ObjectiveTo describe the occurrence of severe polyethylene wear of the acetabular component in 3 dogs that ultimately led to implant failure, and to report the outcome of revision of total hip arthroplasty (THA) in these cases. Study DesignCase report AnimalsDogs (n=3). ResultsThree dogs with hip dysplasia underwent routine THA before 2 years of age. All dogs became acutely lame 4-5 years after THA. Radiographs revealed acetabular cup wear in all cases, shell fragmentation in 2 cases, and additional stem failure in 1 case. All dogs underwent successful revision surgery, during which severe polyethylene wear and metallosis was observed. Revision surgery provided an excellent clinical outcome in all dogs from 10 months to 4 years post revision. ConclusionPolyethylene wear and implant failure may be a significant long-term complication of THA not previously recognized, which may be especially true when placed at an early age in a large breed, active dog, or in a dog with significant morbidity associated with other limbs. A successful revision was achieved in all 3 dogs. Additional investigation is warranted to document the risk factors associated with severe polyethylene wear in canine THA.
引用
收藏
页码:664 / 671
页数:8
相关论文
共 30 条
[1]   Chronic inflammation, immunosuppression and cancer: New insights and outlook [J].
Baniyash, M .
SEMINARS IN CANCER BIOLOGY, 2006, 16 (01) :80-88
[2]  
Bankston A B, 1993, J Arthroplasty, V8, P315
[3]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]   Activity levels and polyethylene wear of patients 10 years post hip replacement [J].
Bennett, D. ;
Humphreys, L. ;
O'Brien, S. ;
Kelly, C. ;
Orr, J. ;
Beverland, D. E. .
CLINICAL BIOMECHANICS, 2008, 23 (05) :571-576
[5]  
BioMedtrix Newsletter, 2014, BIOMEDTRIX NEWSLETTE
[6]  
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[7]   IMPLANTATION OF AN UNCEMENTED TOTAL HIP-PROSTHESIS - TECHNIQUE AND INITIAL RESULTS OF 100 ARTHROPLASTIES [J].
DEYOUNG, DJ ;
DEYOUNG, BA ;
ABERMAN, HA ;
KENNA, RV ;
HUNGERFORD, DS .
VETERINARY SURGERY, 1992, 21 (03) :168-177
[8]  
Dyce J, 2000, VET COMP ORTHOPAED, V13, P78
[9]   The Surgical Options and Clinical Evidence for Treatment of Wear or Corrosion Occurring With THA or TKA [J].
Engh, Charles A., Jr. ;
Ho, Henry ;
Padgett, Douglas E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (12) :3674-3686
[10]   The Impact of Storage Time on the Wear Rates of Ultrahigh-Molecular-Weight Polyethylene Acetabular Liners in Hip Simulators [J].
Fan Weimin ;
Song Huanghe ;
Li Xiang ;
Liu Feng ;
Wang Qing .
JOURNAL OF ARTHROPLASTY, 2009, 24 (04) :543-548