A Graphenic Biosensor for Real-Time Monitoring of Urea During Dialysis

被引:26
|
作者
Bonini, Andrea [1 ]
Vivaldi, Federico Maria [1 ]
Herrera, Elisa [1 ]
Melai, Bernardo [1 ]
Kirchhain, Arno [1 ]
Sajama, Noemi Violeta Poma [1 ]
Mattonai, Marco [1 ]
Caprioli, Raffaele [2 ]
Lomonaco, Tommaso [1 ]
Di Francesco, Fabio [1 ]
Salvo, Pietro [3 ]
机构
[1] Univ Pisa, Dept Chem & Ind Chem, I-56124 Pisa, Italy
[2] Azienda Osped Univ Pisana, Nephrol Unit, I-56124 Pisa, Italy
[3] Natl Council Res, Inst Clin Physiol, I-56124 Pisa, Italy
关键词
Enzymatic biosensor; reduced graphene oxide; 4-amino benzoic acid; urease; pH; chronic kidney failure; dialysis; urea; CHRONIC KIDNEY-DISEASE; SPENT DIALYSATE; REFERENCE ELECTRODES; PH; BLOOD; TEMPERATURE; ABSORBENCY; REDUCTION; REMOVAL; CANNOT;
D O I
10.1109/JSEN.2020.2966456
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A patient at the fifth stage of chronic kidney disease usually needs dialysis or transplantation. The adequacy of dialysis can be assessed by the dimensionless parameter Kt/V, which depends on the concentration of urea in the blood. Kt/V is usually measured before and after treatment, or by an indirect measurement of dialysis fluid conductivity. For the real-time monitoring of the urea concentration during dialysis, this paper proposes a disposable biosensor that combines a pH sensitive reduced graphene oxide layer functionalized with 4-amino benzoic acid and urease. Urease is immobilized on reduced graphene oxide and catalyzes the hydrolysis of urea into carbon dioxide and ammonia. This catalysis increases the pH value locally and elicits a sensor response. Urea concentration was assessed in the plasma of five dialyzed patients using a potentiometric measurement. The response time was 120 s and the error was 6 +/- 3% compared with standard clinical laboratory analysis.
引用
收藏
页码:4571 / 4578
页数:8
相关论文
共 50 条
  • [1] A novel, low power biosensor for real time monitoring of creatinine and urea in peritoneal dialysis
    Premanode, Bhusana
    Toumazou, Chris
    SENSORS AND ACTUATORS B-CHEMICAL, 2007, 120 (02) : 732 - 735
  • [2] Robust, reliable biosensor for continuous monitoring of urea during dialysis
    DellaCiana, L
    Caputo, G
    CLINICAL CHEMISTRY, 1996, 42 (07) : 1079 - 1085
  • [3] Construction of a potentiometric biosensor for monitoring urea during peritoneal dialysis
    Cunningham, DD
    SENSOR LETTERS, 2004, 2 (3-4) : 217 - 220
  • [4] A robust and reliable biosensor for the continuous monitoring of urea during dialysis.
    DellaCiana, L
    Caputo, G
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1996, 212 : 51 - ANYL
  • [5] Fabrication of a Urea Biosensor for Real-Time Dynamic Fluid Measurement
    Kim, Kyunghee
    Lee, Jeongeun
    Moon, Bo Mi
    Seo, Ye Been
    Park, Chan Hum
    Park, Min
    Sung, Gun Yong
    SENSORS, 2018, 18 (08)
  • [6] Real-time liquid crystal-based biosensor for urea detection
    Khan, Mashooq
    Kim, Youngkyoo
    Lee, Joon Hyung
    Kang, Inn-Kyu
    Park, Soo-Young
    ANALYTICAL METHODS, 2014, 6 (15) : 5753 - 5759
  • [7] Real-time liquid crystal-based biosensor for urea detection
    Park, Soo-Young
    Choi, Sung-Mun
    Lee, Jong-Hun
    Khan, Mashooq
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2014, 248
  • [8] Smart sensors for real-time monitoring of patients on dialysis
    Fokko P. Wieringa
    Jeroen P. Kooman
    Nature Reviews Nephrology, 2020, 16 : 554 - 555
  • [9] Smart sensors for real-time monitoring of patients on dialysis
    Wieringa, Fokko P.
    Kooman, Jeroen P.
    NATURE REVIEWS NEPHROLOGY, 2020, 16 (10) : 554 - 555
  • [10] REAL-TIME MONITORING OF DNA MANIPULATIONS USING BIOSENSOR TECHNOLOGY
    NILSSON, P
    PERSSON, B
    UHLEN, M
    NYGREN, PA
    ANALYTICAL BIOCHEMISTRY, 1995, 224 (01) : 400 - 408