First tests of 6Li doped glass scintillators for ultracold neutron detection

被引:17
作者
Ban, G [1 ]
Fléchard, X
Labalme, M
Lefort, T
Liénard, E
Naviliat-Cuncic, O
Fierlinger, P
Kirch, K
Bodek, K
Geltenbort, P
机构
[1] ENSI, CNRS, IN2P3, Phys Corpusculaire Lab, F-14050 Caen, France
[2] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[3] Jagiellonian Univ, Inst Phys, PL-30059 Krakow, Poland
[4] Inst Max Von Laue Paul Langevin, F-38042 Grenoble, France
关键词
glass scintillators; UCN detectors;
D O I
10.6028/jres.110.040
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report the results of test measurements aimed at determining the performances of Li-6 doped glass scintillators for the detection of ultra-cold neutrons. Four types of scintillators, GS1, GS3, GS10 and GS20, which differ by their Li-6 concentrations, have been tested. The signal to background separation is fully acceptable. The relative detection efficiencies have been determined as a function of the neutron velocity. We find that GS10 has a higher efficiency than the others for the detection of neutrons with velocities below 7 m/s. Two pieces of scintillators have been irradiated with a high flux of cold neutrons to test the radiation hardness of the glasses. No reduction in the pulse height has been observed up to an absorbed neutron dose of 1 x 10(13) cm(-3).
引用
收藏
页码:283 / 288
页数:6
相关论文
共 3 条
[1]  
[Anonymous], 1992, NEUTRON NEWS
[2]   Development of solid-state silicon devices as ultra cold neutron detectors [J].
Baker, CA ;
Green, K ;
van der Grinten, MGD ;
Iaydjiev, PS ;
Ivanov, SN ;
Al-Ayoubi, S ;
Harris, PG ;
Pendlebury, JM ;
Shiers, DB ;
Geltenbort, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 487 (03) :511-520
[3]  
Knoll G. F., 1989, RAD DETECTION MEASUR