Predicting the clinical wear performance of orthopaedic bearing surfaces

被引:28
作者
Sauer, WL [1 ]
Anthony, ME [1 ]
机构
[1] Smith & Nephew Inc, Orthopaed Div, Memphis, TN 38116 USA
来源
ALTERNATIVE BEARING SURFACES IN TOTAL JOINT REPLACEMENT | 1998年 / 1346卷
关键词
clinical wear; wear test; historical; biomaterials; surface; cobalt-chrome; UHMWPE; standards;
D O I
10.1520/STP12827S
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A historical review of both laboratory and clinical wear performance of orthopaedic bearing materials and surfaces is presented. Over time, laboratory wear tests have been developed and used more frequently to support the introduction of new material couples. Historically, these methods have been unreliable, and, despite recent improvements, predicting clinical wear performance is still a challenging task. Orthopaedic wear is a highly multi-factorial process-successful simulation depends on the identification, modeling, and control of many critical factors. Clinical results themselves an highly variable, complicating efforts to establish clinical validation criteria for test methods. It is recommended that improvements be made in identifying the appropriate clinical targets for laboratory tests, and that increased effort be directed toward the optimization and standardization of wear test methods.
引用
收藏
页码:1 / 29
页数:29
相关论文
共 167 条
  • [1] METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS
    AGINS, HJ
    ALCOCK, NW
    BANSAL, M
    SALVATI, EA
    WILSON, PD
    PELLICCI, PM
    BULLOUGH, PG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) : 347 - 356
  • [2] AINSWORTH RD, 1977, HUMAN SURG IMPLANT S, P167
  • [3] AMSTUTZ HC, 1968, J BIOMED MATER RES, V3, P547
  • [4] AMSTUTZ HC, 1996, CLIN ORTHOP S, V329, pS11
  • [5] [Anonymous], CLIN ORTHOPAEDICS
  • [6] [Anonymous], CLIN ORTHOP
  • [7] LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .3. THE MEASUREMENT OF INTERNAL VOLUME CHANGES IN EXPLANTED CHARNLEY SOCKETS AFTER 2-16 YEARS INVIVO AND THE DETERMINATION OF WEAR FACTORS
    ATKINSON, JR
    DOWSON, D
    ISAAC, JH
    WROBLEWSKI, BM
    [J]. WEAR, 1985, 104 (03) : 225 - 244
  • [8] BERZINS A, 1995, T 18 SOC BIOM M, P92
  • [9] BIGSBY RJA, 1996, T 5 WORLD BIOM C, P970
  • [10] METALLOSIS ASSOCIATED WITH A STABLE TITANIUM-ALLOY FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT - A CASE-REPORT
    BLACK, J
    SHERK, H
    BONINI, J
    ROSTOKER, WR
    SCHAJOWICZ, F
    GALANTE, JO
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (01) : 126 - 130