Investigation of thermal neutron detection capability of a CdZnTe detector in a mixed gamma-neutron radiation field

被引:1
作者
Yucel, Haluk [1 ]
Nartturk, R. Bora [1 ]
Zumrut, Senem [1 ]
Gedik, Gizem [1 ]
Karadag, Mustafa [2 ]
机构
[1] Ankara Univ, Inst Nucl Sci, Dogol St, TR-06100 Ankara, Turkey
[2] Gazi Univ, TR-06100 Ankara, Turkey
关键词
neutron detection; CdZnTe; prompt gamma ray; thermal neutron; cadmium; neutron sensitivity; Am-241-Be source;
D O I
10.2478/nuka-2018-0007
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The aim of this study was to investigate the thermal neutron measurement capability of a CdZnTe detector irradiated in a mixed gamma-neutron radiation field. A CdZnTe detector was irradiated in one of the irradiation tubes of a Am-241-Be source unit to determine the sensitivity factors of the detector in terms of peak count rate (counts per second [cps]) per neutron flux (in square centimeters per second) [cps/neutron-cm(-2).s(-1)]. The CdZnTe detector was covered in a 1-mm-thick cadmium (Cd) cylindrical box to completely absorb incoming thermal neutrons via Cd-113 (n,gamma) capture reactions. To achieve, this Cd-covered CdZnTe detector was placed in a well-thermalized neutron field (f-ratio = 50.9 +/- 1.3) in the irradiation tube of the Am-241-Be neutron source. The gamma-ray spectra were acquired, and the most intense gamma-ray peak at 558 keV (0.74 gamma/n) was evaluated to estimate the thermal neutron flux. The epithermal component was also estimated from the bare CdZnTe detector irradiation because the epithermal neutron cutoff energy is about 0.55 eV at the 1-mm-thick Cd filter. A high-density polyethylene moderating cylinder box can also be fitted into the Cd filter box to enhance thermal sensitivity because of moderation of the epithermal neutron component. Neutron detection sensitivity was determined from the measured count rates from the 558 keV photopeak, using the measured neutron fluxes at different irradiation positions. The results indicate that the CdZnTe detector can serve as a neutron detector in mixed gamma-neutron radiation fields, such as reactors, neutron generators, linear accelerators, and isotopic neutron sources. New thermal neutron filters, such as Gd and Tb foils, can be tested instead of the Cd filter due to its serious gamma-shielding effect.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 15 条
  • [1] Coulon R, 2016, IEEE NSS S P 2016 IE, P1
  • [2] Gadolinium for neutron detection in current nuclear instrumentation research: A review
    Dumazert, J.
    Coulon, R.
    Lecomte, Q.
    Bertrand, G. H. V.
    Hamel, M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 882 : 53 - 68
  • [3] Compensation scheme for online neutron detection using a Gd-covered CdZnTe sensor
    Dumazert, Jonathan
    Coulon, Romain
    Kondrasovs, Vladimir
    Boudergui, Karim
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 857 : 7 - 15
  • [4] Compensated bismuth-loaded plastic scintillators for neutron detection using low-energy pseudo-spectroscopy
    Dumazert, Jonathan
    Coulon, Romain
    Bertrand, Guillaume H. V.
    Normand, Stephane
    Mechin, Laurence
    Hamel, Matthieu
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 819 : 25 - 32
  • [5] EXFOR Database, 2017, EXFOR DATABASE
  • [6] FASASI M, 1988, RADIAT PROT DOSIM, V23, P429
  • [7] Spectrometric response of large volume CdZnTe coplanar detectors
    González, R
    Pérez, JM
    Vela, O
    de Burgos, E
    Oller, JC
    Gostilo, V
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 2076 - 2084
  • [8] Position-sensitive single carrier CdZnTe detectors
    He, Z
    Knoll, GF
    Wehe, DK
    Miyamoto, J
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 388 (1-2) : 180 - 185
  • [9] Measurement of thermal neutron cross-sections and resonance integrals for 71Ga(n, γ)72Ga and 75As(n, γ)76As by using 241Am-Be isotopic neutron source
    Karadag, M
    Yücel, H
    Tan, M
    Özmen, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 501 (2-3) : 524 - 535
  • [10] Evaluation of CdZnTe as neutron detector around medical accelerators
    Martin-Martin, A.
    Iniguez, M. P.
    Luke, P. N.
    Barquero, R.
    Lorente, A.
    Morchon, J.
    Gallego, E.
    Quincoces, G.
    Marti-Climent, J. M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2009, 133 (04) : 193 - 199