Assessment of accuracy and precision of 3D reconstruction of unicompartmental knee arthroplasty in upright position using biplanar radiography

被引:13
作者
Tsai, Tsung-Yuan [1 ]
Dimitriou, Dimitris [1 ]
Hosseini, Ali [1 ]
Liow, Ming Han Lincoln [1 ]
Torriani, Martin [1 ]
Li, Guoan [1 ]
Kwon, Young-Min [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Orthopaed Surg, Boston, MA 02114 USA
关键词
Unicompartmental knee arthroplasty; Component position; Biplanar radiograph; Lower limb alignment; COMPUTED-TOMOGRAPHY; POLYETHYLENE WEAR; EARLY FAILURE; IN-VIVO; SYSTEM; ALIGNMENT; REPLACEMENT; KINEMATICS; TORSION;
D O I
10.1016/j.medengphy.2016.04.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to evaluate the precision and accuracy of 3D reconstruction of UKA component position, contact location and lower limb alignment in standing position using biplanar radiograph. Two human specimens with 4 medial UKAs were implanted with beads for radiostereometric analysis (RSA). The specimens were frozen in standing position and CT-scanned to obtain relative positions between the beads, bones and UKA components. The specimens were then imaged using biplanar radiograph (EOS). The positions of the femur, tibia, UKA components and UKA contact locations were obtained using RSA and EOS-based techniques. Intraclass correlation coefficient (ICC) was calculated for inter-observer reliability of the EOS technique. The average (standard deviation) of the differences between two techniques in translations and rotations were less than 0.18 (0.29) mm and 0.39 degrees (0.66 degrees) for UKA components. The root-mean-square-errors (RMSE) of contact location along the anterior/posterior and medial/lateral directions were 0.84 mm and 0.30 mm. The RMSEs of the knee rotations were less than 1.70 degrees. The ICCs for the EOS-based segmental orientations between two raters were larger than 0.98. The results suggest the EOS-based 3D reconstruction technique can precisely determine component position, contact location and lower limb alignment for UKA patients in weight-bearing standing position. (c) 2016 IPEM.
引用
收藏
页码:633 / 638
页数:6
相关论文
共 28 条
[1]   Early failure of unicompartmental knee arthroplasty leading to revision [J].
Aleto, Thomas J. ;
Berend, Michael E. ;
Ritter, Merrill A. ;
Faris, Philip M. ;
Meneghini, R. Michael .
JOURNAL OF ARTHROPLASTY, 2008, 23 (02) :159-163
[2]  
[Anonymous], MACH INTELL
[3]   Three-dimensional lower extremity alignment in the weight-bearing standing position in healthy elderly subjects [J].
Ariumi, Akihiro ;
Sato, Takashi ;
Kobayashi, Koichi ;
Koga, Yoshio ;
Omori, Go ;
Minato, Izumi ;
Endo, Naoto .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2010, 15 (01) :64-70
[4]   Polyethylene wear in a non-congruous unicompartmental knee replacement: a retrieval analysis [J].
Ashraf, T ;
Newman, JH ;
Desai, VV ;
Beard, D ;
Nevelos, JE .
KNEE, 2004, 11 (03) :177-181
[5]   Early failure of minimally invasive unicompartmental knee arthroplasty is associated with obesity [J].
Berend, KR ;
Lombardi, AV ;
Mallory, TH ;
Adams, JB ;
Groseth, KL .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (440) :60-66
[6]   An optimized image matching method for determining in-vivo TKA kinematics with a dual-orthogonal fluoroscopic imaging system [J].
Bingham, Jeffrey ;
Li, Guoan .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (04) :588-595
[7]   Experimental assessment of precision and accuracy of radio stereometric analysis for the determination of polyethylene wear in a total hip replacement model [J].
Bragdon, CR ;
Malchau, H ;
Yuan, XH ;
Perinchief, R ;
Kärrholm, J ;
Börlin, N ;
Estok, DM ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) :688-695
[8]   Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays [J].
Chaibi, Y. ;
Cresson, T. ;
Aubert, B. ;
Hausselle, J. ;
Neyret, P. ;
Hauger, O. ;
de Guise, J. A. ;
Skalli, W. .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2012, 15 (05) :457-466
[9]   Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography [J].
Delin, Cyrille ;
Silvera, Stephane ;
Bassinet, Celine ;
Thelen, Philippe ;
Rehel, Jean-Luc ;
Legmann, Paul ;
Folinais, Dominique .
EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) :371-377
[10]   Unicompartmental knee arthroplasty with the Oxford prosthesis in patients with medial compartment arthritis [J].
Emerson, Roger H. ;
Higgins, Linda L. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (01) :118-122