A performance evaluation of eight geometrically different 10 Fr pediatric arterial cannulae under pulsatile and nonpulsatile perfusion conditions in an infant cardiopulmonary bypass model

被引:31
作者
Rider, Alan R. [1 ]
Ji, Bingyang [1 ]
Kunselman, Allen R. [2 ]
Weiss, William J. [3 ,4 ]
Myers, John L. [1 ,3 ]
Uendar, Akif [1 ,3 ,4 ]
机构
[1] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Pediat, Hershey, PA USA
[2] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Hlth Evaluat Sci, Hershey, PA USA
[3] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Surg, Hershey, PA USA
[4] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Bioengn, Hershey, PA USA
关键词
D O I
10.1097/MAT.0b013e3181744071
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This investigation compared pressure drops and surplus hemodynamic energy (SHE) levels in eight commercially available pediatric aortic cannulae (10 Fr) with different geometries during pulsatile and nonpulsatile perfusion conditions in an in vitro infant model of cardiopulmonary bypass. For each trial, the cannula was placed at the distal end of the arterial line, and the insertion tip was fixed to the inlet of the simulated patient. The pseudo patient was subjected to seven pump flow rates ranging from 400 to 1000 ml/min (at 100 ml/min increments), and the mean arterial pressure was set at a constant 40 mm Hg via Hoffman clamp. Of the eight cannulae, the Surgimedics and THI models had significantly larger pressure drops (48.8 +/- 0.3 mm Hg and 48.3 +/- 1.4 mm Hg, respectively; 600 ml/min pulsatile) compared with the RMI cannula (27.6 +/- 1.2 mm Hg; 600 ml/min pulsatile), which created, on average, half of the pressure drop seen in the poorest performing cannulae. When perfusion mode was switched from nonpulsatile to pulsatile, there was a 7-9 fold increase in delivery of SHE recorded at both the pre- and postcannulae sites, regardless of which cannula was being tested. Despite being classified under the same size (10 Fr), these eight cannulae were found to vary considerably in length, inner diameter, and geometrical design. The results suggest that these differences can have a significant impact on pressure drops, as well as generation and delivery of SHE. Furthermore, it was found that pulsatile perfusion produced more "extra" hemodynamic energy when compared with nonpulsatile perfusion, regardless of cannula model.
引用
收藏
页码:306 / 315
页数:10
相关论文
共 17 条
[1]   Effects of pulsatile and nonpulsatile perfusion on vital organ recovery in pediatric heart surgery:: A pilot clinical study [J].
Alkan, Tijen ;
Akcevin, Atif ;
Uendar, Akif ;
Tuerkoglu, Halil ;
Paker, Tufan ;
Aytac, Aydin .
ASAIO JOURNAL, 2006, 52 (05) :530-535
[2]   Comparison of perfusion modes on microcirculation during acute and chronic cardiac support:: is there a difference? [J].
Ji, Bingyang ;
Undar, Akif .
PERFUSION-UK, 2007, 22 (02) :115-119
[3]   An evaluation of the benefits of pulsatile versus nonpulsatile perfusion during cardiopulmonary bypass procedures in pediatric and adult cardiac patients [J].
Ji, Bingyang ;
Uendar, Akif .
ASAIO JOURNAL, 2006, 52 (04) :357-361
[4]   RANDOM-EFFECTS MODELS FOR LONGITUDINAL DATA [J].
LAIRD, NM ;
WARE, JH .
BIOMETRICS, 1982, 38 (04) :963-974
[5]   Pulsatile perfusion during cardiopulmonary bypass procedures in neonates, infants, and small children [J].
Rider, Alan R. ;
Schreiner, Robert S. ;
Uendar, Akif .
ASAIO JOURNAL, 2007, 53 (06) :706-709
[6]   Effects of pulsatile CPB on interleukin-8 and endothelin-1 levels [J].
Sezai, A ;
Shiono, M ;
Nakata, K ;
Hata, M ;
Iida, M ;
Saito, A ;
Hattori, T ;
Wakui, S ;
Soeda, M ;
Taoka, M ;
Umeda, T ;
Negishi, N ;
Sezai, Y .
ARTIFICIAL ORGANS, 2005, 29 (09) :708-713
[7]  
SHEPARD RB, 1966, ARCH SURG-CHICAGO, V93, P730
[8]   Comparison of four different pediatric 10F aortic cannulae during Pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass [J].
Uendar, Akif ;
Ji, Bingyang ;
Rider, Alan ;
Lukic, Branka ;
Kunselman, Allen R. ;
Weiss, William J. ;
Myers, John L. .
ASAIO JOURNAL, 2007, 53 (06) :778-784
[9]   Precise quantification of pulsatility is a necessity for direct comparisons of six different pediatric heart-lung machines in a neonatal CPB model [J].
Uendar, Akif ;
Eichstaedt, Harald C. ;
Masai, Takafumi ;
Bigley, Joyce E. ;
Kunselman, Allen R. .
ASAIO JOURNAL, 2005, 51 (05) :600-603
[10]   The type of aortic cannula and membrane oxygenator affect the pulsatile waveform morphology produced by a neonate-infant cardiopulmonary bypass system in vivo [J].
Undar, A ;
Lodge, AJ ;
Daggett, CW ;
Runge, TM ;
Ungerleider, RM ;
Calhoon, JH .
ARTIFICIAL ORGANS, 1998, 22 (08) :681-686