Determination of three analgesics in pharmaceutical and urine sample on nano poly (3, 4-ethylenedioxythiophene) modified electrode

被引:16
作者
Gopu, Gopalakrishnan [1 ]
Muralidharan, Baladhandapani [2 ]
Vedhi, Chinnapiyan [3 ]
Manisankar, Paramasivam [1 ]
机构
[1] Alagappa Univ, Dept Ind Chem, Karaikkudi 630003, Tamil Nadu, India
[2] Birla Inst Technol & Sci Pilani, Dept Chem, Dubai, U Arab Emirates
[3] VO Chidambaram Coll, Dept Chem, Tuticorin 628008, Tamil Nadu, India
关键词
Acetaminophen; Acetylsalicylic acid; Dipyrone; Nano Poly (3,4-ethylenedioxythiophene); Cyclic voltammetry; Stripping voltammetry; GLASSY-CARBON ELECTRODE; ACETYLSALICYLIC-ACID; SALICYLIC-ACID; VOLTAMMETRIC DETERMINATION; DOSAGE FORMS; SPECTROPHOTOMETRIC DETERMINATION; LIQUID-CHROMATOGRAPHY; PASTE ELECTRODE; ASCORBIC-ACID; HUMAN SERUM;
D O I
10.1007/s11581-011-0619-2
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Three analgesics, acetaminophen, acetylsalicylic acid, and dipyrone were determined by stripping voltammetry using nanosized poly(3,4-ethylenedioxythiophene)-modified glassy carbon electrode. The cyclic voltammetric behavior of the three analgesics was studied in aqueous acid, neutral, and alkaline conditions. One well-defined oxidation peak each for acetaminophen and acetylsalicylic acid and three oxidation peaks for dipyrone were observed in the cyclic voltammograms. The influence of pH, scan rate, and concentration revealed irreversible diffusion controlled reaction. A systematic study of the experimental parameters that affect the differential pulse stripping voltammetric response was carried out. Calibration was made under maximum peak current conditions. The scanning electron microscope analysis confirmed good accumulation of the drugs on the electrode surface. The range of study for both acetaminophen, acetylsalicylic acid were 0.015-0.4 and dipyrone was 0.025-0.4 mu g/ml. The lower limit of determination for both acetaminophen, acetylsalicylic acid was 0.01 mu g/mL and for dipyrone was 0.02 mu g/mL. The suitability of the method for the determination of the three analgesics in pharmaceutical preparations and urine samples was also ascertained.
引用
收藏
页码:231 / 239
页数:9
相关论文
共 41 条
[1]  
Akay C, 2008, TURK J MED SCI, V38, P167
[2]  
Alves L, 2009, ANAL CHIM ACTA, V642, P212
[3]  
Andres C., 2000, TALANTA, V53, P417
[4]  
[Anonymous], KMITL SCI TECH J
[5]   Flow injection-Fourier transform infrared spectrometric determination of paracetamol in pharmaceuticals [J].
Bouhsain, Z ;
Garrigues, S ;
delaGuardia, M .
ANALYST, 1996, 121 (05) :635-639
[6]   Salicylic acid determination in cow urine and drugs using a bienzymatic sensor [J].
Campanella, L. ;
Gregori, E. ;
Tomassetti, M. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 42 (01) :94-99
[7]   Electrochemical study of bergenin on a poly(4-(2-pyridylazo)-resorcinol) modified glassy carbon electrode and its determination in tablets and urine [J].
Chen, Jinghua ;
Zhang, Jing ;
Zhuang, Qian ;
Zhang, Shaobo ;
Lin, Xinhua .
TALANTA, 2007, 72 (05) :1805-1810
[8]   Influence of particle size on the quantitative determination of salicylic acid in a pharmaceutical ointment using FT-Raman spectroscopy [J].
De Beer, T. R. M. ;
Baeyens, Wr. G. ;
Vander Heyden, Y. ;
Remon, J. P. ;
Vervaet, C. ;
Verpoort, F. .
EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2007, 30 (3-4) :229-235
[9]   Derivative-ratio spectrophotometric method for the determination of ternary mixture of aspirin, paracetamol and salicylic acid [J].
El-Yazbi, Fawzi A. ;
Hammud, Hassan H. ;
Assi, Sulaf A. .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2007, 68 (02) :275-278
[10]   Carbon film resistor electrode for amperometric determination of acetaminophen in pharmaceutical formulations [J].
Felix, Fabiana S. ;
Brett, Christopher M. A. ;
Angnes, Lucio .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (05) :1622-1627