Calibration transfer for bioprocess Raman monitoring using Kennard Stone piecewise direct standardization and multivariate algorithms

被引:17
作者
Petillot, Laure [1 ]
Pewny, Fiona [2 ]
Wolf, Martin [2 ]
Sanchez, Celia [1 ]
Thomas, Fabrice [1 ]
Sarrazin, Johan [1 ]
Fauland, Katharina [2 ]
Katinger, Hermann [2 ]
Javalet, Charlotte [1 ]
Bonneville, Christophe [1 ]
机构
[1] RESOLUT Spectra Syst, Meylan, France
[2] Polymun Sci, Klosterneuburg, Austria
基金
欧盟地平线“2020”;
关键词
cell culture monitoring; model transferability; multivariate model maintenance; process analytical technology; Raman spectroscopy; ORTHOGONAL SIGNAL CORRECTION; LEAST-SQUARES; SPECTROSCOPY; IMPROVEMENT; PREDICTION;
D O I
10.1002/eng2.12230
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In the biopharmaceutical industry, Raman spectroscopy is now a proven PAT tool that enables in-line simultaneous monitoring of several CPPs and CQAs in real-time. However, as Raman monitoring requires multivariate modeling, variabilities unknown by models can impact the monitoring prediction accuracy. With the widespread use of Raman PAT tools, it is necessary to fix instrumental variability impacts, encountered for instance during a device replacement. In this work, we investigated the impact of instrumental variability between probes inside a multi-channel analyzer and between two analyzers, and explored solutions to correct them on model prediction errors in cell cultures. It is shown that the Kennard Stone piecewise direct standardization (KS PDS) method enables to lower model prediction errors between probes of a multi-channel analyzer from 20% to 10% on the cell densities (TCD/VCD). Considering the integration of a new device or the replacement of a previous one, it has been determined that a first cell culture monitoring can be directly performed with the new analyzer calibrated by the KS PDS method based on the dataset from the previous analyzer, with an accuracy better than 10% on the main components of the culture like glucose, lactate, and the cell densities. Then, the new data obtained by the new analyzer can be inserted in a global calibration dataset to integrate instrumental variability in the chemometric model: it is shown that only one batch with the new device in a consistent and equilibrated calibration dataset was sufficient to correct the prediction gap induced by instrumental variability, allowing to exploit the data from previous analyzers considering optimized methods. This methodology provides good multivariate calibration model prediction errors throughout the instrumental changes which is a requirement for model maintenance.
引用
收藏
页数:19
相关论文
共 30 条
  • [1] Real Time Monitoring of Multiple Parameters in Mammalian Cell Culture Bioreactors Using an In-Line Raman Spectroscopy Probe
    Abu-Absi, Nicholas R.
    Kenty, Brian M.
    Cuellar, Maryann Ehly
    Borys, Michael C.
    Sakhamuri, Sivakesava
    Strachan, David J.
    Hausladen, Michael C.
    Li, Zheng Jian
    [J]. BIOTECHNOLOGY AND BIOENGINEERING, 2011, 108 (05) : 1215 - 1221
  • [2] Pharmaceutical analysis in solids using front face fluorescence spectroscopy and multivariate calibration with matrix correction by piecewise direct standardization
    Alves, Julio Cesar L.
    Poppi, Ronei J.
    [J]. SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2013, 103 : 311 - 318
  • [3] In-line and real-time prediction of recombinant antibody titer by in situ Raman spectroscopy
    Andre, Silvere
    Saint Cristau, Lydia
    Gaillard, Sabine
    Devos, Olivier
    Calvosa, Eric
    Duponchel, Ludovic
    [J]. ANALYTICA CHIMICA ACTA, 2015, 892 : 148 - 152
  • [4] ASTM International/U.S. Department of Defense, 2014, ASME BOILER PRESSU C, V96, P1, DOI DOI 10.1520/E1840-96R07.2
  • [5] Quick generation of Raman spectroscopy based in-process glucose control to influence biopharmaceutical protein product quality during mammalian cell culture
    Berry, Brandon N.
    Dobrowsky, Terrence M.
    Timson, Rebecca C.
    Kshirsagar, Rashmi
    Ryll, Thomas
    Wiltberger, Kelly
    [J]. BIOTECHNOLOGY PROGRESS, 2016, 32 (01) : 224 - 234
  • [6] CALIBRATION TRANSFER ACROSS NEAR-INFRARED SPECTROMETRIC INSTRUMENTS USING SHENKS ALGORITHM - EFFECTS OF DIFFERENT STANDARDIZATION SAMPLES
    BOUVERESSE, E
    MASSART, DL
    DARDENNE, P
    [J]. ANALYTICA CHIMICA ACTA, 1994, 297 (03) : 405 - 416
  • [7] Improvement of the piecewise direct standardisation procedure for the transfer of NIR spectra for multivariate calibration
    Bouveresse, E
    Massart, DL
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 32 (02) : 201 - 213
  • [8] Applications of Raman Spectroscopy in Biopharmaceutical Manufacturing: A Short Review
    Buckley, Kevin
    Ryder, Alan G.
    [J]. APPLIED SPECTROSCOPY, 2017, 71 (06) : 1085 - 1116
  • [9] Glucose concentration control of a fed-batch mammalian cell bioprocess using a nonlinear model predictive controller
    Craven, Stephen
    Whelan, Jessica
    Glennon, Brian
    [J]. JOURNAL OF PROCESS CONTROL, 2014, 24 (04) : 344 - 357
  • [10] Maintaining the predictive abilities of multivariate calibration models by spectral space transformation
    Du, Wen
    Chen, Zeng-Ping
    Zhong, Li-Jing
    Wang, Shu-Xia
    Yu, Ru-Qin
    Nordon, Alison
    Littlejohn, David
    Holden, Megan
    [J]. ANALYTICA CHIMICA ACTA, 2011, 690 (01) : 64 - 70