTICAL CHEMISTRY, 2015, 87 (09) : 4996 - 5003
  • [2] Influence of immunogenicity on the long-term efficacy of infliximab in Crohn's disease
    Baert, F
    Noman, M
    Vermeire, S
    Van Assche, G
    D'Haens, G
    Carbonez, A
    Rutgeerts, P
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2003, 348 (07) : 601 - 608
  • [3] Bang I, 1913, BIOCHEM Z, V49, P19
  • [4] Is there potential for therapeutic drug monitoring of biologic agents in rheumatoid arthritis?
    Bastida, Carla
    Ruiz, Virginia
    Pascal, Mariona
    Yague, Jordi
    Sanmarti, Raimon
    Soy, Dolors
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 2017, 83 (05) : 962 - 975
  • [5] Burst valves for commercial microfluidics: a critical analysis
    Bauer, Maria
    Ataei, Marzieh
    Caicedo, Maria
    Jackson, Kirsten
    Madou, Marc
    Bousse, Luc
    [J]. MICROFLUIDICS AND NANOFLUIDICS, 2019, 23 (07)
  • [6] Generation and characterization of a unique panel of anti-adalimumab specific antibodies and their application in therapeutic drug monitoring assays
    Bian, Sumin
    Van Stappen, Thomas
    Baert, Filip
    Compernolle, Griet
    Brouwers, Els
    Tops, Sophie
    de Vries, Annick
    Rispens, Theo
    Lammertyn, Jeroen
    Vermeire, Severine
    Gils, Ann
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 125 : 62 - 67
  • [7] CARTER GF, 1978, BRIT J IND MED, V35, P235
  • [8] Influence of haematocrit level on the kinetics of blood spreading on thin porous medium during dried blood spot sampling
    Chao, Tzu Chieh
    Trybala, Anna
    Starov, Victor
    Das, Diganta B.
    [J]. COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2014, 451 : 38 - 47
  • [9] SIMPLE analytical model for smart microfluidic chip design
    Dal Dosso, Francesco
    Bondarenko, Yura
    Kokalj, Tadej
    Lammertyn, Jeroen
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2019, 287 : 131 - 137
  • [10] Innovative Hydrophobic Valve Allows Complex Liquid Manipulations in a Self-Powered Channel-Based Microfluidic Device
    Dal Dosso, Francesco
    Tripodi, Lisa
    Spasic, Dragana
    Kokalj, Tadej
    Lammertyn, Jeroen