Influence of Screw Surface Treatment on Retention of Implant-Supported Fixed Partial Dentures

被引:11
作者
de Moura, Marcos Boaventura [1 ]
Rodrigues, Renata Borges [2 ]
Pinto, Leandro Moreira [1 ]
de Araujo, Cleudmar Amaral [3 ]
Novais, Veridiana Resende [2 ]
Simamoto Junior, Paulo Cezar [1 ]
机构
[1] Univ Fed Uberlandia, Sch Dent, Dept Occlus Fixed Prosthodont & Dent Mat, Uberlandia, MG, Brazil
[2] Univ Fed Uberlandia, Dent Sch, Dept Operat Dent & Dent Mat, Uberlandia, MG, Brazil
[3] Univ Fed Uberlandia, Mech Sch, Mech Projects Lab Henner A Gomide, Uberlandia, MG, Brazil
关键词
dental implants; implant prosthesis; implant screw; mechanical cycling; splinting; FINITE-ELEMENT-ANALYSIS; ABUTMENT SCREW; INTERPROXIMAL CONTACT; PROSTHETIC SCREWS; DENTAL IMPLANTS; TORQUE; PRELOAD; BONE; COMPLICATIONS; RESTORATIONS;
D O I
10.1563/aaid-joi-D-16-00145
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to assess the effect of a diamond-like carbon (DLC) coating on the removal torque of prefabricated implant screws after cyclic loading. Four groups with two crowns supported by two implants (n = 5) were obtained according to splinted and nonsplinted prosthesis, using titanium or DLC screws (splinted crowns with titanium screw [STi], splinted crowns with DLC screw [SC], nonsplinted crowns and titanium screw [NSTi], and nonsplinted crowns and DLC screw [NSC]). The prosthetic screws were tightened at 32 Ncm and retightened, and the specimens were submitted to 106 mechanical cycles (4 Hz/98 N). After cyclic loading, loosening torque was evaluated, and the final measurements were performed. Data were analyzed by two-way analysis of variance and Tukey's test (alpha = .005). There was statistically significance in the interaction of screw X splinting (P = .003). For the group that used titanium screws, NSTi showed smaller removal torque compared with STi. It was concluded that the use of the DLC coating screws in nonsplinted prosthesis maintain the torque after cyclic loading.
引用
收藏
页码:254 / 260
页数:7
相关论文
共 39 条
[1]   A MULTICENTER REPORT ON OSSEOINTEGRATED ORAL IMPLANTS [J].
ALBREKTSSON, T .
JOURNAL OF PROSTHETIC DENTISTRY, 1988, 60 (01) :75-84
[2]   Preload Evaluation of Different Screws in External Hexagon Joint [J].
Assuncao, Wirley Goncalves ;
Delben, Juliana Aparecida ;
Tabata, Lucas Fernando ;
Barao, Valentim Adelino Ricardo ;
Gomes, Erica Alves ;
Garcia, Idelmo Rangel, Jr. .
IMPLANT DENTISTRY, 2012, 21 (01) :46-50
[3]   Loosening torque of Universal Abutment screws after cyclic loading: influence of tightening technique and screw coating [J].
Bacchi, Atais ;
Regalin, Alexandre ;
Brilhante Bhering, Claudia Lopes ;
Alessandretti, Rodrigo ;
Spazzin, Aloisio Oro .
JOURNAL OF ADVANCED PROSTHODONTICS, 2015, 7 (05) :375-379
[4]  
Bickford JH, 1995, INTRO DESIGN BEHAV B, P528
[5]  
Binon PF.S., 1994, INT J ORAL MAX IMPL, V9, P48
[6]   Mechanics of the taper integrated screwed-in (TIS) abutments used in dental implants [J].
Bozkaya, D ;
Müftü, S .
JOURNAL OF BIOMECHANICS, 2005, 38 (01) :87-97
[7]  
Byrne Declan, 2006, J Prosthodont, V15, P164, DOI 10.1111/j.1532-849X.2006.00096.x
[8]   Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study [J].
Cardoso, Mayra ;
Torres, Marcelo Ferreira ;
Veras Lourenco, Eduardo Jose ;
Telles, Daniel de Moraes ;
Silveira Rodrigues, Renata Cristina ;
Ribeiro, Ricardo Faria .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (04) :475-480
[9]  
Ciccio M, 2009, Minerva Stomatol, V58, P9
[10]   FEM Analysis of Mandibular Prosthetic Overdenture Supported by Dental Implants: Evaluation of Different Retention Methods [J].
Cicciu, M. ;
Cervino, G. ;
Bramanti, E. ;
Lauritano, F. ;
Lo Gudice, G. ;
Scappaticci, L. ;
Rapparini, A. ;
Guglielmino, E. ;
Risitano, G. .
COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2015, 2015