共 50 条
Assessing the detectability of antioxidants in two-dimensional high-performance liquid chromatography
被引:3
|作者:
Bassanese, Danielle N.
[1
]
Conlan, Xavier A.
[1
]
Barnett, Neil W.
[1
]
Stevenson, Paul G.
[1
]
机构:
[1] Deakin Univ, Sch Life & Environm Sci, Ctr Chem & Biotechnol, Geelong, Vic 3217, Australia
基金:
澳大利亚研究理事会;
关键词:
Antioxidants;
Countergradient;
Limit of detection;
Solvent strength mismatch;
Two-dimensional high-performance liquid chromatography;
TANDEM MASS-SPECTROMETRY;
DIODE-ARRAY DATA;
STEVIA-REBAUDIANA;
MOBILE-PHASE;
QUANTIFICATION;
PROTEOMICS;
INJECTION;
PEAKS;
2ND-DIMENSION;
COLUMNS;
D O I:
10.1002/jssc.201500054
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
This paper explores the analytical figures of merit of two-dimensional high-performance liquid chromatography for the separation of antioxidant standards. The cumulative two-dimensional high-performance liquid chromatography peak area was calculated for 11 antioxidants by two different methodsthe areas reported by the control software and by fitting the data with a Gaussian model; these methods were evaluated for precision and sensitivity. Both methods demonstrated excellent precision in regards to retention time in the second dimension (%RSD below 1.16%) and cumulative second dimension peak area (%RSD below 3.73% from the instrument software and 5.87% for the Gaussian method). Combining areas reported by the high-performance liquid chromatographic control software displayed superior limits of detection, in the order of 1 x 10(-6) M, almost an order of magnitude lower than the Gaussian method for some analytes. The introduction of the countergradient eliminated the strong solvent mismatch between dimensions, leading to a much improved peak shape and better detection limits for quantification.
引用
收藏
页码:1642 / 1648
页数:7
相关论文